Commands Listed in A lphabetical Order POWer:SWLoss :TOT al:POWer:MEAN? (Query Only) Returns the mean total power loss. Conditions This command requires a DPO4PWR applic ation module for 4000 series models , or an MDO3PW…

100%1 / 1158
Commands Listed in Alphabetical Order
Syntax
POWer:SWLoss:
TOTal:ENERGY:MEAN?
Examples
POWER:SWLOSS:TOTAL:ENERGY:MEAN? might return
:POWER:SWLOSS:TOTAL:ENERGY:MEAN 1 indicating that the mean conduction
energy switching loss calculation is set to 1 J.
POWer:SWLoss:TOTal:ENERGY:MIN? (Query Only)
Returns the minimum total energy for the switching loss calculation.
Conditions
This command requires a DPO4PWR application module for 4000 series models,
or an MDO3PWR application module for MDO3000 series models.
Group
Power
Syntax
POWer:SWLoss:TOTal:ENERGY:MIN?
Examples
POWER:SWLOSS:TOTAL:ENERGY:MIN? might return
:POWER:SWLOSS:TOTAL:ENERGY:MIN 1 indicating that the minimum
conduction energy switching loss calculation is set to 1 J.
POWer:SWLoss:TOTal:POWer:MAX? (Query Only)
Returns the maximum total power loss.
Conditions
This command requires a DPO4PWR application module for 4000 series models,
or an MDO3PWR application module for MDO3000 series models.
Group
Power
Syntax
POWer:SWLoss:TOTal:POWer:MAX?
Examples
POWER:SWLOSS:TOTAL:POWER:MAX? might return
:POWER:SWLOSS:TOTAL:POWER:MAX 1 indicating that the maximum total
power switching loss calculation is set to 1 W.
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-533
Commands Listed in Alphabetical Order
POWer:SWLoss
:TOTal:POWer:MEAN? (Query Only)
Returns the mean total power loss.
Conditions
This command requires a DPO4PWR application module for 4000 series models,
or an MDO3PWR application module for MDO3000 series models.
Group
Power
Syntax
POWer:SWLoss:TOTal:POWer:MEAN?
Examples
POWER:SWLOSS:TOTAL:POWER:MEAN? might return
:POWER:SWLOSS:TOTAL:POWER:MEAN 1 indicating that the mean total power
switching loss calculation is set to 1 W.
POWer:SWLoss:TO Tal:POWer:MIN? (Query Only)
Returns the minimum total power loss.
Conditions
This command requires a DPO4PWR application module for 4000 series models,
or an MDO3PWR application module for MDO3000 series models.
Group
Power
Syntax
POWer:SWLoss:TOTal:POWer:MIN?
Examples
POWER:SWLOSS:TOTAL:POWER:MIN? might return
:POWER:SWLOSS:TOTAL:POWER:MIN 1 indicating that the minimum total
power switching loss calculation is set to 1 W.
POWer:SWLoss:VCEsat
This command specifies VCESAT value for use in switching loss calculations
when the conduction calculation method is VCESAT.
Conditions
This command requires a DPO4PWR application module for 4000 series models,
or an MDO3PWR application module for MDO3000 series models.
2-534 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual
Commands Listed in Alphabetical Order
Group
Power
Syntax
POWer:SWLoss:VCEsat <NR3>
POWer:SWLoss:VCEsat?
Arguments
<NR3> is a floating point number that specifies the VCEsat switching loss
calculation.
Examples
POWER:SWLOSS:VCESAT 5 sets the VCEsat value to 5.0000.
POWER:SWLOSS:VCESAT ? might return :P OWER:SWLOSS:VCEsat 2.0000
indicating that the VCEsat value is set to 2.0000.
POWer:TYPe
This command specifies the power application measurement type.
Conditions
This command requires a DPO4PWR application module for 4000 series models,
or an M
DO3PWR application module for MDO3000 series models.
Group
Power
Syntax
POWer:TYPe {NONe|QUALit y|SWITCHingloss|SOA
|HARMonics|RIPPle|MODULationanalysis|DESKew}
POWer:TYPe?
Arguments
NONe Use to set the measurement type to None.
QUALity Use the power quality functions to obtain measurements and statistics
about the general power quality in your test circuit
SWITCHingloss Use the switching loss functions to obtain the power loss and
energy loss across the acquired waveform, including turn-on loss, turn-off loss,
conduction loss, and total loss. Typically, use these functions to characterize
losses in power supply switching devices, as they switch on and off.
SOA Use the safe operating functions to obtain an X-Y display of the switching
device-under-test's voltage and current. Also use them to perform a mask test of
the X-Y signal relative to the graphical X-Y description of the device specification
table. The safe operating area is typically the voltage and current values that a
semiconductor can operate without damaging itself.
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-535