MDO3000 Programmer Manual.pdf - 第328页

Commands Listed in A lphabetical Order DVM:MEASUrem ent:HIStory:A VErage? (Query Only) Returns the average readout value for the Digital V oltmeter function over the history period. Any qualifier , such as "Measureme…

100%1 / 1158
Commands Listed in Alphabetical Order
Arguments
1 or ON turns on t
he auto range state for the Digital Voltmeter.
0 or OFF turns it off.
Examples
DVM:AUTOR ON turns on the auto range state for the Digital Voltmeter.
DVM:DISPLAYSTYle
Sets (or queries) the display style for the Digital V
oltmeter.
Conditions
MDO3000 and MDO4000C only; requires DVM option key to be installed.
Group
DVM
Syntax
DVM:DISPLAYSTYle {FULl|MINimum}
DVM:DISPLAYSTYle?
Arguments
FULl sets the display style for the DVM to Full.
MINimum sets the display style for the DVM to Minimum.
Examples
DVM:DISPLAYSTYLE MIN sets the display style for the DVM to Minimum.
DVM:MEASUrement:FREQuency? (Query Only)
Returns the current frequency value for the Digital Voltmeter.
Any qualifier, such as "Measurement warning, Clipping positive", is automatically
set as an error event.
Conditions
MDO3000 and MDO4000C only; requires DVM option key to be installed.
Group
DVM
Syntax
DVM:MEASUrement:FREQuency?
Examples
DVM:MEASU:FREQ? might return 100.0000E+3, which represents 100 KHz as
the current frequency value for the DVM.
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-297
Commands Listed in Alphabetical Order
DVM:MEASUrem
ent:HIStory:AVErage? (Query Only)
Returns the average readout value for the Digital Voltmeter function over the
history period.
Any qualifier, such as "Measurement warning, Clipping positive", is automatically
set as an error event.
Conditions
MDO3000 and MDO4000C only; requires DVM option key to be installed.
Group
DVM
Syntax
DVM:MEASUrement:HIStory:AVErage?
Examples
DVM:MEASU:HIS:AVE? might return 100.0000E+3, which represents the
average DVM readout value over the history period.
DVM:MEASUrement:HIStory:MAXimum? (Query Only)
Returns the maximum readout value for the DVM function over the history period.
Any qualifier, such as "Measurement warning, Clipping positive", is automatically
set as an error event.
Conditions
MDO3000 and MDO4000C only; requires DVM option key to be installed.
Group
DVM
Syntax
DVM:MEASUrement:HIStory:MAXimum?
Examples
DVM:MEASU:HIS:MAX? might return 100.0000E+3, which represents the
maximum readout value for the DVM function over the history period.
DVM:MEASUrement:HIStory:MINImum? (Query Only)
Returns the minimum readout value for the DVM function over the history period.
Any qualifier, such as "Measurement warning, Clipping positive", is automatically
set as an error event.
2-298 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual
Commands Listed in Alphabetical Order
Conditions
MDO3000 and MDO
4000C only; requires DVM option key to be installed.
Group
DVM
Synt
ax
DVM:MEASUrement:HIStory:MINImum?
Examples
DVM:MEASU:HIS:MINI? might return 100.0000E+3, which represents the
minimum readout value for the DVM function over the history period.
DVM:MEASUrement:INFMAXimum? (Query Only)
Returns the maximum DVM readout value over the entire time that the DVM has
been on since the last change using the DVM:MODe or DVM:SOUrce commands
or
DVM RESE T.
Any qualifier, such as "Measurement warning, Clipping positive", is automatically
set as an error event.
Conditions
MDO3000 and MDO4000C only; requires DVM option key to be installed.
Group
DVM
Syntax
DVM:MEASUrement:INFMAXimum?
Examples
DVM:MEASU:INFMAX? might return 100.0000E+3, which represents the
maximum readout value of the DVM function over the entire time that the DVM
has been on since the last change using the DVM:MODe or DVM:SOUrce
commands or
DVM RESET.
DVM:MEASUrement:INFMINimum? (Query Only)
Returns the minimum readout value of the DVM function over the entire time
that the DVM has been on since the last change using the DVM:MODe or
DVM:SOUrce commands or
DVM RESE T.
Any qualifier, such as "Measurement warning, Clipping positive", is automatically
set as an error event.
Conditions
MDO3000 and MDO4000C only; requires DVM option key to be installed.
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-299