Commands Listed in Alphabetical Order Arguments <NR3> is a fl oati ng point number that specifies the cursor position relative to ground. Examples CURSOR:HBARS: POSITION1 25. 0E-3 positions Cursor 1 of the horizontal…

100%1 / 1158
Commands Listed in Alphabetical Order
Group
Cursor
Syntax
CURSor:HBArs?
Examples
CURSOR:HBARS? might return the horizontal
bar setting as
:CURSOR:HBARS:POSITION1
320.0000E-03;POSITION2-320.0000E-03;UNITS BASE
CURSor:HBArs:DELTa ? (Query Only)
Returns the vertical difference between the two horizontal bar cursors.
Group
Cursor
Syntax
CURSor:HBArs:DELTa?
Related Commands
CURSor:HBArs:UNIts
Returns
A floating point value with an exponent.
Examples
CURSOR:HBARS:DELTA? might return :CUR SOR:HBARS:DELTA 5.0800E+00
indicating that the difference between the two cursors is 5.08.
CURSor:HBArs:POSITION<x>
This command specifies the horizontal bar cursor position relative to ground,
whichisexpressedinverticalunits(usually volts). The cursor is specified by
x,whichcanbe1or2.
Group
Cursor
Syntax
CURSor:HBArs:POSITION<x> <NR3>
CURSor:HBArs:POSITION<x>?
Related Commands
CURSor:FUNCtion
2-258 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual
Commands Listed in Alphabetical Order
Arguments
<NR3> is a oati
ng point number that specifies the cursor position relative to
ground.
Examples
CURSOR:HBARS:POSITION1 25.0E-3 positions Cursor 1 of the horizontal
cursors at 25 mV.
CURSOR:HBARS:POSITION2? might return :CURSOR:HBARS:POSITION2
-64.0000E-03
indicating that Cursor 2 of the horizontal bar cursors is at -64 mV.
CURSor:HBArs:UNIts
This command specifies the units for the horizontal bar cursors.
Group
Cursor
Syntax
CURSor:HBArs:UNIts {BASE|PERcent}
CURSor:HBArs:UNIts?
Arguments
BASE selects the vertical units for the selected waveform.
PERcent selects ratio cursors.
Examples
CURSOR:HBARS:UNITS might return :C UR
SOR:HBARS:UNITS BASE
indicating
that the units for the horizontal bar cursors are base.
CURSor:HBArs:USE (No Query Form)
This command specifies the horizontal bar cursor measurement scale. This
command is only applicable when the ratio cursors are turned on.
Group
Cursor
Syntax
CURSor:HBArs:USE {CURrent|HALFgrat}
Related Commands
CURSor:HBArs:UNIts
Arguments
CURrent sets the H Bar measurement scale so that 0% is the current position of the
lowest H Bar cursor and 100% is the current position of the highest H Bar cursor.
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-259
Commands Listed in Alphabetical Order
HALFgrat sets H
Bar measurement scale so that half the screen major divisions
(five on the MSO/DPO4000B and MDO4000/B/C, four on the MDO3000) is
100%, where 0% is -2.5 divisions and 100% is +2.5 divisions from the center
horizontal graticule.
Examples
CURSOR:HBARS:USE HALFGRAT sets the H Bar measurement scale so that
five screen major divisions equals 100%.
CURSor:MODe
This command specifies whether the two cursors move linked together in unison
or separately. This command is equivalent to setting Linked to On or Off in the
Cursor menu.
Group
Cursor
Syntax
CURSor:MODe {TRACk|IND ependent}
CURSor:MODe?
Arguments
TRACk ties the navigational functionality of the two cursors together. For cursor 1
adjustments, this ties the movement of the two cursors together; however, cursor 2
continues to move independently of cursor 1. This mode only applies when the
DISplay:XY:WITHYT is set to
OFF.
INDependent allows independent adjustment of the two cursors.
Examples
CURSOR:MODE TRACK specifies that the cursor positions move in unison.
CURSOR:MODE? might return :CURSOR:MODE TRACK indicating that the two
cursorsmoveinunison.
CURSor:SOUrce
This command specifies the cursor source waveform, which can be one of channels
1-4, reference waveforms 1-4, math waveform, bus 1-4 (1-3 for the MDO4000C
and 1-2 for the MDO3000), digital channels 0-15 (MSO/MDO4000/B models
only as well as MDO3000 and MDO400C models with option MDO3MSO or
MDO4MSO installed), the RF time domain traces (MDO4000/B/C models), or
AUTO. When the cursor source is set to AUTO, cursor readouts apply to the
currently selected waveform.
Group
Cursor
2-260 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual