Commands Listed in Alphabetical Order
Group
Mask
Syntax
MASK:USER:AMPLitude <NR3>
MASK:USER:AMPLitude?
Related Commands
MASK:CUSTom,
SAVe:MASK
Arguments
<NR3> is a floating point number that determines the nominal pulse amplitude, in
volts, of a user-defined custom mask.
Examples
MASK:USER:AMPLITUDE 2.4 sets the nominal pulse amplitude of a user-defined
custom mask to 2.4 volts.
MASK:USER:AMPLITUDE? might return
:MASK:USER:AMPLITUDE 100.0000E–3, indicating that the user mask nominal
pulse amplitude is set to 0.1 volts.
MASK:USER:HSCAle
This command specifies the nominal timing resolution, in time/division, tobe
used to draw a user-defined custom mask pulse shape.
A series of examples showing how to use mask commands for typical tasks
is included in an appendix. (See page E-1, Mask/Limit Command Sequence
Examples.)
Conditions
This command requires a DPO4LMT application module for 4000 series models,
andanMDO3LMTapplicati
on module for MDO3000 series models.
Group
Mask
Syntax
MASK:USER:HSCAle <NR3>
MASK:USER:HSCAle?
Related Commands
MASK:CUSTom,
SAVe:MASK
2-398 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual

Commands Listed in Alphabetical Order
Arguments
<NR3> isafloati
ng point value that specifies the timing resolution used to a draw
a user-defined custom mask pulse shape.
Examples
MASK:USER:HSCALE 200E–9 sets the nominal timing resolution used to draw the
mask pulse shape to 200 nanoseconds per division.
MASK:USER:HSCALE? might return :MASK:USER:HSCALE 40.0000E–9,
indicating that the nominal timing resolution used to draw the mask pulse shape
is set to 40 nanoseconds per division.
MASK:USER:HTRIGPOS
This command specifies the nominal trigger position (pulse leading edge), to be
used to draw a user-defined custom mask, as a fraction of the display width.
A series of examples showing how to use mask commands for typical tasks
is included in an appendix. (See page E-1, Mask/Limit Command Sequence
Examples.)
Conditions
This command requires a DPO4LMT application module for 4000 series models,
and an MDO3LMT application module for MDO3000 series models.
Group
Mask
Syntax
MASK:USER:HTRIGPOS <NR3>
MASK:USER:HTRIGPOS?
Related Commands
MASK:CUSTom,
SAVe:MASK
Arguments
<NR3> is
a floating point number in the range of 0.0 to 1.0 that sets the trigger
points as a fraction of the display width, referenced from the left edge of the
graticule.
0.0 represents the left edge.
Examples
MASK:USER:HTRIGPOS 0.6 sets the nominal trigger position of the mask to 0.6.
MASK:USER:HTRIGPOS? might return :MA SK:USER:HTRIGPOS 500.0000E–3,
indicating that the nominal trigger position of the mask is set to 0.5.
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-399

Commands Listed in Alphabetical Order
MASK:USER:LA
Bel
This command specifies a user-defined label for a custom mask.
Conditions
This command requires a DPO4LMT application module.
Group
Mask
Syntax
MASK:USER:LABel <QString>
MASK:USER:LABel?
Related Commands
MASK:CUSTom, MASK:USER:AMPLitude
Arguments
<QString> is a quoted string of up to 32 characters.
MASK:USER:RECOr
dlength
This command specifies the nominal record length to be used for pulse mask
testing with a user-defined custom mask.
A series of examples showing how to use mask commands for typical tasks
is included in an appendix. (See page E-1, Mask/Limit Command Sequence
Examples.)
Conditions
This command requires a DPO4LMT application module for 4000 series models,
and an MDO3LMT application module for MDO3000 series models.
Group
Mask
Syntax
MASK:USER:RECOrdlength <NR1>
MASK:USER:RECOrdlength?
Related Commands
MASK:CUSTom,
SAVe:MASK,
MASK:USER:TRIGTOSAMP
Arguments
<NR1> is an integer that sets the record length value to be used for pulse mask
testing of a user-
defined custom mask.
2-400 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual