MDO3000 Programmer Manual.pdf - 第413页

Commands Listed in Alphabetical Order Syntax MASK:TEMPLate :SOUrce {C H1| CH2|CH3|CH4|REF 1|REF2|REF3|REF4 } MASK:TEMPLate :SOUrce? Related Commands MASK:TEMPLate:CREA TEmask , MASK:TEMPLate:TOLerance:HORizontal , MASK:T…

100%1 / 1158
Commands Listed in Alphabetical Order
Aseriesofexam
ples showing how to use mask commands for typical tasks
is included in an appendix. (See page E-1, Mask/Limit Command Sequence
Examples.)
NOTE. To ensure that Mask commands are enabled, use the command
APPLication:TYPe LIMITMask.
Conditions
This comma
nd requires a DPO4LMT application module for 4000 series models,
and an MDO3LMT application module for MDO3000 series models.
Group
Mask
Syntax
MASK:TEMPLate:CREATEmask
Related Commands
MASK:TEMPLate:SOUrce,
MASK:TEMPLate:TOLerance:HORizontal,
MASK:TEMPLate:TOLerance:VERTical,
SAVe:MASK
MASK:TEMPLate:SOUrce
This command, typically used for limit mask testing, specifies the source
waveform to be used to create a template mask. The source can be either
one of four channels, or one of four saved reference waveforms. See the
SAVe:WAVEform commands for more information on creating reference
waveforms.
A series of examples showing how to use mask commands for typical tasks
is included in an appendix. (See pag
eE-1,Mask/Limit Command Sequence
Examples.)
NOTE. To ensure that Mask commands are enabled, use the command
APPLication:TYPe LIMITMask.
Conditions
This command requires a DPO4LMT application module for 4000 series models,
andanMDO3LMTapplicationmodu
le for MDO3000 series models.
Group
Mask
2-382 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual
Commands Listed in Alphabetical Order
Syntax
MASK:TEMPLate
:SOUrce {C H1|CH2|CH3|CH4|REF1|REF2|REF3|REF4}
MASK:TEMPLate:SOUrce?
Related Commands
MASK:TEMPLate:CREATEmask,
MASK:TEMPLate:TOLerance:HORizontal,
MASK:TEMPLate:TOLerance:VERTical,
SAVe:MASK
MASK:TEMPLate:TOLerance:HORizontal
This command specifies the horizontal limit (tolerance) for a template maskto
be used for limit testing. This indicates how far horizontally from the template
source to create the mask.
A series of examples showing how to use mask commands for typical tasks
is included in an appendix. (See page E-1, Mask/Limit Command Sequence
Examples.)
Conditions
This command requires a DPO4LMT application module for 4000 series models,
and an MDO3LMT application module for MDO3000 series models.
Group
Mask
Syntax
MASK:TEMPLate:TOLerance:HORizontal <NR3>
MASK:TEMPLate:TOLerance:HORizontal?
Related Commands
MASK:TEMPLate:CREATEmask,
MASK:TEMPLate:TOLerance:VERTical,
SAVe:MASK
Arguments
<NR3> is a floating point value that specifies the horizontal limit (tolerence) in
milli-divisions. Range: 0 to 500 mDiv, resolution: 1mDiv, default: 40mDiv.
This value cannot be negative.
Examples
If you create a limit mask after a default setup, then query:
:MASK:TEMPLate:TOLerance:HORizontal?
you get the response: 200.000E-3
Now send the commands:
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-383
Commands Listed in Alphabetical Order
:MASK:TEMPLat
e:TOLerance:HORizontal 0.4
:MASK:TEMPLate:CREATEmask
Notice that the horizontal gap in the mask is twice as tall.
MASK:TEMPLate:TOLerance:VERTical
This comman
d specifies the vertical limit (tolerance) for a template mask to be
used for limit testing. This indicates how far vertically from the templatesource
to create the mask.
A series of examples showing how to use mask commands for typical tasks
is included in an appendix. (See page E-1, Mask/Limit Command Sequence
Examples.)
Conditions
This command requires a DPO4LMT application module for 4000 series models,
and an MDO3LMT application module for MDO3000 series models.
Group
Mask
Syntax
MASK:TEMPLate:TOLerance:VERTical <NR3>
MASK:TEMPLate:TOLerance:VERTical?
Related Commands
MASK:TEMPLate:CREATEmask, MASK:TEMPLate:TOLerance:HORizontal,
SAVe:MASK
Arguments
<NR3> is a floating point value that specifies the vertical limit (tolerence) in
milli-divisions. Range: 0 to 1000 mDiv, resolution: 1 mDiv, default: 40mDiv.
This value cannot be negative.
Examples
If you create a limit mask after a default setup, then query:
:MASK:TEMPLate:TOLerance:VERTical?
you get the response: 200.000E
-3
Now send the commands:
:MASK:TEMPLate:TOLerance:VERTical 0.4
:MASK:TEMPLate:CREATEmask
Notice that the vertical gap in the mask is twice as tall.
2-384 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual