MDO3000 Programmer Manual.pdf - 第168页
Commands Listed in A lphabetical Order AFG:OFFSet Sets (or que ries) the AFG offset value , in volts. The offset values that can be specified for each function and load im pedance are in the table below . (T o specify the…

Commands Listed in Alphabetical Order
The maximum noi
se percent is limited when AFG:AMPLitude is greater than
50% of its maximum setting value for the current function, in accordance with
the following formula:
Noise Percent Max = 100.0 * (Amplitudemax/Amplitude – 1.0)
Conditions
MDO3000 series with option MDO3AFG installed and MDO4000C series with
option MDO4AFG installed only.
Group
AFG
Syntax
AFG:NOISEAdd:PERCent <NR3>
AFG:NOISEAdd:PERCent?
Arguments
Floating point number that represents the AFG additive noise level, as a
percentage.
Examples
AFG:NOISEADD:PERCENT 50 sets the AFG additive noise level to 50 percent.
AFG:NOISEAdd:STATE
Sets (or queries) the AFG additive noise state.
Conditions
MDO3000 series with option MDO3AFG installed and MDO4000C series with
option MDO4AFG installed only.
Group
AFG
Syntax
AFG:NOISEAdd:STATE {0|1|OFF|ON}
AFG:NOISEAdd:STATE?
Arguments
1 orON turns on the AFG additive noise state.
0 or OFF turns it off.
Examples
AFG:NOISEADD:STATE ON turns on the additive noise state.
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-137

Commands Listed in Alphabetical Order
AFG:OFFSet
Sets (or queries) the AFG offset value, in volts. The offset values that can be
specified for each function and load impedance are in the table below. (To
specify the f
unction and load impedance, use the commands AFG:FUNCtion
and AFG:OUTPut:LOAd:IMPEDance.)
NOTE. The AFG:OUTPut:LOAd:IMPEDance command must be used before
using any of the amplitude/offset/highLevel/lowLevel/Preset commands, so that
saved setups are recalled with the proper load factor applied.
AFG:OFFSet Setting Value
Load Impedance
FIFTY Oh
m
Load Impedance
HIGHZ
Min Max Resolution Min Max Resolution
SINE
SQUare
PULSe
RAMP
NOISe
DC
SINC (
Sin(x)/x)
GAUSsian
LORENtz
ERIS
e
EDECAy
HAVERSINe
CARD
Iac
ARBitrary
-1.25V 1.25V 500uV -2.5V 2.5V 1mV
Conditions
MDO3000 series with option MDO3AFG installed and MDO4000C series with
option MDO4AFG installed only.
Group
AFG
Syntax
AFG:OFFSet <NR3>
AFG:OFFSet?
Arguments
Floating point number that represents the AFG offset, in volts.
2-138 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual

Commands Listed in Alphabetical Order
Examples
AFG:OFFSET 1.0
sets the AFG offset to 1.0 volts.
AFG:OUTPut:LOAd:IMPEDance
Sets (or queries) the AFG output load impedance.
Conditions
MDO3000 series with option MDO3AFG installed and MDO4000C series with
option MDO4AFG installed only.
Group
AFG
Syntax
AFG:OUTPut:LOAd:IMPEDance {FIFty|HIGHZ}
AFG:OUTPut:LOAd:IMPEDance?
Arguments
FIFty sets the output load impedance to 50 Ohms.
HIGHZ sets the output load impedance to the high-impedance state.
Examples
AFG:OUTP:LOA:IMPED FIF sets the AFG output load impedance to 50 Ohms.
AFG:OUTPut:STATE
Sets (or queries) the AFG output state.
Conditions
MDO3000 series with option MDO3AFG installed and MDO4000C series with
option MDO4AFG installed only.
Group
AFG
Syntax
AFG:OUTPut:STATE {0|1|OFF|ON}
AFG:OUTPut:STATE?
Arguments
1 orON turns on the AFG output state.
0 or OFF turns it off.
Examples
AFG:OUTPUT:STATE ON turns on the AFG output state.
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-139