Commands Listed in A lphabetical Order POWer:HARMon ics:RESults:P ASSFail? (Query Only) Returns the overall harmonics test result: P ASS, F AIL or NA. Conditions This command requires a DPO4PWR applic ation module for 40…

100%1 / 1158
Commands Listed in Alphabetical Order
Group
Power
Syntax
POWer:HARMonics:RESults:IEC:POHL?
Examples
POWER:HARMONICS:RESULTS:IEC:POHL? might return
:POWER:HARMONICS:RESULTS:IEC:POHL 0.5 indicating that the
IEC POHL is set to 0.5 A.
POWer:HARMonics:RESults:IEC:POWer? (Query Only)
Returns the IEC input power that is used to calculate limits.
Conditions
This command requires a DPO4PWR application module for 4000 series models,
or an MDO3PWR application module for MDO3000 series models.
Group
Power
Syntax
POWer:HARMonics:RESults:IEC:POWer?
Examples
POWER:HARMONICS:RESULTS:IEC:POWER? might return
:POWER:HARMONICS:RESULTS:IEC:POWER 5 indicating that the
IEC power is set to 5 W.
POWer:HARMonics:RESults:IEC:POWERFactor? (Query Only)
Returns the IEC power factor measurement.
Conditions
This command requires a DPO4PWR application module for 4000 series models,
or an MDO3PWR application module for MDO3000 series models.
Group
Power
Syntax
POWer:HARMonics:RESults:IEC:POWERFactor?
Examples
POWER:HARMONICS:RESULTS:IEC:POWERFACTOR? might return
:POWER:HARMONICS:RESULTS:IEC:POWERFACTOR 9.1900E+37 indicating
that the IEC power factor is set to 9.1900E+37.
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-483
Commands Listed in Alphabetical Order
POWer:HARMon
ics:RESults:PASSFail? (Query Only)
Returns the overall harmonics test result: PASS, FAIL or NA.
Conditions
This command requires a DPO4PWR application module for 4000 series models,
or an MDO3PWR application module for MDO3000 series models.
Group
Power
Syntax
POWer:HARMonics:RESults:PASSFail?
Examples
POWER:HARMONICS:RESULTS:PASSFAIL? might return
:POWER:HARMONICS:RESULTS:PASSFAIL PASS indicating that the
harmon
ics test passed.
POWer:HARMonics:RESults:RMS? (Query Only)
Returns the root mean square value of the harmonics source waveform.
Conditions
This command requires a DPO4PWR application module for 4000 series models,
or an MDO3PWR application module for MDO3000 series models.
Group
Power
Syntax
POWer:HARMonics:RESults:RMS?
Examples
POWER:HARMONICS:RESULTS:RMS? might return
:POWER:HARMONICS:RESULTS:RMS 1.0 indicating that the harmonics source
waveform RMS is set to 1.0.
POWer:HARMonics:RESults:SAVe (No Query Form)
Saves the harmonic results to the specified file in CSV format.
Conditions
This command requires a DPO4PWR application module for 4000 series models,
or an MDO3PWR application module for MDO3000 series models.
Group
Power
2-484 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual
Commands Listed in Alphabetical Order
Syntax
POWer:HARMoni
cs:RESults:SAVe <String>
POWer:HARMonics:RESults:THDF? (Query Only)
Returns the Total Harmonic Distortion (THD) in percentage, measured as a ratio
to the RMS value of the fundamental component of the source waveform.
Conditions
This command requires a DPO4PWR application module for 4000 series models,
or an MDO3PWR application module for MDO3000 series models.
Group
Power
Syntax
POWer:HARMonics:RESults:THDF?
Examples
POWER:HARMONICS:RESULTS:THDF? might return
:POWER:HARMONICS:RESULTS:THDF 40 indicating that the THDF is set to 40.
POWer:HARMonics:RESults:THDR? (Query Only)
Returns the Total Harmonic Distortion (THD) in percentage, measured as a ratio
to the RMS value of the source waveform.
Conditions
This command requires a DPO4PWR application module for 4000 series models,
or an MDO3PWR application module for MDO3000 series models.
Group
Power
Syntax
POWer:HARMonics:RESults:THDR?
Examples
POWER:HARMONICS:RESULTS:THDR? might return
:POWER:HARMONICS:RESULTS:THDR 40 indicating that the THDR is set to 40.
POWer:HARMonics:SOURce
This command specifies the source waveform for harmonics tests. The voltage
source w
aveform is specified using the POWer:VOLTAGESOurce command and
the current waveform is specified using the POWer:CURRENTSOurce command.
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-485