60
Conguring Measuring-Module-Specic Settings
Conguring Model U8977 3CH Current Unit Settings
Each of Model 9709, Model CT6860 series, and Model CT6840 series consists of a current sensor
with the sub model-number “-05,” which has a metal connector, and that without the sub model-
number “-05,” which has a black plastic connector.
You can directly connect a current sensor that has the sub model-number “-05,” which has a
metal connector, with Model U8977 U8977 3CH Current Unit. Using Model CT9900 Conversion
Cable, you can connect a current sensor without the sub model-number “-05,” which has a plastic
connector (PL23), with Model U8977 3CH Current Unit.
> [Channel] > [U8977]
3
1
2
Current sensors to be connected determine whether the instrument automatically recognizes mode
or whether you have to choose a current sensor setting.
1
(When the instrument automatically recognizes mode) Conrm the output rate displayed in
the [Mode] area.
Available measurement ranges, which depend on automatically recognized output rates, are as follows:
20A/2V 2 A, 4 A, 10 A, 20 A, 40 A, 100 A
50A/2V 4 A, 10 A, 20 A, 40 A, 100 A, 200 A
200A/2V 20 A, 40 A, 100 A, 200 A, 400 A, 1 kA
500A/2V 40 A, 100 A, 200 A, 400 A, 1 kA, 2 kA
1000A/2V 100 A, 200 A, 400 A, 1 kA, 2 kA, 4 kA
2000A/2V 200 A, 400 A, 1 kA, 2 kA, 4 kA, 10 kA
None When no current sensor is connected.
61
Conguring Measuring-Module-Specic Settings
(When you have to choose a current sensor setting) Click the [Mode] box, and then choose
a connected current sensor.
Available measurement ranges, which depend on chosen current sensors, are as follows:
CT7631/CT7731 200 A
CT7636/CT7736 200 A, 400 A, 1 kA
CT7642/CT7742 2 kA, 4 kA
CT7044/CT7045/
CT7046
2 kA, 4 kA, 10 kA
0.1mV/A 2 kA, 4 kA, 10 kA, 20 kA, 40 kA, 100 kA
1mV/A 200 A, 400 A, 1 kA, 2 kA, 4 kA, 10 kA
10mV/A 20 A, 40 A, 100 A, 200 A, 400 A, 1 kA
100mV/A 2 A, 4 A, 10 A, 20 A, 40 A, 100 A
1000mV/A 0.2 A, 0.4 A, 1 A, 2 A, 4 A, 10 A
IMPORTANT
When Model CT6846 or Model CT6865 connects via Model CT9900 Conversion Cable, the
instrument recognizes the sensor as a 500 A AC/DC sensor. Set the conversion ratio at 2.00 in
the scaling setting.
When a current sensor included in Model CT6700 series is connected via Model CT9920
Conversion Cable, set a current sensor model name or its output rate.
Make sure to execute zero-adjustment after you change the setting. Execute zero-adjustment
without any input.
2
(When you have changed the measurement mode) Click [Zero adjust].
The instrument performs zero-adjustment. Execute zero-adjustment without any input.
3
Click the [Range (f.s.)] box, and then choose a measurement range from the list.
The instrument automatically congures the scaling setting for a measurement range according to a
recognized current sensor.
IMPORTANT
The gure of each measurement range name represents the maximum current that Model U8977
can measure using the range. However, the instrument cannot measure currents that exceed the
rated current of a connected current sensor. Check the specications of the current sensor used.
3
Advanced Functions
62
Conguring Measuring-Module-Specic Settings
Congure Model U8979 Charge Unit Settings
This setting allows you to choose between voltage measurement and acceleration measurement
(charge-output or built-in pre-amplier) for an input channel.
A channel can measure either one of them.
Voltage mode and Preamp mode use BNC connectors, whereas Charge mode uses miniature
connectors.
Model U8979 can automatically recognize TEDS-compliant* sensors.
*: Transducer electronic data sheet
WARNING
Setting the measurement mode to [Preamp] allows Model U8979 Charge Unit
to constantly provide power (3.5 mA, 22 V) to sensors. Set any measurement
mode other than [Preamp] or tum off the instrument before connecting a sensor
or probe with a BNC terminal to avoid an electric shock or damage to the
measurement target.
> [Channel] > [U8979]
3
1
2
4
5
1
Click the [Mode] box, and then choose a measurement mode from the list.
Mode Measurement target Measurement sensitivity
Voltage Voltage
Charge
Charge-output acceleration sensor 0.1 pC/(m/s
2
) to 10 pC/(m/s
2
)
Preamp Acceleration sensor with a built-in pre-
amplier
0.1 mV/(m/s
2
) to 10 mV/(m/s
2
)
2
(When setting mode to [Preamp]) Click [TEDS Load].
Acquires sensitivity of a connected sensor. However, the instrument can acquire sensitivity of TEDS-compliant
acceleration sensors with a built-in pre-amplier only.
When sensor sensitivity has been acquired, it is automatically set.
3
Click the area that includes [A.A.F].
The setting dialog box appears.