Section 2 2-20 ThermoFlex Thermo Scientific

100%1 / 138
Section 2
ThermoFlex 2-19
Thermo Scientific
Shipping crate dimensions
(approximate):
62" (157 cm) wide
78" (198 cm) tall
48" (122 cm) deep
Top View
Rear View
5
9
/
16
"
14.2
8
1
/
16
"
20.5
25"
63.5
The applicable options t
within this envelope, see
Section 5.
25
7
/
8
"
65.7
Process Return
1" FNPT Stainless Steel
2
1
/
4
"
5.7
7"
17.8
20
1
/
8
"
51.1
15
1
/
16
"
38.3
5
1
/
16
"
12.9
14
11
/
16
"
37.3
15
1
/
4
"
38.8
24
5
/
8
"
62.8
24
5
/
8
"
62.8
19
13
/
16
"
50.3
6
5
/
16
"
16.0
6
5
/
16
"
16.0
1
3
/
4
"
4.4
Facility water
connections
Supply
Cast Bronze 3/4" FNPT
Return
Stainless 3/4" FNPT
Facility drain 1/4" FNPT
Process drain 1/4" FNPT
Optional 1/4" FNPT
Auto-Rell Port
Process drain valve
Process Discharge
1" FNPT Wrought Iron
DRAIN
Water-cooled only
Power
Inlet
ThermoFlex15000/20000/24000
• Thermo Fisher Scientic
reserves the right to change
specications without notice.
Section 2
2-20 ThermoFlex
Thermo Scientific
ThermoFlex 3-1
Thermo Scientific
Section 3 Installation
Ambient Temperature Range* 10°C to 40°C (50°F to 104°F)
Relative Humidity Range 10% to 80% (non-condensing)
Operating Altitude* Sea Level to 8000 feet (2438 meters)
Overvoltage Category II
Pollution Degree 2
Degree of Protection IP 20
Never place the chiller in a location where excessive heat, moisture,
inadequate ventilation, or corrosive materials are present.
Note Refer to the nameplate information on the rear of the chiller.
Air-cooled chillers retain their full rated capacity at 20°C setpoint in ambient
temperatures up to 25°C (77°F). For ambient temperatures above 25°C please
de-rate the cooling capacity 3% for every 1°C above 25°C (77°F), up to a
maximum ambient temperature of 40°C (104°F). Note that when operating
at a process temperature lower than 20°C the de-rate percentage may increase
due to additional gains from losses to ambient.
Note Depending on the setpoint and ambient temperatures, there may be a
heat gain or loss through the plumbing resulting in a variation from setpoint
temperature at the application inlet. Applications with large temperature
variations between ambient and setpoint temperatures, and/or long plumbing
lengths, may require additional insulation.
ThermoFlex2500 air-cooled chillers have a two-speed fan. Should the chiller's
internal ambient temperature reach 50°C for 30 seconds, or reach 53°C, the
fan speed will switch from slow speed to high speed to maintain internal
temperatures within acceptable limits. When the temperature reaches 44°C or
below for at least 15 minutes the speed will return to low. When in high speed

Note High speed is required for the chiller to achieve its 2500 watt cooling
capacity. At high-end operating conditions the fan can be set to run at high
speed all the time using the controller's Setup Loop, see Section 4.
Site Requirements
*Because of the decrease in air density, maximum temperature for the air entering an air-cooled
ThermoFlex is reduced by 1°C per 1,000 feet above sea level. In addition, cooling capacity is
reduced 1.2% per 1,000 feet above sea level.