SussPA300UserGuide - 第48页
SUSS / PA300 / User Manual / M10-121841-00 / February 2002 47 Weight Device without table 500 kg Device with table 600 kg Electronics 13 kg Temperature Optimal operating range 19°C ... 24°C Target temperature 22°C Tolera…

46
SUSS / PA300 / User Manual / M10-121841-00 / February 2002
Utilities
Vacuum Less than 200 mbar absolute , flow rate
insignificant.
Compressed Air 0.4 MPa to 1 MPa or 4 bar to 10 bar, flow
rate insignificant.
Vacuum/Air Inputs US: 1/4“, all other: 6 mm.
Ergonomic Information
Probe station Horizontal usage position
Electronics ProberBench Horizontal usage position
Dimensions (W x D x H)
Mechanic (Table device): H = min. 950 mm x 900 mm x 770 mm with
microscope in raised position. Cameras or
laser cutters will require additional
clearance.
Mechanic (Standard device):
H = min. 950 mm x 900 mm x 1430 mm (see above)
Electronics 520 mm x 600 mm x 160 mm
with connectors installed.
Joystick Controller 260 mm x 260 mm x 120 mm
with connector installed.
Packaging
System box Length: 1230 mm
Width: 1200 mm
Height: 1120 mm
Net weight: 498 kg
Gross weight: 607 kg
System table with palette Length: 1050 mm
Width: 1000 mm
Height: 850 mm
Net Weight: 97 kg
Gross weight: 102 kg
Accessories Length: 800 mm
Width: 1200 mm
Height: 800 mm
Gross weight: approx. 80 kg

SUSS / PA300 / User Manual / M10-121841-00 / February 2002
47
Weight
Device without table 500 kg
Device with table 600 kg
Electronics 13 kg
Temperature
Optimal operating range 19°C ... 24°C
Target temperature 22°C
Tolerance 1K
Max. rate of change/hour Not specified
Electronics ProberBench Side ventilators and air expellers must be
kept clear for air circulation at all times.
Humidity
Tool area 40 % ... 60 %
Support equipment area 40 % ... 60 %
Noise
Audio Noise Constant level: 56 dB (A) Impulse: 72 dB (A)
Pollution Level 1 IEC 664
Clean Room Class Class 6 corresponding to
DIN EN ISO 14664-1
Electrical Data
Power 230 V, 50 Hz, 510 VA or 120 V, 60 Hz,
510 VA (according to type)
Protection Class I
Transient overvoltage Overvoltage category II
Fuses
Main Power – external 2 x 2 A sluggish (230 V) or 2 x 4 A sluggish
(120 V) 1/4” x 1 1/4” Spiral type
EMO Main Power Supply (internal) 1x 500 mA sluggish (F1)
1x 125 mA sluggish (F2)
Size 5 mm x 20 mm
Power Supply Probe System 28 V d.c. 5 V d.c.
Power Supply EMO System 24 V a.c.

48
SUSS / PA300 / User Manual / M10-121841-00 / February 2002
Utility Distribution
Concentric Socket INK 1 .. 4 Optional use (DC impulse up to 28 V d.c.)
Concentric Socket Edge 1, 2 Optional use (edge switch)
Airbourne socket J7 Optional use
Pin configuration In the plug
0 V (GND) Contacts 1 and 2, bridged
+ 5 V d.c. Contacts 6 and 7, bridged
+ 28 V d.c. Contacts 3 and 4 and 5, bridged
Affiliated fuses:
+ 5 V d.c. 1 A sluggish (F5)
+ 28 V d.c. 1 A sluggish (F6)
5 mm x 20 mm