Commands Listed in A lphabetical Order Syntax APPLication:L ICENSE:SLOT<x>:T YPe? Related Commands APPLication:LICENSE:SLOT<x>:TRANSFER , APPLication:LICENSE:SLOT<x>:LOCation? Examples APPLication:M ODu…

Commands Listed in Alphabetical Order
If the license c
urrently resides in the scope and the license does not also reside in
the module, the license is transferred from the scope to the module.
If the applica
tion module slot is empty, an error event is posted to the event queue
so indicating and no operation is performed.
If the licen
se resides in both the scope and the module, an error event is posted to
the event queue so indicating and no operation is performed.
If the licen
se resides in neither the scope nor the module, an error event is posted
to the event queue so indicating and no operation is performed.
Group
Miscellaneous
Syntax
APPLication:LICENSE:SLOT<x>:TRANSFER EXECute
Related Commands
APPLication:LICENSE:SLOT<x>:LOCation?,
APPLication:LICENSE:SLOT<x>:TYPe?
APPLication:LICENSE:SLOT<x>:TYPe? (Query Only)
This query returns the application license type of the module that is currently
inserted in the specified application module slot. If there is no application
module in the slot,
NONE is returned. < x> can be slot number 1-4 (1-2 for
MDO3000 models).
NOTE. In order to use many of the advanced oscilloscope features, you must
purchase an application module that contains a license. A license turns on a
feature that is already present in the firmware. The module can be inserted
into any of the four application module slots in the oscilloscope (two for the
MDO3000 series). While only four of the more than twelve physical application
modules available c an be supported at one time (or two for the MDO3000 s eries),
more than this number of applications can be used simultaneously: a license can
be transferred from the physical application module to an internal location within
the oscilloscope. Once a license has been transferred to an internal location, the
application that it enables can be used without the physical application module
being present; the physical application module can be removed, thus freeing up
a slot. However, the license then needs to be transferred back to the physical
application module in order to use the license with another instrument.
Group
Miscellaneous
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-147

Commands Listed in Alphabetical Order
Syntax
APPLication:L
ICENSE:SLOT<x>:TYPe?
Related Commands
APPLication:LICENSE:SLOT<x>:TRANSFER,
APPLication:LICENSE:SLOT<x>:LOCation?
Examples
APPLication:MODule:SLOT1:TYPe? might return DPO4AUTO, indicating that
the DPO4AUTO license is installed in slot 1.
APPLication:TYPe
This command sets or returns the application type. These are the applications
available when the front panel "Test" button is pressed. The query form willreturn
NONe if none of the supported test application modules are installed.
Attempting to set the application type to a type with no application module
installed will result in a settings conflict error event.
NOTE. The Video Picture features are standard on MDO3000 and
MDO4000C series models. They require the
DPO4VID application module for
use with MSO/DPO4000B, MDO4000 and MDO4000B models.
NOTE. The Act on Event features are standard on all models. The Limit/Mask and
Power features are optional for all models.
Conditions
Video Picture commands are available for MDO3000 and MDO4000C series
models. They require the DPO4VID application module for use with
MSO/DPO4000B, MDO4000 and MDO4000B models
The Limit/Mask and Power features require the MDO3LMT or MDO3PWR
application modules for the MDO3000 series models, or the DPO4LMT or
DPO4PWR application modules for the MDO4000/B/C series.
Group
Miscellaneous
Syntax
APPLication:TYPe {POWer|LIMITMask|VIDPic|ACTONEVent|NONe}
APPLication:TYPe?
Related Commands
All of the commands associated with the specified application type.
2-148 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual

Commands Listed in Alphabetical Order
Arguments
POWer sets the a
pplication type to power analysis.
LIMITMask sets the application type to limit mask test.
VIDPic sets the application type to video picture.
ACTONEVent sets the application type to act on event.
NONe
Returns
The query form will return NONe if none of the supported application modules
are installed.
Examples
APPL:TYP? might return LIMITMASK, indicating that the DPO4LMT application
module is currently being used.
AUTOSet (No Query Form)
Sets the vertical, horizontal, and trigger controls of the oscilloscope to
automatically acquire and display the selected waveform.
Group
Miscellaneous
Syntax
AUTOSet {EXECute|UNDo}
Arguments
EXECute autosets the displayed waveform.
UNDo restores the oscilloscope settings to those present prior to the autoset
execution.
Examples
AUTOSET EXECUTE vertical, horizontal, and trigger controls of the oscilloscope to
automatically acquire and display all waveforms which are turned on.
AUTOSet:ENAble
Enables or disables the autoset feature. This is useful for classroom purposes
where the instructor wants the students to achieve the desired instrument settings
without the benefit of the autoset feature. This setting is not saved in setupfilesor
SET? or *LRN? queries. The default state is 1 (autoset enabled).
Group
Miscellaneous
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-149