MR8740T_user_manual_eng_20191016H.pdf - 第121页
11 6 T riggering the Instrument at Regular Intervals (Interval T rigger) 5.8 T riggering the Instrument at Regular Intervals (Interval T rigger) Start triggers can be activated at specied intervals. Setting the recordin…

115
Triggering the Instrument With Logic Signals (Logic Trigger)
3
Click each of the signals under [Trigger pattern] to set a logic-trigger pattern.
The logic-trigger pattern can be set.
X
Ignores a signal.
0 The logic-trigger condition of each logic signal is satised when the signal is at a low
level.
1 The logic-trigger condition of each logic signal is satised when the signal is at a high
level.
Setting example
The logic-trigger conditions differ depending on the combination of the [Condition satised]
setting (logical OR or AND operation) and the [Trigger pattern] setting as follows:
Lx1 [1]
Lx2 [0]
Lx3 [X]
Lx4 [X]
Lx1 [1]
Lx2 [0]
Lx3 [X]
Lx4 [X]
OR
AND
Trigger pattern
Trigger pattern
Not satised Satised
Not satised Satised
Logic-trigger condition
Logic-trigger condition
5
Conguring the Trigger Settings

116
Triggering the Instrument at Regular Intervals (Interval Trigger)
5.8 Triggering the Instrument at Regular Intervals
(Interval Trigger)
Start triggers can be activated at specied intervals. Setting the recording mode to [Repeat] allows
the instrument to record waveforms at regular intervals.
• When using the pre-trigger, the instrument starts monitoring interval-trigger times after the rst
pre-trigger time elapses since the start of measurement.
• No start triggers are activated by any interval triggers while the instrument is lling the pre-trigger
memory. An interval trigger triggers the instrument while the instrument is waiting for a trigger
after the instrument has lled the pre-trigger memory.
• Since the clock is internally corrected, displayed trigger times may not synchronize with the
intervals of the interval trigger.
> [Trigger] > [Common]
2
3
1
1
Click [Interval trigger].
The interval trigger setting dialog box will appear.
2
Click [Interval trigger] to set it to [On].
3
Click the [h], [min], and [s] boxes, and then enter a period of time for interval triggers.
An internal trigger is generated at the start of measurement, and triggers are repeatedly generated at the
specied intervals.

117
Triggering the Instrument at Regular Intervals (Interval Trigger)
Acquiring data at regular intervals (relation between a time interval and a
recording length or recording time)
The instrument is not triggered until having acquired data that has the specied recording length or
recording interval.
When a recording length or recording time is
shorter than a time interval
When a recording length or recording time is
longer than a time interval
The instrument records the specied recording
length or recording time of the data.
Start
Time interval
Finish
Start
Time interval
Finish
The instrument records the specied recording
length or recording time of the data.
5