MDO3000 Programmer Manual.pdf - 第273页
Commands Listed in Alphabetical Order Examples CONFIG:ARB? mi ght return 1, indicating that the hardware is present and the feature is enabled. CONFIGuration:AUXIN? (Query Only) This query returns a boolean value to indi…

Commands Listed in Alphabetical Order
CONFIGuratio
n:APPLications:VIDPIC? (Query Only)
Indicates whether the Video Picture feature is present and enabled.
Conditions
Available for MDO3000 and MDO4000C series models. Requires the DPO4VID
application module for use with MSO/DPO4000B, MDO4000 and MDO4000B
models.
Group
Configuration
Syntax
CONFIGuration:APPLications:VIDPIC?
Returns
1 indicates that the Video Picture featureispresentandisenabled.
0 indicates either the Video Picture feature is not present or the feature is not
enabled.
Examples
CONFIG:APPL:VIDPIC? might return 1, indicating that the hardware is present
and the feature is enabled.
CONFIGuration:ARB? (Query Only)
Indicates whether or not the arbitrary function generator hardware is present, and
the user-defined arbitrary waveform generation feature is enabled. Note that this is
different than the
CONFIGuration:AFG? query. The ability to generate arbitrary
waveforms is an extension of the standard AFG features.
Conditions
Available for MDO3000 series models only.
Group
Configuration
Syntax
CONFIGuration:ARB?
Returns
1 indicates that the arbitrary function generator hardware is present and the
arbitrary function generation feature is enabled.
0 indicates that either the arbitrary function generator hardware is not present, or
the user-defined arbitrary waveform generation feature is not enabled.
2-242 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual

Commands Listed in Alphabetical Order
Examples
CONFIG:ARB? mi
ght return 1, indicating that the hardware is present and the
feature is enabled.
CONFIGuration:AUXIN? (Query Only)
This query returns a boolean value to indicate whether the instrument has an
Aux Input connector.
NOTE. The MSO/DPO4000B series and 2-channel MDO3000 series models
contain an
Aux Input connector.
Group
Configuration
Syntax
CONFIGuration:AUXIN?
Related Commands
(See page 2-91, Vertical Command Group.)
Returns
<NR1> = 1 if the instrument has an Aux Input connector.
<NR1> = 0 if the instrument does not have an Aux Input connector.
CONFIGuration:BUSWAVEFORMS:ARINC429A? (Query Only)
This query returns a Boolean value to indicate whether the optional CAN-FD bus
triggering and analysis feature is present.
Conditions
This feature requires installation of a DPO4AERO application module for
MDO4000C models, or MDO3AERO for MDO3000 models.
Group
Configuration
Syntax
CONFIGuration:BUSWAVEFORMS:ARINC429A?
Related Commands
(See page 2-15, Bus Command Group.)
(See page 2-76, Trigger Command Group.)
(See page 2-61, Search Command Group.)
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-243

Commands Listed in Alphabetical Order
Returns
<NR1> = 1 if the A
RINC429 bus feature is present.
<NR1> = 0 if the ARINC429 bus feature is not present.
CONFIGuration:BUSWAVEFORMS:AUDIO? (Query Only)
This query returns a boolean value to indicate whether the optio
nal audio bus
triggering and analysis feature is present.
Conditions
This feature requires installation of a DPO4AUDIO applicati
on module, or
MDO3AUDIO for MDO3000 models.
Group
Configuration
Syntax
CONFIGuration:BUSWAVEFORMS:AUDIO?
Related Commands
(See page 2-15, Bus Command Group.)
(See page 2-76, Trigger Command Group.)
(See page 2-61, Search Command Group.)
Returns
<NR1> =1iftheaudiobusfeatureispresent.
<NR1> = 0 if the audio bus featu
re is not present.
CONFIGuration:BUSWAVEFORMS:CAN? (Query Only)
This query returns a boolean value to indicate whether the optional CAN bus
triggering and analysis feature is present.
Conditions
This feature requires installation of a DPO4AUTO application module, or
MDO3AUTO for MDO3000 models.
Group
Configuration
Syntax
CONFIGuration:BUSWAVEFORMS:CAN?
Related Commands
(See page 2-15, Bus Command Group.)
(See page 2-76, Trigger Command Group.)
2-244 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual