Section 3 ThermoFlex 3-9 Thermo Scientific Plumbing Requirements Ensure that all shipping plugs are remov ed before installation. Never connect the process uid lines to y our facility water supply or any pressuriz ed liq…

Section 3
3-8 ThermoFlex
Thermo Scientific
Hard Wire Installation
For personal safety and equipment reliability, only a qualied
technician should perform the following procedure.
Note The technician is responsible for installing circuit protection for
incoming power. Before wiring consult the nameplate on the rear of the
chiller. Ensure installation is in accordance with the National Electrical Code
and any other applicable country and local codes.
For ThermoFlex900 through 10000 chillers
Remove the six screws securing the electrical box cover to the chiller.•
Remove the double knock out ( •
7
/
8
" and 1
3
/
32
").
Insert the cable through the hole.•
Refertothelabelintheelectricalboxtocongureyourchiller,seeFigure3-3.•
Secure the cable's ground wire to the ground stud.•
Reinstall the cover.•
For ThermoFlex15000, 20000 and 24000 chillers
Removethevescrewssecuringtheelectricalpaneltothechiller.•
Refertothelabelintheelectricalboxtocongureyourchiller,seeFigure3-3.•
Secure the cable's ground wire to the ground stud.•
Reinstall the panel..•
(2)
Terminal Block
Ground Stud
Knock Out
(2)
(2)
Figure 3-2 Electrical Box
Figure 3-3 Sample Label
WARNING

Section 3
ThermoFlex 3-9
Thermo Scientific
Plumbing
Requirements
Ensure that all shipping plugs are removed before installation.
Never connect the process uid lines to your facility water supply or
any pressurized liquid source.
To prevent damage to the chiller's plate exchanger, centrifugal pumps
require a 4.0 gpm (15.1 lpm) minimum ow rate.
P 1 and P 2 pumps are capable of producing 185 psig. Ensure your
plumbing is rated to withstand this pressure at your operating
temperature. An external pressure relief valve is available, see Section 5.
Note Ensure your plumbing installation develops a back pressure to the
labeled
(PROCESS OUTLET) and ().
Process Fluid Connections (FNPT)
Outlet
ThermoFlex900 - 10000 P 1 P 2 T 0 T 1 1/2" cast bronze
ThermoFlex3500 - 5000 P 3 P 4 3/4" cast bronze
ThermoFlex7500 - 24000 P 3 P 5 T 5 1" wrought copper
Supplied Adapters
P 1 P 2 T 0 T 1 1/2" x 3/8'' Polyethylene and 1/2" x 1/2" Nylon
P 3 P 4 3/4 MPT x 1/2 barb PVC
P 3 P 5 T 5 1" MPT x 1" Barb PVC and 1" MPT x 3/4" Barb PVC
CAUTION
CAUTION
CAUTION
1/4" Female NPT Riton Reservoir
Drain Plug
See Section 2 for the specific
locations on your chiller.
Figure 3-4 Typical Plumbing Connections (1 of 2)
DRAIN
Stainless steel outlet connection for chillers
with P 1/P 2 pumps and a flow transducer

Section 3
3-10 ThermoFlex
Thermo Scientific
Water-cooled Chillers
For water-cooled chillers the facility water plumbing connections are also located
on the rear and are labeled
OUTLET. The connections are ½" Female NPT for ThermoFlex900 -
5000, ¾" Female NPT for ThermoFlex7500 - 24000. Both connections
for ThermoFlex900 to 10000 are cast bronze. The supply connections for
ThermoFlex15000 to 24000 are cast bronze, the return connections are stainless
steel.
Connect the
to your facility water supply. Connect
the
to your facility water return or drain. Ensure
®
tape is used, ensure the tape does not overhang the
Connect the
PROCESS OUTLET to
your application. Ensure all connections are
secure and that the proper sealant/lubricant
®
tape
is used, ensure the tape does not overhang the
and the instrument being cooled as short as
possible. Ensure tubing is straight and without
make them at the inlet and outlet of your
application, not at the ThermoFlex.
Figure 3-4 Typical Plumbing
Connections (2 of 2)
Process Fluid Flow
Process Fluid Flow
Application
Figure 3-5 Typical Plumbing Connections, Water-cooled Chillers
See Section 2 for the specific
locations on your chiller.