ThermoFlex 3-13
Thermo Scientific
To prevent freezing/glazing of the plate exchanger,
ThermoFlex7500 through 24000 chillers require the use of 50/50
EG/water or 50/50 PG/water below 10°C process temperature.
Uninhibited Ethylene Glycol/Water
Ethylene glycol is used to depress the freezing point of water and should
only be used at temperatures where freeze point suppression is required.
Ethylene glycol does not improve heat transfer and is not recommended
for use as a biocide. Because glycols lower the surface tension of
water and do not evaporate as readily as water, they may cause visible
weepage past the pump seals. If weepage cannot be tolerated, seal-less,
magnetically driven pumps should be used where available.
Uninhibited simply means that the glycol does not contain any additives
to prevent corrosion. While uninhibited ethylene glycol is acceptable for
use, the pH level must be closely monitored and the uid may need to
be replaced more often. All glycols produce acids in the presence of air
and the uid should be changed if the pH falls below 8. Note that litmus
paper will not work to test the pH of ethylene glycol/water.
Inhibited Ethylene Glycol/Water
Inhibited glycol can help protect the wetted metals within the cooling
circuit from corrosion caused by poor water quality, ethylene glycol
oxidation (low pH) and mixed metals (electrolysis). The inhibitor works
by either leaving a barrier coating on metal surfaces to buffer them from
the corrosive uid or by creating an oxidized layer that protects the
underlying metal (passivating).
Inhibited automotive glycols are never acceptable. They use either
silicates or Organic Acid Technology (OAT) as the inhibitor and these
components are not compatible with the polymers used in recirculating
chillers including the pump seals and internal hoses.
Inhibitors may also accelerate pump seal wear and seal-less, magnetically
driven pumps should be used where available.
Uninhibited Propylene Glycol/Water
Propylene glycol does not transfer heat as well as ethylene glycol, but can
be used when freeze point suppression is required as well as lower toxicity.
Propylene glycol does not function as a biocide and the pH needs to be
maintained the same as with ethylene glycol as it also produces acid when
oxidized.
Inhibited Propylene Glycol/Water
Inhibited propylene glycol has the same properties as uninhibited
propylene glycol and the same concerns as inhibited ethylene glycol.
CAUTION

Section 3
3-14 ThermoFlex
Thermo Scientific
Process Fluid Permissible (PPM) Desirable (PPM)
Microbiologicals
(algae, bacteria, fungi) 0 0
Inorganic Chemicals
Calcium <25 <0.6
Chloride <25 <10
Copper <1.3 <1.0
0.020 ppm if fluid in contact with aluminum
Iron <0.3 <0.1
Lead <0.015 0
Magnesium <12 <0.1
Manganese <0.05 <0.03
Nitrates\Nitrites <10 as N 0
Potassium <20 <0.3
Silicate <25 <1.0
Sodium <20 <0.3
Sulfate <25 <1
Hardness <17 <0.05
Total Dissolved Solids <50 <10
Other Parameters
pH 6.5-8.5 7-8
Resistivity 0.01* 0.05-0.1*
* MΩ-cm (compensated to 25°C)
Unfavorably high total ionized solids (TIS) can accelerate the rate of galvanic
corrosion. These contaminants can function as electrolytes which increase
the potential for galvanic cell corrosion and lead to localized corrosion such
as pitting. Eventually, the pitting will become so extensive that refrigerant will
leak into the water reservoir.
As an example, raw water in the United States averages 171 ppm (of NaCl).
The recommended level for use in a water system is between 0.5 to 5.0 ppm
(of NaCl).
Recommendation: Initially ll the tank with distilled or deionized water
within a range of 1-3 M
Ω-cm. (It is acceptable to have the uid drop to the
Process Water
Quality and
Standards
When using the ThermoFlex chiller to circulate through aluminum, a
compatible corrosion inhibitor should be utilized to prevent galvanic
corrosion.
Fluid viscosity should be 50 cSt or less at the lowest temperature used.
Visible pump weepage may occur when compatible glycols, oils or other
additives are used. Pump weepage is considered as a normal operating
condition of mechanical seal pumps.
Additional Fluid
Information

Section 3
ThermoFlex 3-15
Thermo Scientific
Facility Water
Quality and
Standards
(water-cooled
chillers)
Facility Water Permissible (PPM) Desirable (PPM)
Microbiologicals
(algae, bacteria, fungi) 0 0
Inorganic Chemicals
Calcium <40 <0.6
Chloride <250 <25
Copper <1.3 <1.0
0.020 ppm if fluid in contact with aluminum
Iron <0.3 <0.1
Lead <0.015 0
Magnesium <12 <0.1
Manganese <0.05 <0.03
Nitrates\Nitrites <10 as N 0
Potassium <20 <0.3
Silicate <25 <1.0
Sodium <20 <0.3
Sulfate <250 <50
Hardness <17 <0.05
Total Dissolved Solids <50 <10
Note A corrosion inhibitor is recommended if mixed metals are in the
facility water loop.
other levels over-time.) Do not use untreated tap water as the total ionized
solids level may be too high. This will reduce the electrolytic potential of
the water and prevent or reduce the galvanic corrosion observed.