Commands Listed in Alphabetical Order
MASK:TESt:COM
PLetion:CRITerion
Arguments
ON turns on this features, so that the instrument stops acquiring data upon failure.
OFF turns off this feature.
<NR1> is an integer. 0 turns off this feature; any other value turns it on.
Examples
MASK:TESt:STOP:FAILure 1 stops data acquisition upon test failure.
MASK:TESt:STOP:FAILure? might return 0, indicating the instrument is not
set to stop acquiring data upon test failure.
MASK:TESt:THReshold
This command
specifies the number of failed tested waveforms needed in a
pass/fail mask test to cause the test status to change to ‘Failing’.
Aseriesofe
xamples showing how to use mask commands for typical tasks
is included in an appendix. (See page E-1, Mask/Limit Command Sequence
Examples.)
Conditions
This comman
d requires a DPO4LMT application module for 4000 series models,
and an MDO3LMT application module for MDO3000 series models.
Group
Mask
Syntax
MASK:TESt:THReshold {<NR1>|INFInite}
MASK:TESt:THReshold?
Related Commands
MASK:TESt:STATE,
MASK:TESt:COMPLetion:CRITerion
Arguments
<NR1> is an integer that specifies the number of tested waveform violations
occurring in each mask test that will change the test status to ‘Failing’. The
maximum number of failed tested waveforms that can be specified is 1,000,000.
The default is 1.
INFInite indicates that there is no limit to the number of failed tested waveforms
that will change a test status to ‘Failing’.
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-395
Commands Listed in Alphabetical Order
Examples
MASK:TESt:STO
P:FAILure 50000
sets the instrument to change the test status
to “failing” after 50,000 failed waveforms.
MASK:TESt:ST
OP:FAILure?
might return INFINITE which indicates the
instrument has been set so that there is no limit to tested waveform violations
per test.
MASK:TESt:
TIME
This command specifies the duration, in seconds, the instrument should run a
pass/fail mask test. The default is INFINITE.
A series of examples showing how to use mask commands for typical tasks
is included in an appendix. (See page E-1, Mask/Limit Command Sequence
Example
s.)
Conditions
This command requires a DPO4LMT application module for 4000 series models,
and an MDO3LMT application module for MDO3000 series models.
Group
Mask
Syntax
MASK:TESt:TIME {<NR1>|INFInite}
MASK:TESt:TIME?
Related Commands
MASK:TESt:STATE,
MASK:TESt:DELay
Arguments
<NR1> is an integer that represents the number of seconds to test. The maximum
number of seconds that can be specified is 172,800.
INFInite indicates that there is no time limit on the test run.
Examples
MASK:TESt:TIME 2 specifies to run a mask test for 2 seconds.
MASK:TESt:TIME? might return INFINITE, indicating there is currently no
specified time limit. This is the default.
MASK:TESt:WAVEform
This command specifies the number of waveforms the instrument should test
during a pass/fail mask test.
2-396 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual
Commands Listed in Alphabetical Order
Aseriesofexam
ples showing how to use mask commands for typical tasks
is included in an appendix. (See page E-1, Mask/Limit Command Sequence
Examples.)
Conditions
This command requires a DPO4LMT application module for 40
00 series models,
and an MDO3LMT application module for MDO3000 series models.
Group
Mask
Syntax
MASK:TESt:WAVEform {<NR1>|INFInite}
MASK:TESt:WAVEform?
Related Commands
MASK:TESt:STATE,
MASK:TESt:TIME,
MASK:TESt:THReshold
Arguments
<NR1> is an integer that specifies the number of w
aveforms to test. The
maximum waveform count that can be specified is 1,000,000.
INFInite indicates that there is no limit on t
he waveform count.
Examples
MASK:TESt:WAVEform 100000 specifies to test 100,000 waveforms.
MASK:TESt:TIME? might return INFINITE, indicating there is currently no
specified waveform limit.
MASK:USER:AMPLitude
This command specifies the nominal pulse amplitude, in volts, to be used for
a user-defined custom mask.
A series of examples showing how to use mask commands for typical tasks
is included in an appendix. (See page E-1, Mask/Limit Command Sequence
Examples.)
NOTE. To ensure that Mask commands are enabled, use the command
APPLication:TYPe LIMITMask.
Conditions
This command requires a DPO4LMT application module for 4000 series models,
and an MDO3LMT application module for MDO3000 series models.
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-397