00196433-02_AI_Vakuumpumpe_X-Serie_70XDE+EN.pdf - 第121页

Assembly instructions: Vacuum pump for SIPLACE X-series 04/2009 Edition 121 2.1 1.2 Checking the vacuum circuit Fig. 2.11 - 1 Measuring point on the vacuum distributor b lock (machine input) 2.1 1.2.1 Checking the vacuum…

100%1 / 128
Assembly instructions: Vacuum pump for SIPLACE X-series
04/2009 Edition
120
2.11 Final tasks
2.11.1 Conditions for switching on
2
Before switching on, run the following checks to prevent any risk of injury or serious damage: 2
Before switching on the compressed air again, make sure that 2
all hose connections / joints are assigned correctly
(and that there is no compressed air in any of the vacuum hoses).
all the requirements for switching on the vacuum pump and the connected placement machine
as
listed in the user manual
are fulfilled.
the vacuum pump is always switched on before the placement machine.
2
: Open the compressed air stop valve.
: Push the pneumatic unit into the placement machine.
: Close all the doors.
: Switch on at the main power switch.
Assembly instructions: Vacuum pump for SIPLACE X-series
04/2009 Edition
121
2.11.2 Checking the vacuum circuit
Fig. 2.11 - 1 Measuring point on the vacuum distributor block (machine input)
2.11.2.1 Checking the vacuum system for the C&P20 holding circuit for leaks
2
2
There should be a vacuum of at least -500 mbar (for very large nozzles 1235) to
-600 mbar (for very small nozzles 1006) at each placement machine/head depending on the noz-
zle configuration. 2
The measurement is taken at a nozzle using an external
measuring device that corresponds to a
closed nozzle (with component). 2
2
Measuring point on the vacuum distributor block, machine input (see Fig. 2.11 - 1): 2
2
2
The vacuum should never drop below -400 mbar in operation. If this does occur, please contact
the SIPLACE Hotline. 2
Assembly instructions: Vacuum pump for SIPLACE X-series
04/2009 Edition
122
Leak test procedure for the C&P20 head: 2
2
To avoid deviations due to offset errors, it is a good idea to take both measurements using the
same measuring device. 2
2
: Measure the vacuum using an external pressure measuring device:
at a connection on the “Distributor, placemen
t head vacuum” on the gantry (see Fig. 2.8 - 5)
and then
at an open nozzle in the holding circuit on the placement head.
The pressure differential must not exceed 30-50 mbar, otherwise there is a leak somewhere. 2
2
The vacuum test (in SITEST) is always measured at an open nozzle. For large nozzles (1235),
the vacuum drops away to values below -100 mbar due to the increased volume of leak air. This
sharp drop – for large nozzles without a component – is caused by the small diameter of the re-
ducing nozzle in the orifice ring compared to the nozzle diameter.
With a component, vacuum values of -500 to -600 mbar are achieved. 2
If the above values are not achieved
, the cause is gen
erally a hose that is not pushed on fully in
the vicinity of the head. Follow these steps: 2
: Use the tubing nipp
ers to make sure that all the hoses are pressed on correctly
: Repeat the vacuum measurement.
2.11.2.2 Checking that the vacuum system for the C&P20 holding circuit is working overall
Run a vacuum test in SITEST to check all the functions of the vacuum system as for the Venturi
principle. 2
The vacuum sensor in the holding circuit merely serves
to check that the small nozzles in the or-
ifice ring are unsoiled. 2
2.11.2.3 Checking the vacuum system for the C&P6/12 holding circuit for leaks
If the vacuum pump is used for the X-series (C&P20), it differs from the vacuum pump for the HS-
60 / HF as follows: 2
The flow rate is 4x higher than with the “Vacuu
m pump for HS-xx / S-27HM / HF“ (item no.
00119017)
The maximum vacuum achieved is now approx. -640 mbar, rather than the previous -850 mbar.
For th
is reason, the minimum permitted vacuum value for the holding circuit must be reduced
from -750 to -400, otherwise, the error message “258 Vacuum in holding circuit is too low (tar-
get/actual)“ will occur during the reference run.