MDO3000 Programmer Manual.pdf - 第477页
Commands Listed in Alphabetical Order Arguments <NR3> is a floati ng point number that specifies the low reference level, in volts. The default is 0.0 V . Examples MEASUREMENT:R EFLEVEL:ABSOLUTE :LOW 0.0 sets the low…

Commands Listed in Alphabetical Order
NOTE. this comm
and affects the associated reference level parameter for all
MEASurements:IMMed and the four periodic measurements.
Group
Measurement
Syntax
MEASUrement:REFLevel:ABSolute:HIGH <NR3>
MEASUrement:REFLevel:ABSolute:HIGH?
Related Commands
MEASUrement:REFLevel:METHod,
MEASUrement:IMMed:TYPe,
MEASUrement:MEAS<x>:TYPe
Arguments
<NR3> is a floating point number that specifies the high reference level, in volts.
The default is 0.0 V.
Examples
MEASUREMENT:REFLEVEL:ABSOLUTE:HIGH 1.71 sets the high reference level
to 1.71 V.
MEASUREMENT:REFLEVEL:ABSOLUTE:HIGH? might return
:MEASUREMENT:REFLEVEL:ABSOLUTE:HIGH 1.7100E+00 indicating that the
absolute high reference level is set to 1.71 V.
MEASUrement:REFLevel:ABSolute:LOW
This command specifies the low reference level, and is the lower reference level
when MEASUrement:REFLevel:METHod is set to Absolute.
NOTE. this command affects the associated reference level parameter for all
MEASurements:IMMed and the four periodic measurements.
Group
Measurement
Syntax
MEASUrement:REFLevel:ABSolute:LOW <NR3>
MEASUrement:REFLevel:ABSolute:LOW?
Related Commands
MEASUrement:REFLevel:METHod,
MEASUrement:IMMed:TYPe,
MEASUrement:MEAS<x>:TYPe
2-446 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual

Commands Listed in Alphabetical Order
Arguments
<NR3> is a floati
ng point number that specifies the low reference level, in volts.
The default is 0.0 V.
Examples
MEASUREMENT:REFLEVEL:ABSOLUTE:LOW 0.0 sets the low reference level
to 0.0 V.
MEASUREMENT:REFLEVEL:ABSOLUTE:LOW? might return
:MEASUREMENT:REFLEVEL:ABSOLUTE:LOW 0.0000E+00 indicating
that the absolute low reference level is set to 0.0 V.
MEASUrement:REFLevel:ABSolute:MID<x>
This command specifies an absolute mid reference level (the 50% reference level)
for the "to" waveform when taking a delay measurement, in volts. There are
two absolute mid reference levels; thus MID<x> can either be MID1 or MID2.
This command applies when MEASUrement:REFLevel:METHod has been set
to Absolute. This command is equivalent to setting the Reference Levels in the
MEASURE menu on the oscilloscope front panel.
NOTE. This command affects the results of period, frequency, delay, and all cyclic
measurements. This command affects the associated reference level parameter for
all MEASurements:IMMed and the four periodic measurements.
Group
Measurement
Syntax
MEASUrement:REFLevel:ABSolute:MID<x> <NR3>
MEASUrement:REFLevel:ABSolute:MID<x>?
Related Commands
MEASUrement:REFLevel:METHod
Arguments
<NR3> is a floating point number that specifies the mid reference level in volts.
Examples
MEASUrement:REFLevel:ABSolute:MID 0.5 sets the absolute MID1
reference level for the delay waveform to 0.5 V.
MEASUrement:REFLevel:ABSolute:MID? might return
:MEASUREMENT:REFLEVEL:ABSOLUTE:MID2 0.5000E+00 indicating that the
absolute MID2 reference level is set to 0.5 V.
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-447

Commands Listed in Alphabetical Order
MEASUrement:
REFLevel:METHod
Specifies or returns the reference level units used for measurement calculations.
NOTE. this command affects the associated reference level parameter
for all MEAS
urements:IMMed and the eight periodic measurements (4 in
MDO3000). To change the parameter for individual measurements, use the
MEASUrement:MEAS<x>:REFLevel commands.
Group
Measurement
Syntax
MEASUrement:REFLevel:METHod {ABSolute|PERCent}
MEASUrement:REFLevel:METHod?
Arguments
ABSolute specifies that the reference levels are set explicitly using the
MEASUrement:REFLevel:ABSolute commands. This method is useful when
precise values are required (for example, when
designing to published interface
specifications, such as RS-232-C).
PERCent specifies that the reference levels ar
e calculated as a percent
relative to HIGH and LOW. The percentages are defined using the
MEASUrement:REFLevel:PERCent commands.
Examples
MEASUREMENT:REFLEVEL:METHODABSOLUTE specifies that explicit user-defined
values are used for the reference levels.
MEASUREMENT:REFLEVEL:METHOD? might return
:MEASUREMENT:REFLEVEL:METHOD PERCENT indicating that the reference
level units used are calculated as a percent relative to HIGH and LOW.
MEASUrement:REFLevel:PERCent:HIGH
This command specifies the percent (where 100% is equal to HIGH) used to
calculate the high reference level when MEASUrement:REFLevel:METHod is set
to Percent. This command affects the results of rise and fall measurements.
NOTE. This command affects the associated reference level parameter for all
MEASurements:IMMed and the four periodic measurements.
Group
Measurement
2-448 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual