IPC-TM-650 EN 2022 试验方法-- - 第583页

cur re nt pre se t on t he Hi pot tes t in str umen t , ca us ing the instrument to indicate a failure when in fact there was n one. The ch a rgi ng cur r ent o f the c ap acit or i s a ff ecte d by t he change in voltag…

100%1 / 824
3.3 Test Specimen Conditioning
All qualification test
specimens
be conditioned at 23 °C ± 3 °C and 50% ±
10% RH for 24 hours, before testing. For conformance test-
ing, such conditioning is optional.
4 Apparatus
4.1 Hipot Test Instrument
A Hipot test instrument is a
piece of equipment capable of supplying a range of DC test
voltages appropriate for the materials under test with adjust-
able ramp rate and hold-time settings. The Hipot equipment
have an adjustable threshold current setting. (see 5.2.4)
The user
ensure that the Hipot test instrument satisfies
the original manufacturer’s technical specifications.
4.2 High Voltage Connections
Contacts (conductor
plates) apply the voltage from the Hipot test instrument to the
test specimen’s Top Imaged Foil and Bottom Foil (see Figure
1). These contacts should not contain sharp points that could
damage either the copper foil or the dielectric layers of the test
specimens.
Dangerous voltages may be present on the test
connections. Use proper machine guarding and/or machine
interlocking.
5 Procedure
5.1
This test method be performed on fresh test
specimens. Hipot testing
be conducted on test
specimens that have previously been exposed to high voltage
levels or other similar testing.
Some dielectrics may show acceptable Hipot results
(i.e., ‘‘Pass’’) after defects have been ‘‘burned out’’ at high
voltage (see Section 6).
5.2
Program the Hipot test instrument with the appropriate
peak voltage, voltage ramp rate, hold time at peak voltage
and current threshold level. These values
be recorded.
5.2.1
The peak voltage should be as specified in the mate-
rial Specification Sheet (see IPC- 4821) under the parameter
‘‘Hipot (Volts DC).’’
5.2.2
For qualification testing, the voltage ramp rate
be 5% of the peak voltage per second, unless otherwise
specified. For conformance testing, the voltage ramp rate
be 5% of the peak voltage per second or AABUS.
5.2.3
For qualification testing, the hold time at peak voltage
be 30 seconds +3 / -0 seconds. For conformance test-
ing, the hold time at peak voltage
be a minimum of 10
seconds or AABUS.
5.2.4
The threshold settings be set to a value greater
than the in-rush current (due to the charging of the test speci-
men) observed when the voltage is increased (see 6.1). Many
commercial Hipot test instruments display the current during
the test. The in-rush current can be determined by setting the
threshold current to a high value and then observing the cur-
rent spikes as the voltage is ramped to the peak voltage. After
several test specimens have been tested and the currents
observed, set the threshold current to be greater than the
highest current observed. For example, if the in-rush current is
20 microamperes and the current at peak voltage is 1 micro-
ampere, set the threshold current to 40 microamperes.
5.3
The test specimen be placed between the con-
tacts of the Hipot test equipment (see Figure 1). Start the
Hipot sequence.
5.4
Upon completion of the test, the Hipot sequence should
include the discharge of the test specimen.
Larger test specimens, with high capaci-
tance density, may take more time than expected to dis-
charge.
5.5 Reporting
The Hipot test instrument indicates either
Pass or Fail of the material under test. A current surge above
the threshold current setting indicates a Failure. This includes
very short term current surges or ‘‘arcs’’ that occur due to the
burnout of defects. Such unacceptable current surges may
also be the result of dielectric failure or manufacturing defects.
For qualification testing, if the test specimen Passes, record
the leakage current per unit area and the passing voltage. If
the qualification test specimen Fails, record the failure voltage
and threshold current per unit area of each test specimen.
For conformance testing, reporting requirements should be
AABUS.
6 Notes
6.1
When the Hipot test instrument voltage changes from
one level to the next higher level during the ramp-up to the
final voltage, the in-rush current will initially surge above the
steady state current because the capacitor is charging. It is
possible that this surge in current could exceed the threshold
Number
2.5.7.2
Subject
Dielectric Withstanding Voltage (Hipot Method) - Thin Dielectric
Layers for Printed Boards
Date
11/2009
Revision
A
IPC-TM-650
shall
shall
shall
shall
shall
shall
shall
CAUTION:
SAFETY
NOTE:
NOTE:
shall
shall
not
shall
shall
Page
2
of
3
current preset on the Hipot test instrument, causing the
instrument to indicate a failure when in fact there was none.
The charging current of the capacitor is affected by the
change in voltage from one ramp step to another, the dielec-
tric constant of the dielectric, the thickness of the dielectric
and the area of the capacitor. High dielectric constant, very
thin dielectric thickness and large area of the capacitor plates
will all cause the charging current to increase. As a result, the
threshold current setting on the Hipot test instrument may
need to be adjusted to avoid generating a false failure condi-
tion.
6.2
Some thin and filled dielectrics will require a higher
threshold current setting, compared to unfilled materials. This
is particularly true of dielectrics containing ferroelectric com-
pounds, such as barium titanate. These materials may show
a nonlinear response between current and voltage. This is not
an issue at most operating voltages, which are normally low,
but can be an issue for the Hipot test. At high voltage levels,
these materials may trigger a false failure because they allow
more current than the threshold setting.
6.3
Materials, especially very thin and/or highly filled materi-
als, may have a leakage current per unit area that is area
dependent when tested at the specified test voltage. Thus,
results from the qualification test specimen or other small area
test structures may not reflect the actual leakage current per
unit area when the material is tested in a significantly larger
area format, such as that commonly done in conformance
testing by a material supplier.
6.4 Reference Documents
‘‘Dielectric Breakdown Voltage of Solid Electri-
cal Insulating Materials at Commercial Power Frequencies’’
Number
2.5.7.2
Subject
Dielectric Withstanding Voltage (Hipot Method) - Thin Dielectric
Layers for Printed Boards
Date
11/2009
Revision
A
IPC-TM-650
ASTM
D149
Page
3
of
3
Figure 1 Suggested Flex Circuit Layout for Insulation
Resistance Test
Figure 2 Preattachment of the ACF Strips to the Flex
Circuit
The Institute for Interconnecting and Packaging Electronic Circuits
2215 Sanders Road Northbrook, IL 60062
Material in this Test Methods Manual was voluntarily established by Technical Committees of the IPC. This material is advisory only
and its use or adaptation is entirely voluntary. IPC disclaims all liability of any kind as to the use, application, or adaptation of this
material. Users are also wholly responsible for protecting themselves against all claims or liabilities for patent infringement.
Equipment referenced is for the convenience of the user and does not imply endorsement by the IPC.
Page 1 of 2
Number
Testable
Area
should
be
25mm
IPC-TM-650
TEST
METHODS
MANUAL
1
Scope
The
purpose
of
this
test
method
is
to
quickly
assess
the
adequacy
of
a
given
Anisotropically
Conductive
Adhesive
Film
(ACF)
construction
and
bonding
process
for
avoiding
short
circuits
between
adjacent
traces
of
a
flex
circuit
being
bonded
to
a
low
profile
circuit
substrate.
1.1
Purpose
ACF
materials
are
often
used
to
interconnect
fine-pitch
flexible
circuitry
to
substrates
such
as
flat-panel
dis¬
plays.
A
center
to
center
pitch
range
of
80
pm
to
200
pm
is
not
uncommon
in
circuits
for
flat
panel
display
applications.
It
is
critical
that
the
particle
dispersion
within
the
ACF
be
of
suf¬
ficient
quality
such
that
there
is
no
inherent
tendency
for
short
circuits
between
adjacent
traces.
In
addition,
it
is
important
that
a
bonding
process
is
used,
which
doesn't
create
any
undue
accumulation
of
particles,
which
will
lead
to
short
cir¬
cuits.
2
Applicable
Documents
None
2.5.10.1
Subject
Insulation
Resistivity
for
Adhesive
Interconnection
Bonds
Date
Revision
11/98
Originating
Task
Group
SMT
Mounting
Adhesives
Task
Group
(5-24d)
3
Test
Specimens
3.1
In
order
to
perform
this
test,
a
custom-designed
and
fabricated
flex
circuit
substrate
will
need
to
be
produced.
A
suggested
flex
circuit
construction
of
a
design
is
shown
in
Figure
1
.
Flex
circuit
materials
should
be
selected
to
be
repre¬
sentative
of
what
is
being
used
in
the
application.
The
traces
alternate
between
anodic
and
cathodic
polarity
as
shown.
Trace
thickness,
width,
and
pitch,
should
be
selected
in
accordance
with
the
application.
Total
trace
count
should
be
at
least
1
00,
and
total
width
of
the
pattern
should
be
slightly
less
than
the
thermode
length.
Total
length
of
the
traces
should
be
sufficient
to
allow
at
least
four
bonds
to
be
accom¬
modated
as
shown
in
Figure
2
and
Figure
3.
N
is
the
number
of
circuit
traces
(at
least
100).
I
is
the
measured
leakage
current
in
amps
after
10
seconds
@
50V.
g
is
the
gap
spacing
between
adjacent
traces
on
the
circuit
in
mm
(of
the
order
0.04
mm
to
0.1
mm).
wi
a
h
IPC-2-5-1
0-1-1
IPC-2-5-10-1-2