HS50_advance_level 1_20200522_221201 (1) - 第514页

S tud ent Guid e HS-50 A dvanced I 0 6/2002 E dition 12 Pr eventive Maintenance 81 Y ou will need the f oll o wing tools an d consum ables for m aintaining the valve plungers: – Lint-free cloth s – ISOF LEX TOP AS NCA 52…

100%1 / 567
06/2002 Edition Student Guide HS-50 Advanced I
12 Preventive Maintenance
80
12.12 Maintenance work on the valve plungers
12
Student Guide HS-50 Advanced I 06/2002 Edition
12 Preventive Maintenance
81
You will need the following tools and consumables for maintaining the valve plungers:
Lint-free cloths
ISOFLEX TOPAS NCA 52 (Item no. 00328369-01, 00330850-01, 50 g tube)
O-ring (Item no. 00345127-01)
New valve plungers, if necessary (Item no. 00330711-02).
a Wipe the valve plungers with a dry lint-free cloth.
s Check the O-ring and the sealing lips of the valve plungers for damage and to ensure that they
are not seated too loosely.
: If the bottom four sealing lips are damaged or are seated too loosely, replace them with new
valve plungers.
d Grease the valve plungers extremely sparingly with a lint-free cloth moistened with ISOFLEX
TOPAS NCA 52.
There must be no visible grease residues.
Make sure that there is no grease in the groove.
12
06/2002 Edition Student Guide HS-50 Advanced I
12 Preventive Maintenance
82