MEASUREMENT:M
EAS3:DELAY:DIRECTION BACKWARDS
starts searching from
the end of the waveform record.
MEASUREMENT:
MEAS3:DELAY:DIRECTION?
might return
:MEASUREMENT:MEAS3:DELAY:DIRECTION BACKWARDS indicating
that the current search direction is backwards.
MEASUremen
t:MEAS<x>:DELay:EDGE<x>
This command specifies the slope of the edge used for the delay "from" or "to"
waveform when taking an immediate delay measurement. The waveform is
specified
by MEASUrement:MEAS<x>:SOUrce<x>.
Group
Measurement
Syntax
MEASUrement:MEAS<x>:DELay:EDGE<x> {FALL|RISe}
MEASUrement:MEAS<x>:DELay:EDGE<x>?
Arguments
<x> specifies which waveform to use, where <x> = 1 is the "from" waveform, and
<x> = 2 is the "to" waveform.
FALL specifies the falling edge.
RISe specifies the rising edge.
Examples
MEASUREMENT:MEAS1:DELAY:EDGE1 RISE specifies that the "from" waveform
rising edge be used for the immediate delay measurement.
MEASUREMENT:MEAS1:DELAY:EDGE1? returns either RISE or FALL.
MEASUrement:MEAS<x>:MAXimum? (Query Only)
Returns the maximum value found for this measurement since the last statistical
reset. Measurements are specified by x, which ranges from 1 through 8.
Group
Measurement
Syntax
MEASUrement:MEAS<x>:MAXimum?
Examples
MEASUREMENT:MEAS3:MAXIMUM? might return
:MEASUREMENT:MEAS3:MAXIMUM 4.18E-9.
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-435
Commands Listed in Alphabetical Order
MEASUrement:
MEAS<x>:MEAN? (Query Only)
Returns the mean value accumulated for this measurement since the last statistical
reset. Measurements are specified by x, which ranges from 1 through 8.
Group
Measurement
Syntax
MEASUrement:MEAS<x>:MEAN?
Examples
MEASUREMENT:MEAS1:MEAN? might return :MEASUREMENT:MEAS1:MEAN
514.71E-09.
MEASUrement:MEAS<x>:MINImum? (Query Only)
Returns the minimum value for this measurement since the last statistical reset.
Measurements are specified by <x>, which ranges from 1 through 8.
Group
Measurement
Syntax
MEASUrement:MEAS<x>:MINImum?
Examples
MEASUREMENT:MEAS1:MINIMUM? might return
:MEASUREMENT:MEAS1:MINIMUM 1.75E-09.
MEASUrement:MEAS<x>:SOUrce<x>
For SOURce1: This command specifies the source for all single channel
measurements. For delay or phase measurements, This command specifies the
waveform to measure "from". This is equiv
alent to setting the "From:" waveform
in the "Measure Delay" side menu or the "Measure Phase" side menu. SOUrce is
equivalent to SOURCE1.
For SOUrce2: This command specifies the waveform to measure "to" when taking
a delay measurement or phase measurement. This is equivalent to setting the"To:"
waveform in the "Measure Delay" side menu or the "Measure Phase" side menu.
Measurements are specified by <x>, which ranges from 1 to 8.
NOTE. If you do not specify a numeric sufx,thesourceisassumedtobeSOUrce1.
2-436 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual
Commands Listed in Alphabetical Order
Group
Measurement
Syntax
MEASUrement:MEAS<x>:SOUrce<x>
{CH1|CH2|CH3|CH4|MATH|R1|R2|R3|R4|D0|D1|D2|D3|D4|D5|D6|D7
|D8|D9|D10|D11|D12|D13|D14|D15|HIStogram|RF_AMPlitude|
RF_FREQuency|RF_PHASe}
MEASUrement:MEAS<x>:SOUrce<x>?
Related Commands
MEASUrement:MEAS<x>:TYPe
Arguments
CH1–4 is an analog channel to use as the source waveform.
MATH is the math waveform.
REF1–4 is a reference waveform to use as the source waveform.
D0–D15 is a digital channel to use as the source waveform.
(MSO/MDO4000/B models only as well as MDO3000 and MDO400C models
with option MDO3MSO or MDO4MSO installed.)
HIStogram indicates the histogram is the object to be measured. HIStogram only
applies to SOUrce1; it is not allowed on SOUrce2.
RF_AMPlitude|RF_FREQuency|RF_PHASe are the RF time domain traces.
(MDO4000/B/C models only.)
Examples
MEASUREMENT:MEAS4:SOURCE2 CH1 specifies CH1 as the delay "to" source
when making delay measurement.
MEASUREMENT:MEAS2:SOURCE2? might return
:MEASUREMENT:MEAS2:SOURCE2 MATH1 indicating that Math 1 is the
measurement 2 source.
MEASUrement:MEAS<x>:STATE
This command specifies whether the specified measurement slot is computed
and displayed. The measurement slot is specified by <x>, which ranges from
1 through 8.
For a measurement to display, you must have selected a source waveform
and defined the measurement you want to take and display. You select the
measurement using the MEASUrement:MEAS<x>:SOUrce<x> command.
You define the measurement type using the MEASUrement:MEAS<x>:TYPe
command.
Group
Measurement
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-437