Commands Listed in Alphabetical Order
Arguments
1, ON stops acqu
isitions when the specified event occurs.
0, OFF turns off this action.
Examples
ACTONEVENT:ACTION:STOPACQ:STATE ON specifies to stop acquisitions when
the specified event occurs.
ACTONEVent:ACTION:VISUAL:STATE
Displays a visual notification when a specified event occurs, or queries the state of
the “display a visual notification” action. The default state is 0 (off).
To specify an event, use the command ACTONEVent:EVENTTYPe.
Group
Act on Event
Syntax
ACTONEVent:ACTION:VISUAL:STATE <0|1|OFF|ON>
ACTONEVent:ACTION:VISUAL:STATE?
Arguments
1, ON displays a visual notification when the specified event occurs.
0, OFF turns off this action.
Examples
ACTONEVENT:ACTION:VISUAL:STATE ON specifies to display a visual
notification when the specified event occurs.
ACTONEVent:EVENTTYPe
Specifies (or queries) which event to act on (TRIGger, ACQCOMPLete, or
NONe) when using an Act on Event command. The default is
NONe.
To specify the action to take, use the
ACTONEVent:ACTION commands.
Group
Act on Event
Syntax
ACTONEVent:EVENTTYPe <NONe|TRIGger|ACQCOMPLete>
ACTONEVent:EVENTTYPe?
Arguments
NONe no event (this is the default).
TRIGger specifies to act when a trigger occurs.
2-122 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual
Commands Listed in Alphabetical Order
ACQCOMPLete sp
ecifies to act when acquisition completes.
Examples
ACTONEVENT:EVENTTYPE ACQCOMPLETE specifies to take action when
acquisition
completes.
ACTONEVent:NUMACQs
Sets (or queries) the number of acquisitions to complete for the event type
ACQCOMPLete. The default is 1 acquisition.
To set the event type to
ACQCOMPLete,usetheACTONEVent:EVENTTYPe
command.
Group
Act on Event
Syntax
ACTONEVent:NUMACQs <NR1>
ACT
ONEVent:NUMACQs?
Arguments
Integer which can be set as follows:
Table 2-50:
From To Resolution
1 100 1
100 1000 10
1000 10,000 100
10,000 1,000,000 1000
Examples
ACTONEVENT:NUMACQS 350 specifies to take the specified action when 350
acquisitions have been completed.
ACTONEVent:REPEATCount
Sets or returns the number of events to run.
To specify an event, use the command ACTONEVent:EVENTTYPe.
Group
Act on Event
Syntax
ACTONEVent:REPEATCount <NR1>
ACTONEVent:REPEATCount?
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-123
Commands Listed in Alphabetical Order
Arguments
NR1 is an intege
r that specifies the number of events to run. The default
is 1 event and the maximum is 1000000 events. A repeat count greater
than 1000000 specifies to run forever. In this case, the action on event can be
terminated by setting the state for all actions to OFF. When the repeat count is set
to infinite, the query returns 9.9E+37 (IEEE positive infinity).
AFG:AMPLitude
Sets (or qu
eries) the AFG amplitude in volts, peak to peak. The minimum and
maximum allowable settings are dependent upon the output load impedance setting
(AFG:OUTPut:LOAd:IMPEDance) and the selected function (AFG:FUNCtion).
NOTE. The AFG:OUTPut:LOAd:IMPEDance command must be used before
using any of the amplitude/offset/highLevel/lowLevel/Preset commands, so that
saved setups are recalled with the proper load factor applied.
The following values can be set for AFG:AMPLitud e:
AFG AMPLitude Setting Value
Load I
mpedance
FIFTY Ohm
Load I
mpedance
HIGHZ
Function
Min Max Resolution Min Max Resolution
SINE
SQUare
PULS
e
RAMP
NOISe
DC
1
HAVERSINe
CARDIac
AR
Bitrary
10.0
mV
2.5V 500u
V
20.0
mV
5.0V 1mV
SINC
10.0mV 1.5V 500uV 20.0mV 3.0V 1mV
LORENtz
1
0.0mV
1
.2V
5
00uV
2
0.0mV
2
.4V
1
mV
G
AUSsian
ERISe
EDECAy
10.0mV 1.25V 500uV 20.0mV 2.5V 1mV
1
The DC function does not use the amplitude setting. The DC level is set using :AFG:OFFSet. Ampltude settings that are set when the :AFG:FUNCtion is DC
are remembered and applied when the function is changed to other than DC.