Commands Listed in Alphabetical Order
POWer:QUALit
y:DISplay:FREQuency
This command specifies the display state for the frequency readout.
Conditions
This command requires a DPO4PWR application module for 4000 series models,
or an MDO3PWR application module for MDO3000 series models.
Group
Power
Syntax
POWer:QUALity:DISplay:FREQuency {OFF|ON|0|1}
POWer:QUALity:DISplay:FREQuency?
Arguments
OFF or 0 turns off the frequency display.
ON or 1 turns on the frequency display.
Examples
POWER:QUALITY:DISPLAY:FREQUENCY 1 turns on the frequency readout
display.
POWER:QUALITY:DISPLAY:FREQUENCY? might return
:POWER:QUALITY:DISPLAY:FREQUENCY 0 indicating that the
frequency readout display is off.
POWer:QUALity:DISplay:ICRESTfactor
This command specifies the display state for the current crest factor readout.
Conditions
This command requires a DPO4PWR application module for 4000 series models,
or an MDO3PWR application module for MDO3000 series models.
Group
Power
Syntax
POWer:QUALity:DISplay:ICRESTfactor {OFF|ON|0|1}
POWer:QUALity:DISplay:ICRESTfactor?
Arguments
OFF or 0 turns off the current crest factor display.
ON or 1 turns on the current crest factor display.
Examples
POWER:QUALITY:DISPLAY:ICRESTFACTOR 1 turns on the current crest factor
display.
2-490 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual
Commands Listed in Alphabetical Order
POWER:QUALITY
:DISPLAY:ICRESTFACTOR?
might return
:POWER:QUALITY:DISPLAY:ICRESTFACTOR 0 indicating that the
current crest factor display is off.
POWer:QUALi
ty:DISplay:IRMS
This command specifies the display state for the rms current readout.
Conditions
This command requires a DPO4PWR application module for 4000 series models,
or an MDO3PWR application module for MDO3000 series models.
Group
Power
Syntax
POWer:QUALity:DISplay:IRMS {OFF|ON|0|1}
POWer:QUALity:DISplay:IRMS?
Arguments
OFF or 0 turns off the rms current display.
ON or 1 turns on the rms current display.
Examples
POWER:QUALITY:DISPLAY:IRMS 1 turns on the rms current display.
POWER:QUALITY:DISPLAY:IRMS? might return
:POWER:QUALITY:DISPLAY:IRMS 0 indicating that the rms current display
is off.
POWer:QUALity:DISplay:PHASEangle
This command specifies the display state for the phase angle readout.
Conditions
Th
is command requires a DPO4PWR application module for 4000 series models,
or an MDO3PWR application module for MDO3000 series models.
Group
Power
Syntax
POWer:QUALity:DISplay:PHASEangle {OFF|ON|0|1}
POWer:QUALity:DISplay:PHASEangle?
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-491
Commands Listed in Alphabetical Order
Arguments
OFF or 0 turns of
f the phase angle display.
ON or 1 turns on the phase angle display.
Examples
POWER:QUALITY:DISPLAY:PHASEANGLE 1 turns on the phase angle display.
POWER:QUALITY:DISPLAY:PHASEANGLE? might return
:POWER:QUALITY:DISPLAY:PHASEANGLE 0 indicating that the
phase angle display is off.
POWer:QUALity:DISplay:POWERFACtor
This command specifies the display state for the power factor readout.
Conditions
This command requires a DPO4PWR application module for 4000 series models,
or an MDO3PWR application module for MDO3000 series models.
Group
Power
Syntax
POWer:QUALity:DISplay:POWERFACtor {OFF|ON|0|1}
POWer:QUALity:DISplay:POWERFACtor?
Arguments
OFF or 0 tu
rns off the power factor display.
ON or 1 turns on the power factor display.
Examples
POWER:QUALITY:DISPLAY:POWERFACTOR 1 turns on the power factor display.
POWER:QUALITY:DISPLAY:POWERFACTOR? might return
:POWER:QUALITY:DISPLAY:POWERFACTOR 0 indicating that the
power factor display is off.
POWer:QUALity:DISplay:REACTpwr
This command specifies the display state for the reactor power readout.
Conditions
This command requires a DPO4PWR application module for 4000 series models,
or an MDO3PWR application module for MDO3000 series models.
Group
Power
2-492 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual