70 Conguring Generator-Module-Specic Settings 4 Output conguration ( Method ) Y ou can congure the output settings. The ground potential is cross-channel and not insulated. With the open-collector output setting • Li…

100%1 / 236
69
Conguring Generator-Module-Specic Settings
Conguring Model MR8791 Pulse Generator Unit settings
1
Click [MR8791].
6
1
3
4
5
2
2
Click the [Mode] box, and then choose an output type from the list.
PULSE
Outputs a pulse waveform.
PATTERN Outputs a pattern waveform.
3
Congure the pulse output settings.
Frequency
Allows you to enter a pulse output frequency. (In 0.1 Hz increments)
0.0 Hz to 20000.0 Hz
Duty ratio
Allows you to enter a pulse duty ratio. (In 0.1 percent point increments)
With the 0% setting, the low-level signal is outputted; with the 100% setting, the high-level signal is
outputted. No pulse is outputted.
With the 100% setting, the low-level signal is outputted even if the output setting is disabled.
The output pulse width setting of 1 µs, which can be derived from the frequency and duty ratio settings, may
cause absence of pulses.
0.0% to 100.0%
3
Advanced Functions
70
Conguring Generator-Module-Specic Settings
4
Output conguration (Method)
You can congure the output settings.
The ground potential is cross-channel and not insulated.
With the open-collector output setting
Limit the collector-emitter voltage to 50 V.
Output signals require a response time of about 5 µs or less to rise from 10% to 90%. (With a load capacity
of 1000 pF and a pull-up resistamce of 1 k
Ω
, values for reference purposes)
TTL
TTL-level output (amplitude: 0-5 V)
OC Open-collector output
5
Output
Allows you to switch waveform output settings.
The instrument may momentary output a high-level signal at the times of power-on and power-off.
OFF
Does not output any waveform
ON Outputs a waveform
6
Conrmation of the setting
After nishing the setting, click
[Send setting]
.
The instrument conrms the settings and outputs a waveform.
Conguring the pattern output settings
Pattern number
Enter an pattern number.
When no output is enabled, the instrument outputs the pattern signal registered rst.
Turning off the instrument clears pattern data sets. After turning on the instrument, re-register pattern data
set.
1 to 16
71
Conguring Generator-Module-Specic Settings
Pattern reproducing frequency
Enter a clock frequency of pulse pattern. (In 10 Hz increments)
Settable clock period range: 0 to 0.1 seconds
You can freely specify the clock cycle as long as it falls into the above-mentioned range. The cycle of the
actually outputted waveform becomes the nearest value among the reciprocals of settable frequencies.
0 Hz to 120000 Hz
Output conguration
Choose an output status.
The ground potential is cross-channel and not insulated.
TTL
TTL-level output (amplitude: 0-5 V)
OC Open-collector output
With the open-collector output setting
Limit the collector-emitter voltage to 50 V.
Output signals require a response time of about 5 µs or less to rise from 10% to 90%. (With a load capacity
of 1000 pF and a pull-up resistance of 1 k
Ω
, values for reference purposes)
Output
Switch waveform output settings.
The instrument may momentary output a high-level signal at the times of power-on and power-off.
OFF
Does not output a waveform
ON Outputs a waveform
Conrmation of the setting
After nishing the setting, click
[Send setting]
.
The instrument conrms the settings and outputs a waveform.
3
Advanced Functions