Commands Listed in A lphabetical Order CALibrate:RE Sults:F ACtory:RF? (Query Only) This query returns the factory calibration status for the RF portion of the instrument, if present. This query is synonymous with CALibr…

Commands Listed in Alphabetical Order
common, howeve
r, to use uncalibrated probes (particularly when the oscilloscope
inputs are connected into a test system with coaxial cables).
Group
Calibration
and Diagnostic
Syntax
CALibrate:RESults?
Related Commands
*CAL?
CALibrate:RESults:FACtory? (Query Only)
Returns the status of internal and factory calibration, without performing any
calibration operations.
Group
Calibration and Diagnostic
Syntax
CALibrate:RESults:FACtory?
CALibrate:RESults:FACtory:AFG? (Query Only)
This query returns the factory calibration status for the Arbitrary Function
Generator portion of the instrument, if present.
This query is synonymous with
CALibrate:FACtory:STATus:AFG?
Conditions
Available for MDO3000 and MDO4000C series only.
Group
Calibration and Diagnostic
Syntax
CALibrate:RESults:FACtory:AFG?
Returns
PASS - factory calibration has passed
INIT - factory calibration is unadjusted
RUNNING - factory calibration is running
Examples
CALIBRATE:RESULTS:FACTORY:AFG? might return RUNNING, indicating that
the factory calibration for the AFG portion of the instrument is in progress.
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-215

Commands Listed in Alphabetical Order
CALibrate:RE
Sults:FACtory:RF? (Query Only)
This query returns the factory calibration status for the RF portion of the
instrument, if present.
This query is synonymous with
CALibrate:FACtory:STATus:RF?
Group
Calibration and Diagnostic
Syntax
CALibrate:RESults:FACtory:RF?
Returns
PASS - factory calibration has passed
INIT - factory calibration is unadjusted
RUNNING - factory calibration is in progress
Examples
CALIBRATE:RESULTS:FACTORY:RF? might return RUNNING, indicating that the
factory calibration for the RF portion of the instrument is in progress.
CALibrate:RESults:FACtory:SCOPE? (Query Only)
This query returns the factory calibration status for the oscilloscope (doesn't
include RF or AFG) of the instrument.
This query is synonymous with the following query:
:CALibrateFACtory:STATus:SCOPE?
Group
Calibration and Diagnostic
Syntax
CALibrate:RESults:FACtory:SCOPE?
Returns
This query will return one of the following:
INIT indicates the oscilloscope portion of the instrument has not been calibrated.
PASS indicates the oscilloscope internal calibration completed successfully.
RUNNING indicates the oscilloscope internal calibration is currently running.
Examples
CAL:RESULTS:SPC:SCOPE? might return INIT indicating that the oscilloscope
portion of the instrument has not been calibrated.
2-216 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual

Commands Listed in Alphabetical Order
CALibrate:RE
Sults:SPC? (Query Only)
Returns the status of the SPC operation. This query does not initiate a SPC.
Group
Calibration and Diagnostic
Syntax
CALibrate:RESults:SPC?
Related Commands
*CAL?
Returns
INIT indicates that SPC has never successfully completed.
PASS indicates that the last SPC operation passed.
FAIL indicates that the last SPC operation failed.
RUNNING indicates that the SPC operation is running.
Examples
CALIBRATE:RESULTS:SPC? might return FAIL, indicating that the last SPC
operation failed.
CALibrate:RESults:SPC:RF? (Query Only)
This query returns the status of the last SPC run for the RF portion of the
instrument (doesn't include analog channels) . This query is synonymous with
CALibrate:INTERNal:STATus:RF?
Group
Calibration and Diagnostic
Syntax
CALibrate:RESults:SPC:RF?
Returns
This query will return one of the following:
INIT indicates the RF portion of the instrument has not been calibrated.
PASS indicates the RF internal calibration completed successfully.
RUNNING indicates the oscilloscope internal calibration is currently running.
FAIL indicates the RF internal calibration did not complete successfully.
Examples
CAL:RESULS:SPC:RF? might return INIT indicating that the RF portion of the
instrument has not been internally calibrated.
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-217