MDO3000 Programmer Manual.pdf - 第544页
Commands Listed in A lphabetical Order Syntax POWer:SOA:MAS K:NR_Pt? Examples POWER:SOA:MAS K:NR_PT? might return :POWER:SOA:MASK: NR_ PT 4 indicating that the number of mask points defined is 4. POWer:SOA:MASK:ST A TE Th…

Commands Listed in Alphabetical Order
Syntax
POWer:SOA:MAS
K:MAXVolts <NR3>
POWer:SOA:MASK:MAXVolts?
Arguments
<NR3> is a floating point number that specifies the maximum voltage applied
to SOA mask testing.
Examples
POWER:SOA:MASK:MAXVOLTS 200 sets the SOA mask testing voltage to 200 V.
POWER:SOA:MASK:MAXVOLTS ? might return :POWER:SO A:MASK:MAXVOLTS
300.0000
indicating that the SOA mask testing maximum voltage is set to 300 V.
POWer:SOA:MASK:MAXWatts
This command specifies the maximum power applied to SOA mask testing.
Conditions
This command requires a DPO4PWR application module for 4000 series models,
or an MDO3PWR application module for MDO3000 series models.
Group
Power
Syntax
POWer:SOA:MASK:MAXWatts <NR3>
POWer:SOA:MASK:MAXWatts?
Arguments
<NR3> is a floating point number that specifies the maximum power applied to
SOA mask testing.
Examples
POWER:SOA:MASK:MAXWATTS 1000 sets the SOA mask testing power to
1.0000E+3 W.
POWER:SOA:MASK:MAXWATTS ? might return :POWER:SO A:MASK:MAXWATTS
750.0000
indicating that the SOA mask testing maximum power is set to 750 W.
POWer:SOA:MASK:NR_Pt? (Query Only)
Returns the number of mask points defined.
Conditions
This command requires a DPO4PWR application module for 4000 series models,
or an MDO3PWR application module for MDO3000 series models.
Group
Power
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-513

Commands Listed in Alphabetical Order
Syntax
POWer:SOA:MAS
K:NR_Pt?
Examples
POWER:SOA:MASK:NR_PT? might return :POWER:SOA:MASK: NR_PT 4
indicating that the number of mask points defined is 4.
POWer:SOA:MASK:STATE
This command specifies the state of the mask for SOA calculations.
Conditions
This command requires a DPO4PWR application module for 4000 series models,
or an MDO3PWR application module for MDO3000 series models.
Group
Power
Syntax
POWer:SOA:MASK:STATE {OFF|LIMITS|POINTS}
POWer:SOA:MASK:STATE?
Arguments
OFF disables mask testing.
LIMITS enables mask testing based on limits specified using
POWer:SOA:MASK:MAXAmps, POWer:SOA:MASK:MAXVolts,and
POWer:SOA:MASK:MAXWatts commands
POINTS enables mask testing based on masks points defined.
Examples
POWER:SOA:MASK:STATEPOINTS enables mask testing power based on points
defined.
POWER:SOA:MASK:STATE? might return :POWER:SOA:MASK:STATE LIMITS
indicating that the mask testing is enabled based on limits.
POWer:SOA:MASK:STOPOnviol
This command specifies the enabled state of the mask stop on violation condition.
Conditions
This command requires a DPO4PWR application module for 4000 series models,
or an MDO3PWR application module for MDO3000 series models.
Group
Power
2-514 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual

Commands Listed in Alphabetical Order
Syntax
POWer:SOA:MAS
K:STOPOnviol {OFF|ON|0|1}
POWer:SOA:MASK:STOPOnviol?
Arguments
OFF or 0 enables mask stop on violations.
ON or 1 disables mask stop on violations.
Examples
POWER:SOA:MASK:STOPONVIOL 1 disables the state of the mask stop on
violation condition.
POWER:SOA:MASK:STOPONVIOL ? might return
:POWER:SOA:MASK:STOPONVIOL 0 indicating that the mask stop is enabled.
POWer:SOA:PLOTTYPe
This command specifies the Safe Operating Area (SOA) plot type.
Conditions
This command requires a DPO4PWR application module for 4000 series models,
or an MDO3PWR application module for MDO3000 series models.
Group
Power
Syntax
POWer:SOA:PLOTTYPe {LOG|LINear}
POWer:SOA:PLOTTYPe?
Arguments
LOG for logarithmic SOA plot type.
LINear for linear SOA plot type.
Examples
POWER:SOA:PLOTTYPE LINear sets the SOA plot type to Linear.
POWER:SOA:PLOTTYPE ? might return :POWER:SOA :PLOTTYPE LO G indicating
that the SOA plot type is set to Log.
POWer:SOA:RES
ult:FAILures:QTY? (Query Only)
Returns the number of failures in the test.
Conditions
This command requires a DPO4PWR application module for 4000 series models,
or an MDO3PWR application module for MDO3000 series models.
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-515