MR8740T_user_manual_eng_20191016H.pdf - 第144页

139 Conguring the Numerical Calculation Settings Calculation target channels and setting contents of calculation conditions for each calculation type Calculation type Setting Description Example of screen A verage value…

100%1 / 236
138
Conguring the Numerical Calculation Settings
4
Click the [Area] box, and then choose an option for a calculation range.
You can specify a calculation range for each calculation.
Whole
Makes calculations using whole waveforms.
A_B Makes calculations using waveforms between trace cursors A and B.
C_D Makes calculations using waveforms between trace cursors C and D.
Choosing [A_B] or [C_D] allows you to specify the calculation range with the trace cursors on the waveform
screen. When the instrument does not acquire any waveforms, perform another measurement and specify a
range. Doing so allows you to make calculations using data acquired within the specied range from the next
measurement.
5
Click a calculation target channel.
The channel selection screen appears.
Refer to “Calculation target channels and setting contents of calculation conditions for each calculation type”
(p. 139).
6
Set calculation conditions.
Refer to “Calculation target channels and setting contents of calculation conditions for each calculation type”
(p. 139).
7
Execute calculation.
(To calculate existing data)
Click [Execute].
(To automatically make calculation after measurement)
Click the start icon.
139
Conguring the Numerical Calculation Settings
Calculation target channels and setting contents of calculation conditions for
each calculation type
Calculation type Setting Description Example of screen
Average value
RMS
P-P
Maximum
Time to maximum
Minimum
Time to minimum
Standard
deviation
Area
High level
Low level
Intermediate value
Amplitude
Overshoot
Undershoot
Accumulation
Target channel
(Analog, waveform
calculation)
Allows you to specify
calculation target
channels.
With the [Area] or
[Accumulation] setting,
set [Method].
(Total
, Absolute value,
Positive, Negative)
Calculates an area or
accumulation according
to the method you
specify here.
7
Numerical Calculation Function
140
Conguring the Numerical Calculation Settings
Calculation type Setting Description Example of screen
Period
Frequency
Pulse width
Duty ratio
Target channel
(Analog, logic, waveform
calculation)
Allows you to specify
calculation target
channels.
Level*
The instrument makes
calculations using
values based on a
period of time when a
waveform crossed the
level specied here. No
setting is available for
logic channels.
Slope
No slope can
be specied for
calculating the
duty ratio.
Makes calculations
using values based on
a period of time when a
waveform crossed the
specied level in the
positive direction.
Makes calculations
using values based on
a period of time when a
waveform crossed the
specied level in the
negative direction.
Filter(samples)
(Off
, 10 to 10,000)
Allows you to enter
duration used to
determine that a
waveform crosses the
specied level.
Only if a waveform does
not cross the level again
within the specied lter
duration after it has
crossed the level, the
waveform is considered
to have crossed the
level.
This is useful to
eliminate false level-
crossing events caused
by noise.
Stat. Top
Detects a value
obtained rst from the
beginning within the
specied range.
Average Calculates the
average, maximum, or
minimum value for each
parameter within the
specied range.
Max
Min
*: The range you can specify are between −9.9999E+29 and −1.0000E-29, 0, and between +1.0000E-29 and
+9.9999E+29. A value to ve signicant gures can be set (a value to 10 signicant gures can be set for a