MDO3000 Programmer Manual.pdf - 第634页

Commands Listed in A lphabetical Order files or a .csv ex tension for comma-separated value format files. The fileformat is specified using the SA V e:W A VEform:FILEFormat:RF_BB_IQ command. Examples SAVE:WAVEFORM CH1,REF1 s…

100%1 / 1158
Commands Listed in Alphabetical Order
domain wavefor
ms. Waveform sources with record lengths greater than 131072
points saved to EMEM are decimated to fit in the 131072 point edit memory.
D0-D15 -
These digital waveform sources can only be saved to CSV files and, for
instrument
s with the arbitrary waveform feature, to EMEM.
EMEM (instruments with the arbitrary waveform feature) - EMEM is the arbitrary
waveform ed
it memory. EMEM can be saved to a CSV file or to one of the
4 internal arbitrary waveform storage locations (ARB1-4).
ARB1-4 is o
ne of the internal arbitrary waveform storage locations. These sources
can only be saved to ARB1-4 or to a CSV file. ARB1-4 can be copied to another
of the internal arbitrary waveform storage locations: e.g. :SAVe:WAVEform
ARB1,ARB2. To recall from an arbitrary storage location to EMEM, use the
:RECAll:WAVEform command.
RF_NORMal, RF_AVErage, RF_MAXHold, RF_MINHold - These sources can
be saved to REF1-4, an ISF or CSV file
RF_BB_IQ specifies to save baseband I & Q data. This file format is
automatically saved using a .TIQ file extension. You can import .TIQ files into
Tektronix SignalVu-PC software (PC based), SignalVu software (oscilloscope
based
), RSAVu software (PC based) or into a Tektronix real-time spectrum
analyzer for pulse analysis and demodulation analysis.
ALL -
Specifies all displayed waveforms subject to the restrictions indicated inthe
note below for saving ALL waveforms.
<de
stination> can be one of:
REF1-4 - For saving a single scope waveform to one of the 4 internal reference
wa
veform memory storage locations (see descriptions for source waveforms
above for which waveforms can be saved to REF1-4).
E
MEM
- For instruments with the arbitrary waveform feature, scope waveforms
(CH1-4, MATH, REF1-4, D0-D15) can be saved to arbitrary waveform edit
memory. MATH and REF1-4 waveforms can only be time domain.
ARB1-ARB4 - For instruments with the arbitrary waveform feature, EMEM
can be saved to one of the 4 internal arbitrary waveform memorylocations.
Additionally, ARB1-4 can be copied to another of the internal arbitrary waveform
memory locations: e.g. :SAVe:WAVEform ARB1,ARB2
<file p ath> - Specifies the destination file path to save the specified
waveform(s). The file path is a quoted string that specifies the file name and
path. Specify the file path using the form <drive>:<dir>/<filename>. <drive> and
one or more <dir>s are optional. If you do not specify them, the instrument will
write the waveform to the current working directory. <filename> stands for a
filename of up to 125 characters, followed by a period (".") and the three-character
extension. Waveform files should have a .isf extension for internal save format
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-603
Commands Listed in Alphabetical Order
files or a .csv ex
tension for comma-separated value format files. The fileformat is
specified using the SAVe:WAVEform:FILEFormat:RF_BB_IQ command.
Examples
SAVE:WAVEFORM CH1,REF1 saves the CH1 waveform to reference memory
location 1.
SAVE:WAVEFORM ALL,"E:/ test_folder/test1_all.csv"
saves all displayed waveforms (excluding serial bus waveforms) to
E:/test_folder/test1_all.csv when the file format is set to
SPREADSHEET.
SAVe:WAVEform ALL,"E:/ test_folder/test1" saves all displayed
waveforms (excluding serial bus waveforms and digital channels for MSO/MDO
models) to individual files named
E:/test_folder/test1_<wfm>.isf (for
example
test1_CH1.isf) when the file format is set to INTERNAL.
SAVe:WAVEform:FILEFormat
This command specifies the file format to be used when saving waveforms either
an internal format, .ISF, or an external comma-delimited spreadsheet format,
.CSV, that includes waveform header and timing information.
NOTE. If the waveform source is RF_BB_IQ (baseband I&Q), then the
SAVe:WAVEf orm:FILEFormat
is not applicable. Use the command
SAVe:WAVEf orm:FILEFormat:RF_BB_IQ
instead.
Group
Save and Recall
Syntax
SAVe:WAVEform:FILEFormat {INTERNal|SPREADSheet}
SAVe:WAVEform:FILEFormat?
Related Commands
CURVe,
DATa,
DATa:STARt,
DATa:STOP,
SAVe:WAVEform,
WFMInpre:NR_Pt,
WFMOutpre:NR_Pt?
2-604 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual
Commands Listed in Alphabetical Order
Arguments
INTERNal speci
fies to save waveforms using an internal format. The file name
should be specified with .ISF filename extension. These files can be recalled as
reference waveforms.
SPREADSheet specifies that waveform data is saved in a format that contains
comma delimited values. These waveform data files should be named using the
.CSV filename extension. Saving waveforms in .CSV format enables spreadsheet
programs to import the data.
Examples
SAVE:WAVEFORM:FILEFORMAT INTERNAL specifies that the internal .ISV file
format is the format specified for saved waveforms.
SAVE:WAVEFORM:FILEFORMAT? might return :SAVE:WAVEFORM:FILEFORMAT
SPREADSHEET
indicating that waveforms are saved using the comma-delimited
.CSV format, enabling them to be imported by spreadsheet programs.
SAVe:WAVEform:FILEFormat:RF_BB_IQ
This command specifies the file format for saving the RF baseband I & Q data.
The default format is TIQ. See the description of the SAVe:WAVEform command
for detailed information about these formats.
Group
Save and Recall
Syntax
SAVe:WAVEform:FILEFormat:RF_BB_IQ {TIQ|MATLAB}
SAVe:WAVEform:FILEFormat:RF_BB_IQ?
Arguments
TIQ specifies to use the TIQ file format for saving the RF baseband I & Q data.
MATLAB specifies to use the MATLAB file format for saving the RF baseband I
&Qdata.
Examples
SAVE:WAVEFORM:FILEFORMAT:RF_BB_IQ TIQ specifies the the TIQ file format
for saving the RF baseband I & Q data.
SAVE:WAVEFORM:FILEFORMAT:RF_BB_IQ? might return
:SAVE:WAVEFORM:FILEFORMAT MATLAB indicating that MATLAB is
the currently selected file format for saving the RF baseband I & Q data.
SAVe:WAVEform:GATIng
Specifies whether save waveform operations should save the entire waveform
(NONe) or a specified portion of the waveform.
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-605