Commands Listed in A lphabetical Order Arguments NR1 is an intege r that specifies the number of events to run. The default is 1 event and the maximum is 1000000 events. A repeat count greater than 1000000 s pecifies to ru…

Commands Listed in Alphabetical Order
ACQCOMPLete sp
ecifies to act when acquisition completes.
Examples
ACTONEVENT:EVENTTYPE ACQCOMPLETE specifies to take action when
acquisition
completes.
ACTONEVent:NUMACQs
Sets (or queries) the number of acquisitions to complete for the event type
ACQCOMPLete. The default is 1 acquisition.
To set the event type to
ACQCOMPLete,usetheACTONEVent:EVENTTYPe
command.
Group
Act on Event
Syntax
ACTONEVent:NUMACQs <NR1>
ACT
ONEVent:NUMACQs?
Arguments
Integer which can be set as follows:
Table 2-50:
From To Resolution
1 100 1
100 1000 10
1000 10,000 100
10,000 1,000,000 1000
Examples
ACTONEVENT:NUMACQS 350 specifies to take the specified action when 350
acquisitions have been completed.
ACTONEVent:REPEATCount
Sets or returns the number of events to run.
To specify an event, use the command ACTONEVent:EVENTTYPe.
Group
Act on Event
Syntax
ACTONEVent:REPEATCount <NR1>
ACTONEVent:REPEATCount?
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-123

Commands Listed in Alphabetical Order
Arguments
NR1 is an intege
r that specifies the number of events to run. The default
is 1 event and the maximum is 1000000 events. A repeat count greater
than 1000000 specifies to run forever. In this case, the action on event can be
terminated by setting the state for all actions to OFF. When the repeat count is set
to infinite, the query returns 9.9E+37 (IEEE positive infinity).
AFG:AMPLitude
Sets (or qu
eries) the AFG amplitude in volts, peak to peak. The minimum and
maximum allowable settings are dependent upon the output load impedance setting
(AFG:OUTPut:LOAd:IMPEDance) and the selected function (AFG:FUNCtion).
NOTE. The AFG:OUTPut:LOAd:IMPEDance command must be used before
using any of the amplitude/offset/highLevel/lowLevel/Preset commands, so that
saved setups are recalled with the proper load factor applied.
The following values can be set for AFG:AMPLitud e:
AFG AMPLitude Setting Value
Load I
mpedance
FIFTY Ohm
Load I
mpedance
HIGHZ
Function
Min Max Resolution Min Max Resolution
SINE
SQUare
PULS
e
RAMP
NOISe
DC
1
HAVERSINe
CARDIac
AR
Bitrary
10.0
mV
2.5V 500u
V
20.0
mV
5.0V 1mV
SINC
10.0mV 1.5V 500uV 20.0mV 3.0V 1mV
LORENtz
1
0.0mV
1
.2V
5
00uV
2
0.0mV
2
.4V
1
mV
G
AUSsian
ERISe
EDECAy
10.0mV 1.25V 500uV 20.0mV 2.5V 1mV
1
The DC function does not use the amplitude setting. The DC level is set using :AFG:OFFSet. Ampltude settings that are set when the :AFG:FUNCtion is DC
are remembered and applied when the function is changed to other than DC.
2-124 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual

Commands Listed in Alphabetical Order
Conditions
MDO3000 series
with option MDO3AFG installed and MDO4000C series with
option MDO4AFG installed only.
Group
AFG
Syntax
AFG:AMPLitude <NR3>
AFG:AMPLitude?
Arguments
Floating point number that represents the AFG amplitude, peak to peak, in volts.
Examples
AFG:AMPLITUDE 1.0 sets the AFG amplitude to 1.0 volts, peak to peak.
AFG:ARBitrary:ARB<x>:DATE? (Query Only)
Returns the date that the data in the specified arbitrary waveform slot 1-4 was
saved.
Conditions
MDO3000 series with option MDO3AFG installed and MDO4000C series with
option MDO4AFG installed only.
Group
ARB
Syntax
AFG:ARBitrary:ARB<x>:DATE?
Examples
AFG:ARB:ARB3:DATE? might return 11-11-13, which represents the date that
thedatainslot3wassaved.
AFG:ARBitrary:ARB<x>:LABel
Sets (or queries) the waveform label for arbitrary waveform slots 1- 4.
Conditions
MDO3000 series with option MDO3AFG installed and MDO4000C series with
option MDO4AFG installed only.
Group
ARB
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-125