Command Groups T able 2-41: T rigger Commands (cont.) Command Description TRIGger:A:BUS:B<x>:MIL1553B:ST A T us:BIT : DBCA When the MIL-STD-1553 bus trigger condition is set to STATus , this command speci fi es the …

Command Groups
Table 2-41: Trigger Commands (cont.)
Command
Description
TRIGger:A:BUS:B<x>:MIL1553B :CO MMAND:
ADDRess:HIVALue
When the M IL-STD-1553 bus trigger condition is set to COMMAN D, and the
qualifier is set to
INrange or OUTrange, this command specifies the upper
limit of the range for the remote terminal address field.
TRIGger:A:BUS:B<x>:MIL1553B :CO MMAND:
ADDRess:QUALifier
When the MIL-STD-1553 bus trigger condition is set to COMMAND,this
command specifies the qualifier to be used with the remote terminal address
field.
TRIGger:A:BUS:B<x>:MIL1553B :CO MMAND:
ADDRess:VALue
When the MIL-STD-1553 bus trigger condition is set to COMMAND,and
the qualifier is set to
LESSthan, MOREthan, EQual, UNEQual,
LESSEQual
or MOREEQual, this command specifies the value of the 5–bit
remote terminal address to be used in the trigger.
TRIGger:A:BUS:B<x>:MIL1553B :CO MMAND:
COUNt
When the MIL-STD-1553 bus trigger condition is set to COMMAND,this
command specifies the bit pattern for the 5–bit Word C ount/Mode Code
sub-address fi eld that is to be used in the trigger.
TRIGger:A:BUS:B<x>:MIL1553B :CO MMAND:
PARity
When the MIL-STD-1553 bus trigger condition is set to COMMAND,this
command specifies the Command word parity that is to be used in the trigger.
TRIGger:A:BUS:B<x>:MIL1553B :CO MMAND:
SUBADdress
When the MIL-STD-1553 bus trigger condition is set to COMMAND,this
command specifies the 5 bit s ub-address that is to be used in the trigger.
TRIGger:A:BUS:B<x>:MIL1553B :CO MMAND:
TRBit
When the MIL-STD-1553 bus trigger condition is set to COMMAND,this
command specifies that the transmit/receive bit (bit 9) is to be used in the
trigger.
TRIGger:A:BUS:B<x>:MIL1553B :CONDition This command specifies a word type or condition within a MIL-STD-1553 bus
word to trigger on.
TRIGger:A:BUS:B<x>:MIL1553B :DATa:PARity When the MIL-STD-1553 bus trigger condition is set to DATa, this c ommand
specifies the data parity bit to be used in the trigger.
TRIGger:A:BUS:B<x>:MIL1553B :DATa:VALue When the MIL-STD-1553 bus trigger condition is set to DATa, this command
specifies the data binary pattern to be used in the trigger.
TRIGger:A:BUS:B<x>:MIL1553B :ERRTYPE When the MIL-STD-1553 bus trigger condition is set to ERRor, this command
specifies the signaling error type to be used in the trigger: Parity, Sync,
Manchester or Data.
TRIGger:A:BUS:B<x>:MIL1553B :STATus:
ADDRess:HIVALue
When the MIL-STD-1553 bus trigger condition is set to STATus, and the
qualifier is set to
INrange or OUTrange, this command specifies the upper
limit for the 5 bit remote terminal address fi eld of the Status word.
TRIGger:A:BUS:B<x>:MIL1553B :STATus:
ADDRess:QUALifier
When the MIL-STD-1553 bus trigger condition is set to STATus, this command
specifies the qualifier to be used with the address field.
TRIGger:A:BUS:B<x>:MIL1553B :STATus:
ADDRess:VALue
When the MIL-S TD-1553 bus trigger condition is set to STATus, and
the qualifier is set to
LESSthan, MOREthan, EQual, UNEQual,
LESSEQual
or MOREEQual, this command specifies the value of the 5–bit
remote terminal address to be used in the trigger.
TRIGger:A:BUS:B<x>:MIL1553B :STATus:BIT:BCR When the MIL-STD-1553 bus trigger condition is set to STATus, this command
specifies the status word broadcast command received (BCR) bit value (bit
15) to be used in the trigger.
TRIGger:A:BUS:B<x>:MIL1553B :STATus:BIT:
BUSY
When the MIL-STD-1553 bus trigger condition is set to STATus, this command
specifies the status word busy bit value (bit 16) to be used in the trigger.
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-85

Command Groups
Table 2-41: Trigger Commands (cont.)
Command
Description
TRIGger:A:BUS:B<x>:MIL1553B:STATus:BIT:
DBCA
When the MIL-STD-1553 bus trigger condition is set to STATus, this command
specifies the status word dynamic bus contro l acceptance (DBCA) bit value (bit
18) to be used in the trigger.
TRIGger:A:BUS:B<x>:MIL1553B:STATus:BIT:
INSTR
When the MIL-STD-1553 bus trigger condition is set to STATus,this
command specifies the status word instrumentation bit value (bit 10) to be
used in the trigger.
TRIGger:A:BUS:B<x>:MIL1553B:STATus:BIT:ME When the MIL-STD-1553 bus trigger condition is set to STATus,this
command specifies the status word message error bit value ( bit 9) to be used
in the trigger.
TRIGger:A:BUS:B<x>:MIL1553B:STATus:BIT:SRQ When the MIL-STD-1553 bus trigger condition is set to STATus, this command
specifies the status w ord service request (SRQ) bit value (bit 11) to be used
in the trigger.
TRIGger:A:BUS:B<x>:MIL1553B:STATus:BIT:
SUBSF
When the MIL-STD-1553 bus trigger condition is set to STATus,this
command specifies the status word subsystem flag bit value (bit 17) to be
used in the trigger.
TRIGger:A:BUS:B<x>:MIL1553B:STATus:BIT:TF When the M IL-STD-1553 bus trigger condition is set to STATus, this command
specifies the status word terminal flag bit value (bit 19) to be used in the trigger.
TRIGger:A:BUS:B<x>:MIL1553B:STATus:PARity When the MIL-STD-1553 bus trigger condition is set to STATus, this command
specifies the status parity bit value to be used in the trigger.
TRIGger:A:BUS:B<x>:MIL1553B:TIMe:LESSLimit When the MIL-STD-1553 bus trigger condition is set to TIMe, this command
specifies either the minimum remote terminal response time (RT) limit for
the amount of time the terminal has to transmit, or it specifies the minimum
inter-message gap (IMG).
TRIGger:A:BUS:B<x>:MIL1553B:TIMe:MORELimit When the MIL-STD-1553 bus trigger condition is set to TI Me, this command
specifies either the maximum remote terminal response time (RT) limit for
the amount of time the terminal has to transmit, or it specifies the maximum
inter-message gap (IMG).
TRIGger:A:BUS:B<x>:MIL1553B:TIMe:QUALifier When the MIL-STD-1553 bus trigger condition is set to TIMe, this command
specifies the trigger data time qualifier.
TRIGger:A:BUS:B<x>:PARallel:VALue This command specifies the binary data string to b e u sed for a Parallel trigger
TRIGger:A:BUS:B<x>:RS232C:CONDition This command specifies the condition for an RS-232C trigger
TRIGger:A:BUS:B<x>:RS232C:RX:DATa:SIZe This command specifies the length of the data string in Bytes for an RX
RS-232 Trigger
TRIGger:A:BUS:B<x>:RS232C:RX:DATa:VALue This command specifies the binary data string for an RX RS-232 trigger
TRIGger:A:BUS:B<x>:RS232C:TX:DATa:SIZe This command specifies the length of the data string for a TX RS-232 trigger
TRIGger:A:BUS:B<x>:RS232C:TX:DATa:VALue This command specifies the binary data string for an RS-232 trigger if the
trigger condition involves TX
TRIGger:A:BUS:B<x>:SPI:C ONDition This command specifies the trigger condition for SPI triggering
TRIGger:A:BUS :B<x>:SPI:D ATa{:IN|:MISO}:
VALue
This command specifies the binary data string to be used for S PI trigger
TRIGger:A:BUS :B<x>:SPI:DATa{:OUT|:MOSI}:
VALue
This command specifies the binary data string used for the SPI trigger
2-86 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual

Command Groups
Table 2-41: Trigger Commands (cont.)
Command
Description
TRIGger:A:BUS:B <x>:SPI:DATa:SIZe This command specifies the length of the data s tring in bytes to be used for
SPI trigger
TRIGger:A:BUS:B <x >:USB:ADDRess:HIVALue This command specifies the high limit for the USB trigger address
TRIGger:A:BUS:B<x>:USB :ADDRess:VALue This command specifies the value for the USB trigger address
TRIGger:A:BUS:B <x>:USB:CONDition This command specifies the USB trigger condition
TRIGger:A:BUS:B<x>:USB :DATa:HIVALue This command specifi es the high limit for the USB trigger data
TRIGger:A:BUS:B <x>:USB:DATa:OFFSet This command specifies the data offset for the U SB trigger data
TRIGger:A:BUS:B <x>:USB:DATa:SIZe This command specifies the number of data bytes for the USB trigger
TRIGger:A:BUS:B <x>:USB:DATa:TYPe This command specifies the data type for the USB trigger
TRIGger:A:BUS:B <x>:USB:DATa:VALue This command specifies the data value for the USB trigger
TRIGger:A:BUS:B<x>:USB :ENDPoint:VALue This command specifies the endpoint value for the USB trigger
TRIGger:A:BUS:B <x>:USB:ERRTYPE This command specifies the error type for the USB trigger
TRIGger:A:BUS:B<x>:USB :HANDSHAKEType This command specifies the handshake type for the USB trigger
TRIGger:A:BUS:B <x >:USB:QUALifier This command specifies the qualifier for USB trigger
TRIGger:A:BUS:B <x >:USB:SOFFRAMENUMber This command specifies the SOF number for the USB trigger
TRIGger:A:BUS:B<x>:USB :SPECIALType This command specifies the special packet type for the USB trigger
TRIGger:A:BUS:B<x>:USB :SPLit:ET:VALue When triggering on a high-speed USB split transaction, this command specifies
the split transaction endpoint type value to trigger on.
TRIGger:A:BUS:B<x>:USB :SPLit:HUB:VALue When triggering on a high-speed USB split transaction, this command specifies
the split transaction hub address value to trigger on.
TRIGger:A:BUS:B<x>:USB :SPLit:PORT:VALue When triggering on a high-speed USB split transaction, this command specifies
the split transaction port address value to trigger on.
TRIGger:A:BUS:B<x>:USB :SPLit:SC:VALue When triggering on a high-speed USB split transaction, this command specifies
whether to trigger on the start or complete phase of the split transaction, based
on the Start/Complete bit field value.
TRIGger:A:BUS:B<x>:USB :SPLit:SE:VALue When triggering on a high-speed USB split transaction, this command specifies
the split transaction start/end bit value to trigger on.
TRIGger:A:BUS:B <x>:USB:TOKENType This command speci fies the token type for the USB trigger
TRIGger:A:BUS:S OUrce This command specifies the source for a bus trigger
TRIGger:A:EDG E? Returns the source, coupling and source for the A edge trigger
TRIGger:A:EDGE:COUPling This command specifies the type of coupling for the A edge trigger
TRIGger:A:EDGE:SLOpe This command specifies the slope for the A edge trigger: rising, falling or either.
TRIGger:A:EDGE:SOUrce This command specifies the source for the A edge trigger
TRIGger:A:HOLDoff? Returns the A trigger holdoff parameters
TRIGger:A:HOLDoff:TIMe This command specifies the A trigger holdoff time
TRIGger:A:LEVel:AUXin Sets the threshold v oltage level for an Edge, Pulse Width, Runt or Rise/Fall
(aka Transition, aka Slew Rate) trigger to use when triggering on the Aux
Input connector signal.
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-87