Conguring the Numerical Calculation Settings
Calculation target channels and setting contents of calculation conditions for
each calculation type
Calculation type Setting Description Example of screen
Average value
RMS
P-P
Maximum
Time to maximum
Minimum
Time to minimum
Standard
deviation
Area
High level
Low level
Intermediate value
Amplitude
Overshoot
Undershoot
Accumulation
Target channel
(Analog, waveform
calculation)
Allows you to specify
calculation target
channels.
With the [Area] or
[Accumulation] setting,
set [Method].
(Total
, Absolute value,
Positive, Negative)
Calculates an area or
accumulation according
to the method you
specify here.
7
Numerical Calculation Function

140
Conguring the Numerical Calculation Settings
Calculation type Setting Description Example of screen
Period
Frequency
Pulse width
Duty ratio
Target channel
(Analog, logic, waveform
calculation)
Allows you to specify
calculation target
channels.
Level*
The instrument makes
calculations using
values based on a
period of time when a
waveform crossed the
level specied here. No
setting is available for
logic channels.
Slope
No slope can
be specied for
calculating the
duty ratio.
Makes calculations
using values based on
a period of time when a
waveform crossed the
specied level in the
positive direction.
Makes calculations
using values based on
a period of time when a
waveform crossed the
specied level in the
negative direction.
Filter(samples)
(Off
, 10 to 10,000)
Allows you to enter
duration used to
determine that a
waveform crosses the
specied level.
Only if a waveform does
not cross the level again
within the specied lter
duration after it has
crossed the level, the
waveform is considered
to have crossed the
level.
This is useful to
eliminate false level-
crossing events caused
by noise.
Stat. Top
Detects a value
obtained rst from the
beginning within the
specied range.
Average Calculates the
average, maximum, or
minimum value for each
parameter within the
specied range.
Max
Min
*: The range you can specify are between −9.9999E+29 and −1.0000E-29, 0, and between +1.0000E-29 and
+9.9999E+29. A value to ve signicant gures can be set (a value to 10 signicant gures can be set for a
time of the level at time setting).

141
Conguring the Numerical Calculation Settings
Calculation type Setting Description Example of screen
Rise time
Fall time
Target channel
(Analog, waveform
calculation)
Allows you to specify
calculation target
channels.
Time (%)
(5%→95%
to 30% →
70% or 95% → 5%
to
70%→30%)
Allows you to specify
which part of a
waveform between the
upper and lower limits is
used for calculating the
rise time (or fall time).
A rise time (or fall time)
is calculated based on
the percentage, which
is specied here, of
the range between the
upper and lower limit
values.
Stat. Top
Detects a value
obtained rst from the
beginning within the
specied range.
Average Calculates the
average, maximum, or
minimum value for each
parameter within the
specied range.
Max
Min
X-Y area
Angle of XY
waveform
Target channel
(X, Y)
(Analog, waveform
calculation)
Allows you to assign
channels to the X- and
Y-axis.
With the [X-Y area]
setting, set [Method].
(Coordinate
method
, Trapezoidal
approximation)
Calculates an X-Y
area according to a
calculation method.
7
Numerical Calculation Function