MDO3000 Programmer Manual.pdf - 第536页
Commands Listed in A lphabetical Order Syntax POWer:RIPPle: RESults:MEAN? Examples POWER:RIPPLE: RESULTS:MEAN? might return :POWER:RIPPLE :RESULTS:MEAN 1.0 indicating that the me an peak-to -peak amplitude is set to 1.0 …

Commands Listed in Alphabetical Order
POWer:RIPPle
:RESults:AMPLitude? (Query Only)
Returns the peak-to-peak ripple measurement.
Conditions
This command requires a DPO4PWR application module for 4000 series models,
or an MDO3PWR application module for MDO3000 series models.
Group
Power
Syntax
POWer:RIPPle:RESults:AMPLitude?
Examples
POWER:RIPPLE:RESULTS:AMPLITUDE? might return
:POWER:RIPPLE:RESULTS:AMPLITUDE 1 indicating that the peak-to-peak
amplitude is set to 1 V.
POWer:RIPPle:RESults:MAX? (Query Only)
Returns the maximum of the peak-to-peak ripple measurements.
Conditions
This command requires a DPO4PWR application module for 4000 series models,
or an MDO3PWR application module for MDO3000 series models.
Group
Power
Syntax
POWer:RIPPle:RESults:MAX?
Examples
POWER:RIPPLE:RESULTS:MAX? might return :POWER:RIPPLE:RESULTS:MAX
1.1
indicating that the maximum peak-to-peak amplitude is set to 1.1 V.
POWer:RIPPle:RESults:MEAN? (Query Only)
Returns the mean of the peak-to-peak ri
pple measurements.
Conditions
This command requires a DPO4PWR application module for 4000 series models,
or an MDO3PWR application module for MDO3000 series models.
Group
Power
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-505

Commands Listed in Alphabetical Order
Syntax
POWer:RIPPle:
RESults:MEAN?
Examples
POWER:RIPPLE:RESULTS:MEAN? might return
:POWER:RIPPLE:RESULTS:MEAN 1.0 indicating that the mean peak-to-peak
amplitude is set to 1.0 V.
POWer:RIPPle:RESults:MIN? (Query Only)
Returns the minimum of the peak-to-peak ripple measurement.
Conditions
This command requires a DPO4PWR application module for 4000 series models,
or an MDO3PWR application module for MDO3000 series models.
Group
Power
Syntax
POWer:RIPPle:RESults:MIN?
Examples
POWER:RIPPLE:RESULTS:MIN? might return :POWER:RIPPLE:R ESULTS:MIN
0.9
indicating that the minimum peak-to-peak amplitude is set to 0
.9 V.
POWer:RIPPle:RESults:STDdev? (Query Only)
Returns the standard deviation of the peak-to-peak ripple measurements.
Conditions
This command requires a DPO4PWR application module for 4000 series models,
or an MDO3PWR application module for MDO3000 series models.
Group
Power
Syntax
POWer:RIPPle:RESults:STDdev?
Examples
POWER:RIPPLE:RESULTS:STDDEV? might return
:POWER:RIPPLE:RESULTS:STDDEV 0.14 indicating that the standard
deviation of the peak-to-peak amplitude is set to 0.14.
2-506 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual

Commands Listed in Alphabetical Order
POWer:RIPPle
:SOUrce
This command specifies the source waveform for ripple tests. The voltage source
waveform is specified using the POWer:VOLTAGESOurce command and the
current wave
form is specified using the POWer:CURRENTSOurce command.
Conditions
This command requires a DPO4PWR application module for 4000 series models,
or an MDO3PWR application module for MDO3000 series models.
Group
Power
Syntax
POWer:RIPPle:SOUrce {VOLTage|CURRent}
Arguments
VOLTage specifies voltage source waveform for ripple tests.
CURRent specifies current source waveform for ripple tests.
Examples
POWER:RIPPLE:SOURCE CURRent sets the ripple source to Current.
POWER:RIPPLE:SOURCE? might return :POWER:RIPPLE:SOURCE VOLT
indicating that the ripple source is set to Voltage.
POWer:SOA:LINear:XMAX
This command specifies the user XMAX value for use in linear SOA calculations.
Conditions
This command requires a DPO4PWR application module for 4000 series models,
or an MDO3PWR application module for MDO3000 series models.
Group
Power
Syntax
POWer:SOA:LINear:XMAX <NR3>
POWer:SOA:LINear:XMAX?
Arguments
<NR3> is a floating point number that specifies the XMAX value used for linear
SOA calculations.
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-507