MDO3000 Programmer Manual.pdf - 第605页
Commands Listed in Alphabetical Order Group RF Syntax RF:RF_AMPlitu de:VERTical:SCAl e <NR3> RF:RF_AMPlitu de:VERTical:SCAl e? Arguments <NR3> is a floating point number that specifies the vertica l scale. Exam…

Commands Listed in Alphabetical Order
Syntax
RF:RF_AMPlitu
de:LABel <QString>
RF:RF_AMPlitude:LABel?
Related Commands
RF:RF_FREQuency:LABel, RF:RF_PHASe:LABel, RF:LABel
Arguments
<QString> is a quoted string of up to 32 characters.
RF:RF_AMPlitude:VERTical:POSition
This command specifies the vertical position of the RF Amplitude vs. Time trace.
The position value determines the vertical graticule location at which thetrace
is displayed. Increasing the position value of a waveform causes the waveform
to move up. Decreasing the position value causes the waveform to move down.
The minimum is -50 divisions and the maximum is 50 divisions with a resolution
of 0.02 divisions.
Conditions
MDO4000/B/C models only.
Group
RF
Syntax
RF:RF_AMPlitude:VERTical:POSition <NR3>
RF:RF_AMPlitude:VERTical:POSition?
Arguments
<NR3> is a floating point number.
Examples
RF:RF_AMPlitude:VERTical:POSition —1 sets the vertical position of the
RF Amplitude vs. Time trace to -1 division.
RF:RF_AMPlitude:VERTical:POSition? might return 0.0E+0, indicating a
vertical position of 0.
RF:RF_AMPlitude:VERTical:SCAle
This command specifies the vertical scale for the RF Amplitude vs. Time trace.
For a signal with constant amplitude, increasing the scale causes the waveform to
be displayed smaller. Decreasing the scale causes the waveform to be displayed
larger.
Conditions
MDO4000/B/C models only.
2-574 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual

Commands Listed in Alphabetical Order
Group
RF
Syntax
RF:RF_AMPlitude:VERTical:SCAle <NR3>
RF:RF_AMPlitude:VERTical:SCAle?
Arguments
<NR3> is a floating point number that specifies the vertical scale.
Examples
RF:RF_AMPlitude:VERTical:SCAle 50E-3 sets the vertical scale to
50mV/division.
RF:RF_AMPlitude:VERTical:SCAle? might return 10.0000E-3, indicating
that the vertical scale has been set to 10mV/division.
RF:RF_AVErage:COUNt? (Query Only)
This query returns the number of RF traces that have been accumulated to create
the RF Average frequency domain trace.
Conditions
MDO3000, MDO4000/B, or MDO4000C models with option SA3 or SA6 only.
Group
RF
Syntax
RF:RF_AVErage:COUNt?
Returns
This command returns an integer that ranges from 0 to the number of acquisitions
specified in the command RF:RF_AVErage:NUMAVg.
Examples
RF:RF_AVErage:COUNt? might return 15, indicating that 15 traces have been
accumulated to create the RF Average trace.
RF:RF_AVErage:NUMAVg
This command specifies the number of acquisitions to be used when creating the
RF Average frequency domain trace, which displays the average of values from
multiple acquisitions at each trace point. The range is 2 – 512, in exponential
increments.
Conditions
MDO3000, MDO4000/B, or MDO4000C models with option SA3 or SA6 only.
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-575

Commands Listed in Alphabetical Order
Group
RF
Syntax
RF:RF_AVErage:NUMAVg <NR1>
RF:RF_AVErage:NUMAVg?
Related Commands
RF:RF_AVErage:COUNt?
Arguments
<NR1> is an integer.
Examples
RF:RF_AVErage:NUMAVg 100 sets the number of acquisitions to be used for the
RF average trace to 128 (the closest power of 2 increment).
RF:RF_AVErage:NUMAVg? might return 16, indicating that the number of
averages for the RF average trace has been set to 16.
RF:RF_FREQuency:LABel
This command specifies the label for the RF Frequency vs. Time trace.
Conditions
MDO4000/B/C models only.
Group
RF
Syntax
RF:RF_FREQuency:LABel <QString>
RF:RF_FREQuency:LABel?
Related Commands
RF:RF_AMPlitude:LABel, RF:RF_PHASe:LABel, RF:LABel
Arguments
<QString> is a quoted string of up to 32 characters.
RF:RF_FREQuency:VERTical:POSition
This command specifies the vertical position for the RF Frequency vs. Time trace.
The position value determines the vertical graticule location at which thetrace
is displayed. Increasing the position value of a waveform causes the waveform
to move up. Decreasing the position value causes the waveform to move down.
The minimum is -50 divisions and the maximum is 50 divisions with a resolution
of 0.02 divisions.
2-576 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual