2 Maintenance
2.7 Cooling Tract
Operating Instructions
Version 1.5
2.7.3 Clean cooling module 1 VXP+
Fig. 2-36 Disconnect power supply
Fig. 2-37 Lower nozzle field
Fig. 2-38 Unlatch, remove the cooling mod-
ule
Consumables, tools:
Oven cleaner
Cleaning rags
Cleaning bath
Procedure:
1. First disconnect the electric
heater supply at the back of
the plant. Then pull out the
heater at the back.
Now remove cooling module 1
at the front of the machine for
cleaning, as follows.
2. Rotate the wing screws at the
right and left at the nozzle field
by 90°, remove the metal safety
plate with a knurled screw and
lower the nozzle field for
cleaning.
3. Release the cooling cartridge’s
tension lock, raise the cartridge
and lift it forwards and out.
4. Place the cooler cassette in the
cleaning bath.
5. Also clean the inside area as
well as the fan wheel.
6. Re-fit the cleaned cooler
cassette.
7. Push the heater at the back of
the machine back in and plug in
the power supply.
Page 30 Vision XP+ VAC
2 Maintenance
2.7 Cooling Tract
Operating Instructions
Version 1.5
2.7.4 Clean trapped heat cooler
2.7.5 Inspecting and Cleaning the Fan Impeller
Fig. 2-39 Trapped heat cooler
Fig. 2-40 Trapped heat cooler
Consumable materials, tools:
Vacuum cleaner
Brush
Procedure:
1. Occasionally check the cooling
fins of the trapped heat cooler
for dirt
2. Vacuum the cooling fins as
soon as they are covered with
dust and dirt. Also use a brush,
if necessary.
Fig. 2-41 Cleaning the Fan Impeller
The fan impellers are located in the
process chamber behind the cooler
cartridges. (optional for undercool-
ing)
Consumable materials, tools:
Oven cleaner
Rags
Procedure:
1. Remove the cooler cartridge
(see instructions in chapter
2.7.2).
2. Inspect the fan impellers for
contamination, and clean with
oven cleaner and rags if
necessary.
Vision XP+ VAC Page 31
2 Maintenance
2.7 Cooling Tract
Operating Instructions
Version 1.5
2.7.6 Replacing the Fan Impeller
Fig. 2-42 Replacing Fan Impeller (step 1)
Fig. 2-43 Replacing Fan Impeller (step 2)
Fig. 2-44 Replacing Fan Impeller (step 3
Fig. 2-45 Replacing Fan Impeller (step 4)
Consumable materials, tools:
17 mm box wrench or open-end
wrench
Puller
Procedure:
1. Open the cooling zone.
2. Remove the self-locking bolt at
the fan impeller.
3. Attach the puller carefully. The
fan deducted evenly
4. Set the fan impeller into a rins-
ing bath, or replace it with a new
one.
5. Put on the fan wheel properly.
Now, put on the locking screw
again.
6. Screw tightly the fan wheel with
locking screw. Close again the
cover with clamping levers.