SussPA300UserGuide - 第52页

SUSS / PA300 / User Manual / M10-121841-00 / February 2002 51 Indicator Part SESC Criteria Actual Value Notes Fist to wrist access: Height; Wrist; loading tray min. 89 mm min. 127 mm >89 x 127 mm Yes 2 arms to shoulde…

100%1 / 62
50
SUSS / PA300 / User Manual / M10-121841-00 / February 2002
Indicator Part
SESC
Criteria
Actual Value Notes
Full body clearance
standing/ sitting
position
to be supplied by
the customer
see
section
2.9.1
Any posture: upper
body clearance
(shoulder width)
min. 610 mm
see
section
2.9.1
Standing position:
Height;
Width;
min. 1980 mm
min. 690 mm
to be supplied by
the customer
see
section
2.9.1
Sitting Position (on
floor):
Height;
Width;
Working height;
min. 1000 mm
min. 690 mm
min. 280 mm
to be supplied by
the customer
see
section
2.9.1
Squatting (on floor):
Height;
Width;
Working height;
min. 1220 mm
min. 790 mm
min. 460 mm
to be supplied by
the customer
see
section
2.9.1
Kneeling (on floor):
Height;
Width;
Working height;
min.1450 mm
min. 1220 mm
min. 640 mm
to be supplied by
the customer
see
section
2.9.1
Kneeling crawl (on
floor):
N/A
Stooping (on floor):
Height;
Width;
Working height;
min. 1450 mm
min. 1020 mm
min. 640 mm
to be supplied by
the customer
see
section
2.9.1
Supine (lying on back): N/A
Prone (lying on front): N/A
Hand/Arm Clearance
(for use with clean-
room gloves only)
Finger access
(diamater):
2, 3, or 4 finger access:
min. 32 mm
object + 58 mm
i.e.: PH150: 2
finger access,
>32 mm
Yes
Flat hand to wrist
access:
Height, palm thickness;
Width, palm width;
loading tray min. 89 mm
min. 114 mm
>89 x 114 mm
2
Yes
SUSS / PA300 / User Manual / M10-121841-00 / February 2002
51
Indicator Part
SESC
Criteria
Actual Value Notes
Fist to wrist access:
Height;
Wrist;
loading tray min. 89 mm
min. 127 mm
>89 x 127 mm Yes
2 arms to shoulders N/A
2 hands to wrists N/A
1 arm to shoulder N/A
1 arm to elbow N/A
Maintenance and
Service Access
Access openings for
frequent service
intervention
optical and
physical access
openings, if
necessary
N/A
Enclosures and covers supports present removeable Yes
Enclosures and covers handles
not for frequent
service
N/A
Enclosures and covers handles
not for frequent
service
N/A
Replaceable
Components
not to be
replaced by
customer
N/A
Connectors, plugs etc.
should be labelled
demanded if not
valued in SEMI
S2
N/A
Circuit boards mounted
in card cage
configuration should
have mounting and
removal aids
finger access,
handles etc.
Handles from PB
electronic socket
units are not
accessible from
the outer side
Yes
Display Units
control box /
monitor
installation left to
the customer
see
section
2.9.2
Location of operator
primary interface
(standing)
1 or 2 screens
are possible
to be determined
by the customer
optional
Height of monitor (from
floor to middle of
screen)
max. 1470 mm
min. 1320 mm
location to be
determined by
the customer
see
section
2.9.2
Height of monitor (from
floor to top of screen)
max. 1680 mm
according to
configuration, 1
or 2 monitors,
location to be
determined by
the customer
see
section
2.9.2
52
SUSS / PA300 / User Manual / M10-121841-00 / February 2002
Indicator Part
SESC
Criteria
Actual Value Notes
Height of monitors used less
than once an hour (from floor
to top of screen)
max. 1680 mm N/A
Height of monitors used less
than once an day (from floor
to top of screen)
max. 2130 mm N/A
Height of touch screen
monitor
max. 1370 mm
min. 910 mm
N/A
Tilt of touch screen between
41" and 48"
min. 30° upwards N/A
Tilt angle of touch screen <
41"
min. 45° upwards N/A
Location of Operator
Primary Interface (sitting)
computer:
sitting
position
when using a PC
Height of monitor (from floor
to middle of screen)
max. 1190 mm
min. 940 mm
according to
configuration, 1
or 2 monitors,
location to be
determined by
customer
see
section
2.9.2
Height of monitor (from floor
to top of screen)
max. 1400 mm
min. 940 mm
according to
configuration, 1
or 2 monitors,
location to be
determined by
customer
see
section
2.9.2
Tilt angle of screen >55"
min. 15°
downwards
N/A
Height of touch screen (from
floor to active block)
max. 1070 mm
min. 760 mm
N/A
Display Characteristics
Lateral distance from display
to input device
max. 320 mm
location to be
determined by
customer
see
section
2.9.2
Character height (specific to
Chinese, Korean and
Japanese characters)
character height
>= visual
distance (457
mm - 762 mm)
divided by 143
N/A
Character height (except for
Chinese, Korean and
Japanese characters)
character height
>= visual
distance (457
mm - 762 mm)
divided by 215
devices to be
provided by
customer
see
section
2.9.2