Commands Listed in A lphabetical Order Examples MEASUREMENT:R EFLEVEL:PERCENT: LOW 15 sets the high reference level to 15% of HIGH. MEASUREMENT: REFLEVEL:PERC ENT:LOW? might return :MEASUREMENT: REFLEVEL:PERCENT :LOW 10 …

100%1 / 1158
Commands Listed in Alphabetical Order
Syntax
MEASUrement:R
EFLevel:PERCent:HIGH <NR3>
MEASUrement:REFLevel:PERCent:HIGH?
Related Commands
MEASUrement:REFLevel:METHod,
MEASUrement:IMMed:TYPe,
MEASUrement:MEAS<x>:TYPe
Arguments
<NR3> is a floating point number that specifies the high reference level, ranging
from 0 to 100%. The default high reference level is 90%.
Examples
MEASUREMENT:REFLEVEL:PERCENT:HIGH95 sets the high reference level to
95% of HIGH.
MEASUREMENT:REFLEVEL:PERCENT:HIGH? might return
:MEASUREMENT:REFLEVEL:PERCENT:HIGH 90 indicating that the
percentage high reference level is set to 90% of HIGH.
MEASUrement:REFLevel:PERCent:LOW
This command specifies the percent (where 100% is equal to HIGH) used to
calculate the low reference level when MEASUrement:REFLevel:METHod is set
to Percent. This command affects the results of rise and fall measurements.
NOTE. this command affects the associated reference level parameter for a ll
MEASurements:IMMed and the periodic measurements.
Group
Measurement
Syntax
MEASUrement:REFLevel:PERCent:LOW <NR3>
MEASUrement:REFLevel:PERCent:LOW?
Related Commands
MEASUrement:REFLevel:METHod,
MEASUrement:IMMed:TYPe,
MEASUrement:MEAS<x>:TYPe
Arguments
<NR3> is a floating point number that speci
fies the low reference level, ranging
from 0 to 100%. The default low reference level is 10%.
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-449
Commands Listed in Alphabetical Order
Examples
MEASUREMENT:R
EFLEVEL:PERCENT:LOW 15
sets the high reference level to
15% of HIGH.
MEASUREMENT:
REFLEVEL:PERCENT:LOW?
might return
:MEASUREMENT:REFLEVEL:PERCENT:LOW 10 indicating that the
percentage high reference level is set to 10% of HIGH.
MEASUremen
t:REFLevel:PERCent:MID<x>
This command specifies the percent (where 100% is equal to HIGH) that is
used to calculate a mid reference level. There are two mid reference levels;
thus, MID
<x> can be either MID1 or MID2. This command applies when the
command MEASUrement:REFLevel:METHod has been set to Percent. This
command is equivalent to setting the Reference Levels in the MEASURE menu
on the oscilloscope front panel.
NOTE. This command affects the re sults of period, frequency, delay, a nd all cyclic
measurements. This command affects the associated reference level parameter for
all MEA
Surements:IMMed and the four periodic measurements.
Group
Measurement
Syntax
MEASUrement:REFLevel:PER
Cent:MID<x> <NR3>
MEASUrement:REFLevel:PERCent:MID<x>?
Related Commands
MEASUrement:REFLevel:METHod
Arguments
<NR3> is a floating point number that specifies the mid ref
erence level, ranging
from 0 to 100%. The default mid reference level is 50%.
Examples
MEASUrement:REFLevel:PERCent:MID 60 sets the MID1 reference level
to 60% of HIGH.
MEASUrement:REFLevel:PERCent:MID? might return
:MEASUREMENT:REFLEVEL:PERCENT:MID2 65 indicating that the
MID2 reference level has been set to 65% of HIGH.
MEASUrement:STATIstics (No Query Form)
Clears all of the statistics accumulated for all periodic measurement
s(MEAS1
through MEAS4).
2-450 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual
Commands Listed in Alphabetical Order
The query form r
eturns statistic settings.
Group
Measurement
Syntax
MEASUrement:STATIstics RESET
MEASUrement:STATIstics?
Arguments
RESET clears the measurements.
MEASUrement:STATIstics:MODe
Controls the operation and display of management statistics.
Group
Measurement
Syntax
MEASUrement:STATIstics:MODe {OFF|ALL}
MEASUrement:STATIstics:MODe?
Related Commands
MEASUrement:STATIstics
Arguments
OFF turns all measurements off. This is the default value.
AL turns on statistics and displays all statistics for each measurement.
Examples
MEASUREMENT:STATISTICS:MODE OFF turns statistic measurements off.
MEASUREMENT:STATISTICS:MODE? might return
:MEASUREMENT:STATISTICS:MODE ON indicating that measurement statistics
are turned on and all statistics are being displayed for each measurement.
MEASUrement:STATIstics:WEIghting
This command specifies the time constant for mean and standard deviation
statistical accumulations.
Group
Measurement
Syntax
MEASUrement:STATIstics:WEIghting <NR1>
MEASUrement:STATIstics:WEIghting?
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-451