148 Evaluating Calculation Results on a Pass/Fail Basis 7.3 Evaluating Calculation Results on a Pass/Fail Basis Y ou can specify evaluation criteria ( [Up] and [Low] ) to evaluate numerical calculation results on a pass/…

100%1 / 236
147
Conguring the Numerical Calculation Settings
Displaying numerical calculation results
You can check calculation results on the waveform screen.
> [Numeric Calc.]
a
b c
You can display or hide the screen of the numerical calculation results every time you click [CH
info].
If no periods are found or a calculation is aborted, the character string [******] appears instead of
a calculation result.
For channels not specied as a calculation target, the character, [–] appears.
Click [ ] under the [CH Info] toggle switch, and then choose [Calcu Result] from the list (
a
).
To save calculation results after measurement
Refer to “Freely selecting data items to be saved and save les (save icon)” (p. 90).
To change the display width of the calculation result screen
You can change the display width of the calculation results by clicking [–] or [+] (
b
).
You can choose calculation numbers to be displayed by clicking the buttons (
c
).
You can specify up to four calculation results to be displayed.
Choose calculation numbers you want to display from the list.
7
Numerical Calculation Function
148
Evaluating Calculation Results on a Pass/Fail Basis
7.3 Evaluating Calculation Results on a Pass/Fail
Basis
You can specify evaluation criteria ([Up] and [Low]) to evaluate numerical calculation results on a
pass/fail basis.
Each numerical calculation can has different evaluation criteria.
The waveform acquisition process varies depending on the recording mode setting ([Single] or
[Repeat]) and the stop condition specied to stop the measurement depending on an evaluation
([PASS], [FAIL], or [PASS & FAIL]).
Conguring the calculation
settings
Starting a measurement
Acquiring data
Calculation
Displaying calculation
results
Evaluating calculation
results
Saves data.
Finishing the measurement
Stop condition
Recording mode
No stop
condition is
satised.
A stop condition is satised,
or stop condition is set to
[PASS & FAIL].
[Repeat]
[Single]
Measurements repeat
until the stop condition is
satised.
Measurements repeat
until you click the stop
icon.
Auto-save is not carried out until the stop condition is satised at the end of calculation evaluation.
149
Evaluating Calculation Results on a Pass/Fail Basis
> [Calculation] > [Numerical calculation]
4
1
2
3
1
Click the [Judge stop condition] box, and then from the list, choose a stopping condition
applied according to judgments.
PASS Stops measurement when a calculation result falls within a criteria range (pass
judgment).
FAIL Stops measurement when a calculation result is outside a criteria range (fail judgment).
PASS & FAIL
Stops measurement regardless whether a pass or fail judgment is given.
2
Click the [Judge] box, and then from the list, choose whether or not to evaluate calculation
results.
Off
Evaluates calculation results.
On Yields a fail judgment if a result is outside a criteria range.
Calculated values in a channel with the fail judgment are highlighted in red.
3
Click the [Up] and [Low] boxes in turn, and then enter evaluation criteria values in each of
them.
Enterable range: −9.9999E+29 to −1.0000E29, 0, +1.0000E29 to +9.9999E-29
(To evaluate existing data)
Click [Execute].
(To automatically execute evaluation after measurement)
Click the start icon.
About upper and lower limit values
You cannot specify an upper limit value lower than a lower limit value.
Neither can you specify a lower limit value higher than a upper limit value.
To record all calculation results
Choose [PASS & FAIL] for the stopping condition.
7
Numerical Calculation Function