MDO3000 Programmer Manual.pdf - 第326页
Commands Listed in A lphabetical Order Related Commands All CURSOR:XY c ommands. DVM:DISPLA YSTYle Arguments 1 or ON turns on simultaneous display of the XY and YT waveforms when in TRIGgered XY display mode. 0 or OFF tu…

Commands Listed in Alphabetical Order
Syntax
DISplay:XY {OF
F|TRIGgered}
DISplay:XY?
Related Commands
All CURSOR:XY commands.
Arguments
OFF – The channels are displayed indivi
dually as a function of time.
TRIGgered – The channels are displayed in “X-Y” pairs with CH1 being
displayed as a function of CH2, and so o
n.
Examples
DISplay:XY TRIG specifies to use the XY display mode.
DISPLAY:XY? might return OFF, indicating that the XY disp
lay mode is not
currently on.
DISplay:XY:WITHYT
Sets or returns the state of simultaneous display of the XY and YT waveforms
when in TRIGgered XY display mode. (To set the mode to TRIGgered XY
display, first use the command
:DISplay:XY TRIGgered.)
When both the XY and YT waveforms are displayed, the YT waveform is
displayed in the upper graticule, and the XY waveform is displayed in the lower
graticule.
This command applies to the MDO3000 and MDO4000C series only. For this
platform, the default state is 0 (XY display only)
.
For the DPO/MSO4000B and MDO4000/B platforms, this command is ignored
and the state is always 1 (to display both the XY an
d YT waveforms when in
XY display mode).
NOTE. For models with the DVM feature, such as the MDO3000 and MDO4000C,
this command does not take effect if the :DVM:DISPLAYSTYle is set to FULl,
but will take effect once the :DVM:DISPLAYSTYle is set to MINimum or if the
DVM is turned off.
Group
Display
Syntax
DISplay:XY:WITHYT {0|1|OFF|ON}
DISplay:XY:WITHYT?
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-295

Commands Listed in Alphabetical Order
Related Commands
All CURSOR:XY c
ommands.
DVM:DISPLAYSTYle
Arguments
1 or ON turns on simultaneous display of the XY and YT waveforms when in
TRIGgered XY display mode.
0 or OFF turns simultaneous display off.
Examples
DISplay:XY:WITHYT turns simultaneous display of the XY and YT waveforms
when in TRIGgered XY display mode
DVM (No Query Form)
Resets the Digital Voltmeter measurements and history.
Conditions
MDO3000 and MDO4000C only; requires DVM option key to be installed.
Group
DVM
Syntax
DVM {RE SET}
Arguments
None
Examples
DVM:RESET resets the DVM measurement and history.
DVM:AUTORange
Sets (or queries) the auto range state for the Digital Voltmeter.
Conditions
MDO3000 and MDO4000C only; requires DVM option key to be installed.
Group
DVM
Syntax
DVM:AUTORange {0|1
|OFF|ON}
DVM:AUTORange?
2-296 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual

Commands Listed in Alphabetical Order
Arguments
1 or ON turns on t
he auto range state for the Digital Voltmeter.
0 or OFF turns it off.
Examples
DVM:AUTOR ON turns on the auto range state for the Digital Voltmeter.
DVM:DISPLAYSTYle
Sets (or queries) the display style for the Digital V
oltmeter.
Conditions
MDO3000 and MDO4000C only; requires DVM option key to be installed.
Group
DVM
Syntax
DVM:DISPLAYSTYle {FULl|MINimum}
DVM:DISPLAYSTYle?
Arguments
FULl sets the display style for the DVM to Full.
MINimum sets the display style for the DVM to Minimum.
Examples
DVM:DISPLAYSTYLE MIN sets the display style for the DVM to Minimum.
DVM:MEASUrement:FREQuency? (Query Only)
Returns the current frequency value for the Digital Voltmeter.
Any qualifier, such as "Measurement warning, Clipping positive", is automatically
set as an error event.
Conditions
MDO3000 and MDO4000C only; requires DVM option key to be installed.
Group
DVM
Syntax
DVM:MEASUrement:FREQuency?
Examples
DVM:MEASU:FREQ? might return 100.0000E+3, which represents 100 KHz as
the current frequency value for the DVM.