Section 4 4-10 ThermoFlex Thermo Scientific Figure 4-7 Setup Loop (Optional Features) Continued on next page. xx xxx xx xx FLt indC xx di t SEc HZ ALr di t FiLL LoFLO HiFLO HiFLO enter enter enter enter enter enter enter …

Section 4
ThermoFlex 4-9
Thermo Scientific
Figure 4-6 Setup Loop (All Chillers)
• StArt is used to enable/disable the auto restart
function. When enabled the chiller will automatically
restart after a power failure or power interruption
condition.
Range:
on or oFF Default: oFF
NOTE Consider any possible risks before enabling
this mode of operation.
• CArE is used to set the preventive care cleaning
frequency reminder for the chiller's air and uid
lters, in hours. The time selected is based on your
operating environment, see Section 6.
Range:
off Default: L1
L1 (1000 hours)
L2 (2000 hours)
L3 (3000 hours)
Off disables the reminder. Exceeding this limit
ashes FLtrS, see Section 6.
Note If your chiller is equipped with any of the
Optional Features refer to the next page.
When the display indicates StorE press
to save all changes or press to not save
all changes. The display will return either the process
uid temperature or, if the chiller was off when you
entered the loop, a blank screen.
Continued from previous page
xx
on
oFF
Lx
CArE
AUto
StorE
StArt
CArE
enter
enter
enter
enter
Save all changes
Do not save all changes

Section 4
4-10 ThermoFlex
Thermo Scientific
Figure 4-7 Setup Loop (Optional Features)
Continued on next page.
xx
xxx
xx
xx
FLt
indC
xx
di t
SEc
HZ
ALr
di t
FiLL
LoFLO
HiFLO
HiFLO
enter
enter
enter
enter
enter
enter
enter
enter
enter
OPT
HZ
CArE
• HZ is used to identify the incoming frequency for
chiller's with P 3 - P 5 pumps and the capability to run
on either 50 Hz or 60 Hz. The selected frequency
automatically adjusts the rmware's xed high pressure
default setting.
Range:
50 Hz or 60 Hz Default: 60 Hz
• OPt is used to congure the analog in/out mode of
operation. See Appendix C.
• FiLL is used to set the time limit the auto rell has for
lling the reservoir to the normal operating level.
Range: 0 to 900 seconds
Default: 45 seconds ThermoFlex900 - 5000
80 seconds ThermoFlex7500 - 24000
Exceeding the time limit ashes rEFiL and the auto rell
will shut off. The chiller's reaction depends on the alarm
ALr setting, Flt is shut down, indC is continue to run.
Note Setting the time limit to 0 disables the auto rell
option.
See Section 5 for additional information.
• di t is used to set the preventive care cleaning frequency
reminder for the chiller's DI lter cartridge.
Range:
0 to 9999 hours Default: 448 hours
Exceeding the limit ashes di, see Section 6.
• HiFLO is used to set the high ow alarm limit.
T 0 T 1 Range: 0.0 to 10.5 GPM Default: 0.0 GPM
T 5 Pump Range: 0.0 to 15.0 GPM Default: 0.0 GPM
P 1 Pump Range: 0.0 to 10.5 GPM Default: 0.0 GPM
P 2 Pump Range: 0.0 to 10.5 GPM Default: 0.0 GPM
P 3 Pump Range: 0.0 to 30.0 GPM Default: 0.0 GPM
P 4 Pump Range: 0.0 to 30.0 GPM Default: 0.0 GPM
P 5 Pump Range: 0.0 to 30.0 GPM Default: 0.0 GPM
Exceeding a high limit ashes HiFLO and, if enabled,
sounds the alarm. The chiller's reaction depends on the
alarm (ALr) setting.
Note This feature is not enabled until the value is
changed to something other than 0.0.
Shaded displays only appear
on chillers equipped with that option.

Section 4
ThermoFlex 4-11
Thermo Scientific
yES
no
drAin
StorE
drAin
enter
enter
SEr
xxxx
LoFLO
LoFLO
enter
enter
Figure 4-7 Setup Loop (Optional Features)
• LoFLO is used to set the low ow alarm limit.
T 0 Pump Range: 0.0 to 10.5 GPM Default: 0.0 GPM
T 1 Pump Range: 0.0 to 10.5 GPM Default: 0.0 GPM
T 5 Pump Range: 0.0 to 15.0 GPM Default: 0.0 GPM
P 1 Pump Range: 0.0 to 10.5 GPM Default: 0.0 GPM
P 2 Pump Range: 0.0 to 10.5 GPM Default: 0.0 GPM
P 3 Pump Range: 0.0 to 30.0 GPM Default: 0.0 GPM
P 4 Pump Range: 0.0 to 30.0 GPM Default: 0.0 GPM
P 5 Pump Range: 0.0 to 30.0 GPM Default: 0.0 GPM
Going below the low limit ashes LoFLO and, if
enabled, sounds the alarm. The chiller's reaction depends
on the alarm (ALr) setting.
This feature is not enabled until the value is changed to
something other than 0.0. If the feature is not enabled
and the ow rate drops below the ow rate listed below
the chiller will continue to run and the controller will ash
between FLo and LoFLo.
P
1, T 0, T 1 and T 5 Pump 0.3 GPM
P
2 Pump 1.0 GPM
P
3, P 4 and P 5 Pump 4.0 GPM
• SEr is used to congure the serial communications
mode of operation. See Appendix D.
Note Keypad operation is still available with serial com-
munications enabled.
• drAin is used to open and close the chiller's anti
drainback valve for draining, see Section 5.
Range:
yes or no Default: no
Note The chiller must be off to drain the valve. The valve
automatically closes when you exit the drAin display.
When the display indicates StorE press
to save all changes or press to not save all
changes. The display will return either the process uid
temperature or, if the chiller was off when you entered
the loop, a blank screen.
Continued from previous page
Save all changes
Do not save all changes