ThermoFlex-Manual - 第44页

Section 3 3-6 ThermoFlex Thermo Scientific ThermoFlex900 V oltage ±10% Frequency Phase Branch Circuit Requirements Line Cord Plug 200/208/230 V AC 60 Hz 1Ø 15A - 200/230 V AC 50 Hz 1Ø **16A 1 , 15A 2 , 13A 3 - ThermoFlex1…

100%1 / 138
Section 3
ThermoFlex 3-5
Thermo Scientific
ThermoFlex15000/20000
(Air-cooled)
Voltage ±10% Frequency Phase MCA MOPD Line Cord
Plug
208-230 VAC P 3 Pump 60 Hz 32.2 60 Hard wire
208-230 VAC P 5 Pump 60 Hz 35.8 60 Hard wire
400 VAC P 3 Pump 50 Hz 15.9 30 Hard wire
400 VAC P 5 Pump 50 Hz 18.1 30 Hard wire
ThermoFlex15000/20000
(Water-cooled)
Voltage ±10% Frequency Phase MCA MOPD Line Cord
Plug
208-230 VAC P 3 Pump 60 Hz 28.7 50 Hard wire
208-230 VAC P 5 Pump 60 Hz 32.3 60 Hard wire
400 VAC P 3 Pump 50 Hz 14.5 25 Hard wire
400 VAC P 5 Pump 50 Hz 16.7 30 Hard wire
ThermoFlex24000
(Air-cooled)
Voltage ±10% Frequency Phase MCA MOPD Line Cord
Plug
208-230 VAC P 3 Pump 60 Hz 43.9 70 Hard wire
208-230 VAC P 5 Pump 60 Hz 47.5 80 Hard wire
400 VAC P 3 Pump 50 Hz 20.1 35 Hard wire
400 VAC P 5 Pump 50 Hz 22.3 40 Hard wire
ThermoFlex24000
(Water-cooled)
Voltage ±10% Frequency Phase MCA MOPD Line Cord
Plug
208-230 VAC P 3 Pump 60 Hz 37.1 70 Hard wire
208-230 VAC P 5 Pump 60 Hz 40.7 70 Hard wire
400 VAC P 3 Pump 50 Hz 18.8 35 Hard wire
400 VAC P 5 Pump 50 Hz 21.0 35 Hard wire
For installation information on variable voltage chillers refer to Appendix B.
Refer to the nameplate label located on the rear of the chiller for specic
electrical requirements.
* United States and Japan only. All other plugs are country specific.
ThermoFlex900 Voltage ±10% Frequency Phase Branch Circuit
Requirements
Line Cord
Plug
115 VAC 60 Hz 15A 5-15P*
100 VAC 50/60 Hz 15A 5-15P*
ThermoFlex1400 Voltage ±10% Frequency Phase Branch Circuit
Requirements
Line Cord
Plug
115 VAC 60 Hz 20A -
100 VAC 50/60 Hz 20A -
Electrical Service Requirements (Variable voltage chillers):
Section 3
3-6 ThermoFlex
Thermo Scientific
ThermoFlex900 Voltage ±10% Frequency Phase Branch Circuit
Requirements
Line Cord
Plug
200/208/230 VAC 60 Hz 15A -
200/230 VAC 50 Hz **16A
1
, 15A
2
, 13A
3
-
ThermoFlex1400 Voltage ±10% Frequency Phase Branch Circuit
Requirements
Line Cord
Plug
200/208/230 VAC 60 Hz 15A -
200/230 VAC 50 Hz **16A
1
, 15A
2
, 13A
3
-
ThermoFlex2500 Voltage ±10% Frequency Phase Branch Circuit
Requirements
Line Cord
Plug
200 VAC T 1 Pump 60 Hz 15A -
208-230 VAC T 1 Pump 60 Hz 20A -
230 VAC 50 Hz *16A
1
, 15A
2
, 13A
3
-
ThermoFlex3500/5000 Voltage ±10% Frequency Phase Branch Circuit
Requirements
Line Cord
Plug
200/208-230 VAC P 1 P 3 Pump 50/60 Hz 15A -
200/208-230 VAC T 1 P 3 Pump 50/60 Hz 20A -
200/208-230 VAC P 4 Pump 50/60 Hz 30A Hard wired
Electrical Service Requirements (Global Voltage chillers):
ThermoFlex7500/10000
(Air-cooled)
Voltage ±10% Frequency Phase MCA MOPD Line Cord
Plug
400 VAC P 2 Pump 50 Hz 8.8 15 Hard wire
400 VAC P 3 Pump 50 Hz 10.1 20 Hard wire
400 VAC P 5 Pump 50 Hz 12.3 20 Hard wire
400 VAC T 5 Pump 50 Hz 9.1 15 Hard wire
460 VAC P 2 Pump 60 Hz 8.8 15 Hard wire
460 VAC P 3 Pump 60 Hz 10.1 20 Hard wire
460 VAC P 5 Pump 60 Hz 12.3 20 Hard wire
460 VAC T 5 Pump 60 Hz 9.1 15 Hard wire
ThermoFlex7500/10000
(Water-cooled)
Voltage ±10% Frequency Phase MCA MOPD Line Cord
Plug
400 VAC P 2 Pump 50 Hz 8.4 15 Hard wire
400 VAC P 3 Pump 50 Hz 9.7 20 Hard wire
400 VAC P 5 Pump 50 Hz 11.9 20 Hard wire
400 VAC T 5 Pump 50 Hz 8.8 15 Hard wire
460 VAC P 2 Pump 60 Hz 8.4 15 Hard wire
460 VAC P 3 Pump 60 Hz 9.7 20 Hard wire
460 VAC P 5 Pump 60 Hz 11.9 20 Hard wire
460 VAC T 5 Pump 60 Hz 8.8 15 Hard wire
Continued on next page.
Section 3
ThermoFlex 3-7
Thermo Scientific
** Chillers selected for 230 VAC operation have a range of -10% to +7%. Refer to Appendix A for
country specific ratings.
For installation information on global voltage chillers refer to Appendix B.
 
electrical requirements.
ThermoFlex15000/20000
(Air-cooled)
Voltage ±10% Frequency Phase MCA MOPD Line Cord
Plug
400 VAC P 3 Pump 50 Hz 16.2 30 Hard wire
400 VAC P 5 Pump 50 Hz 18.4 30 Hard wire
460 VAC P 3 Pump 60 Hz 16.2 30 Hard wire
460 VAC P 5 Pump 60 Hz 18.4 30 Hard wire
ThermoFlex15000/20000
(Water-cooled)
Voltage ±10% Frequency Phase MCA MOPD Line Cord
Plug
400 VAC P 3 Pump 50 Hz 14.5 25 Hard wire
400 VAC P 5 Pump 50 Hz 16.7 30 Hard wire
460 VAC P 3 Pump 60 Hz 14.5 25 Hard wire
460 VAC P 5 Pump 60 Hz 16.7 30 Hard wire
ThermoFlex24000
(Air-cooled)
Voltage ±10% Frequency Phase MCA MOPD Line Cord
Plug
400 VAC P 3 Pump 50 Hz 20.1 35 Hard wire
400 VAC P 5 Pump 50 Hz 22.3 40 Hard wire
460 VAC P 3 Pump 60 Hz 20.1 35 Hard wire
460 VAC P 5 Pump 60 Hz 22.3 40 Hard wire
ThermoFlex24000
(Water-cooled)
Voltage ±10% Frequency Phase MCA MOPD Line Cord
Plug
400 VAC P 3 Pump 50 Hz 18.8 35 Hard wire
400 VAC P 5 Pump 50 Hz 21.0 35 Hard wire
460 VAC P 3 Pump 60 Hz 18.8 35 Hard wire
460 VAC P 5 Pump 60 Hz 21.0 35 Hard wire
MCA = Minimum Current Ampacity
MOPD = Maximum Overcurrent Protective Device
Values reflect those on the nameplate located on the rear of the chiller .