MDO3000 Programmer Manual.pdf - 第217页

Commands Listed in Alphabetical Order Examples BUS:B1:MIL155 3B:RESPonsetime: MAXimum 12.0E-6 specifies the maximum response time to a valid command received to be 12.0000E-6 microseconds. BUS:B1:MIL15 53B:RESPonseti me:M…

100%1 / 1158
Commands Listed in Alphabetical Order
Group
Bus
Syntax
BUS:B<x>:MIL1553B:POLarity {NORMal|INVERTed}
BUS:B<x>:MIL1553B:POLarity?
Related Commands
TRIGger:A:BUS:B<x>:MIL1553B:CONDition
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:MIL1553B:CONDition
BUS:B<x>:MIL1553B:SOUrce
Arguments
NORMal - A high-low transition represents a 1 on the Data+ line.
INVERTed - A high-low transition represents a 0 on the Data+ line.
Examples
BUS:B1:MIL1553B:POLarity INVERTED sets the polarity so that a high-low
transition represents a 0 on the Data+ line.
BUS:B1:MIL1553B:POLarity? might return NORMAL.
BUS:B<x>:MIL1553B:RESPonsetime:MAXimum
This command specifies the maximum response time to a valid command issued
for the MIL-STD-1553 bus.
NOTE. The MIL-STD-1553 bus speci cation requires devices to respond to a
valid command within 4 to 12 microseconds.
Conditions
This command requires a DPO4AERO application module for 4000 series models,
and an MDO3AERO application module for MDO3000 series models.
Group
Bus
Syntax
BUS:B<x>:MIL1553B:RESPonsetime:MAXimum <NR3>
BUS:B<x>:MIL1553B:RESPonsetime:MAXimum?
Related Commands
BUS:B<x>:MIL1553B:RESPonsetime:MINimum
Arguments
<NR3> is a floating point number that specifies the maximum response time, in
seconds.
2-186 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual
Commands Listed in Alphabetical Order
Examples
BUS:B1:MIL155
3B:RESPonsetime:MAXimum
12.0E-6 specifies the maximum
response time to a valid command received to be 12.0000E-6 microseconds.
BUS:B1:MIL15
53B:RESPonsetime:MAXimum?
might return 12.0000E-6.
BUS:B<x>:MIL1553B:RESPonsetime:MINimum
This command specifies the minimum response time to a valid command issued
for the MIL-STD-1553 bus.
NOTE. The MIL-STD-1553 bus specication requires devices to respond to a
valid command within 4 to 12 microseconds.
Conditions
This command requires a DPO4AERO application module for 4000 series models,
and an MDO3AERO application module for MDO3000 series models.
Group
Bus
Syntax
BUS:B<x>:MIL1553B:RESPonsetime:MINimum <NR3>
BUS:B<x>:MIL1553B:RESPonsetime:MINimum?
Related Commands
BUS:B<x>:MIL1553B:RESPonsetime:MAXimum
Arguments
<NR3> is a floating point number that specifies the minimum response time in
seconds.
Examples
BUS:B1:MIL1553B:RESPonsetime:MINimum 4.0E-6 specifies the minimum
responsetimetoavalidcommandreceivedto4.0E-6microseconds.
BUS:B1:MIL1553B:RESPonsetime:MINimum? might return 4.0000E-6.
BUS:B<x>:MIL1553B:SOUrce
This command specifies the source for differential input for the MIL-STD-1553
bus. The supported source waveforms are channels 1–4, math waveform, and
reference waveforms 1–4. The default is channel 1.
Conditions
This command requires a DPO4AERO application module for 4000 series models,
and an MDO3AERO application module for MDO3000 series models.
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-187
Commands Listed in Alphabetical Order
Group
Bus
Syntax
BUS:B<x>:MIL1553B:SOUrce
{CH1|CH2|CH3|CH4|MATH|REF1|REF2|REF3|REF4}
BUS:B<x>:MIL1553B:SOUrce?
Related Commands
BUS:B<x>:ETHERnet:PRObe
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:MIL1553B:CONDition
TRIGger:A:BUS:B<x>:MIL1553B:CONDition
Arguments
CH1–4 specifies an analog channel as the source for differential input.
MATH specifies the math waveform as the MIL-STD-1553 bus source for
differential input
REF1–4 specifies a reference waveform as the source for differential input.
Examples
BUS:B1:MIL1553B:SOUrce ref4 specifies to use reference waveform 4 as
the source.
BUS:B1:MIL1553B:SOUrce? might return CH2, indicating that channel 2 is
the currently specified source.
BUS:B<x>:PARallel:BIT<x>:SOUrce
Specifies the bit source waveform for the parallel bus.
Conditions
Requires an MSO/MDO model, with the MDO3MSO option installed for the
MDO3000 series. The MSO4000B and MDO4000/B/C models can trigger on
parallel buses without an application module.
Group
Bus
Syntax
BUS:B<x>:PARallel:BIT<x>:SOUrce {CH1|CH2|CH3|CH4|
D0|D1|D2|D3|D4|D5|D6|D7|D8|D9|D10|D11|D12|D13|D14|D15}
BUS:B<x>:PARallel:BIT<x>:SOUrce?
2-188 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual