Page 32 Vision XP+ V AC 2 M aintenance 2.8 V acu um Unit Operating Instructions V ersion 1.5 2.8 V acuum Unit 2.8.1 Detail view of the vacuum chamber Fig. 2-4 6 Detail vi ew of the vacuum chamber 1. Clean the sensor glas…

100%1 / 142
Vision XP+ VAC Page 31
2 Maintenance
2.7 Cooling Tract
Operating Instructions
Version 1.5
2.7.6 Replacing the Fan Impeller
Fig. 2-42 Replacing Fan Impeller (step 1)
Fig. 2-43 Replacing Fan Impeller (step 2)
Fig. 2-44 Replacing Fan Impeller (step 3
Fig. 2-45 Replacing Fan Impeller (step 4)
Consumable materials, tools:
17 mm box wrench or open-end
wrench
Puller
Procedure:
1. Open the cooling zone.
2. Remove the self-locking bolt at
the fan impeller.
3. Attach the puller carefully. The
fan deducted evenly
4. Set the fan impeller into a rins-
ing bath, or replace it with a new
one.
5. Put on the fan wheel properly.
Now, put on the locking screw
again.
6. Screw tightly the fan wheel with
locking screw. Close again the
cover with clamping levers.
Page 32 Vision XP+ VAC
2 Maintenance
2.8 Vacuum Unit
Operating Instructions
Version 1.5
2.8 Vacuum Unit
2.8.1 Detail view of the vacuum chamber
Fig. 2-46 Detail view of the vacuum chamber
1. Clean the sensor glass.
2. Clean and lubricate splined shaft and spindles.
3. Grease the articulated shaft.
4. Clean or grease the transport chains
Note!
The emergency stop function must first be deactivated before undertaking
any maintenance work in the vacuum chamber.
Note!
After the maintenance the chamber seal should be lubricated with
vacuum grease.
1
1
3
2
4
Vision XP+ VAC Page 33
2 Maintenance
2.8 Vacuum Unit
Operating Instructions
Version 1.5
2.8.2 Check that the vacuum chamber is sealed
2.8.3 Vacuum chamber transfer height
Fig. 2-48 Vacuum chamber transfer height
Detailed data for adjusting the transfer height into or out of the vacuum
chamber.
Fig. 2-47 vacuum chamber
To check that the vacuum unit is
sealed, a vacuum cycle should be
run. If the cycle finishes in the spec-
ified time and with the preset pres-
sure (approx. 10 mbar), the vacuum
unit is sealed.
However, if the preset time is ex-
ceeded or the pressure not reached,
the vacuum unit has a leak, which
must be addressed.
pass-through direc-
tion