Section 7 ThermoFlex 7-13 Thermo Scientific V erifying/ Adjusting the Controller PID V alues (T une Loop) The controller controls temperature using a Propor tional-Integ ral- Deriv ativ e (PID) alg orithm. Should your chi…

100%1 / 138
Section 7
7-12 ThermoFlex
Thermo Scientific
Chiller will not circulate process uid
Check the reservoir level. Fill, if necessary.
Ensure the reservoir bag lter is not clogged.
Check the application for restrictions in the cooling lines.
Chiller requires >3 PSIG application pressure drop. If a bypass valve
has been installed add some restriction to the bypass line.
The pump motor overloaded. The pump's internal overtemperature
overcurrent device will shut off the pump causing the ow to stop. This
can be caused by low uid, debris in system, operating chiller in a high
ambient temperature condition or excessively conned space. Allow
time for the motor to cool down.
Make sure supply voltage matches the chiller's nameplate rating ±10%.
Inadequate temperature control
Verify the setpoint.
If the chiller is over-cooling, recycle the power.
Make sure the condenser/air lter is free of dust and debris.
Check the uid concentration, see Section 3.
Ensure chiller installation complies with site requirements, see Section 3.
Make sure supply voltage matches chiller nameplate rating ±10%.
For ThermoFlex900 - 5000 global voltage chillers ensure the chiller is
properly congured, see Appendix B.
If the temperature continues to rise, make sure your application's heat
load does not exceed the rated specications.
Check for high thermal gradients (e.g., the application load is being
turned on and off or rapidly changing).
Verify/adjust controller PID values, see next page.
Please contact Thermo Fisher Scientic Sales Service and Customer
Support if you need any additional information, see this manual's inside
cover for contact instructions.
Section 7
ThermoFlex 7-13
Thermo Scientific
Verifying/
Adjusting the
Controller
PID Values
(Tune Loop)
The controller controls temperature using a Proportional-Integral-
Derivative (PID) algorithm. Should your chiller experience temperature
control issues, verifying/adjusting the controller's PID values may correct
the condition.
Note Thermo Fisher recommends that only a qualied technician adjust
the PID values. Incorrect values will hamper chiller performance.
Pro
proportional value
P = % of span (100°C)
Range: 0.0 to 99.9
Factory Preset:
ThermoFlex900-5000
10.0
ThermoFlex7500-10000
20.0
ThermoFlex15000-20000
30.0
ThermoFlex24000
40.0
int
integral value
I = repeats/minute
Range: 0.00 to 9.99
Factory Preset: 0.50
dEr
derivative value
D = minutes
Range: 0.0 to 9.9
Factory Preset: 0.0
Figure 7-1 Verifying/Adjusting PID Values
indicates press and hold.
tunE
Pro
int
dEr
enter
StorE
xx.x C
mode
mode
CAL
mode
xx.x
xx.x
xx.x
xx.x
x.xx
x.x
enter
enter
enter
enter
enter
enter