SEARCH:SEARC
H<x>:TRIGger:A:BUS:B<x>:MIL1553B:STATus:ADDRess:QUALifier
When the MIL-STD-1553 bus search condition is set to STATus, this command
specifies the qualifier to be used with the address field.
SEARCH<x> is the search
number, whic
hisalways1,and
B<x>
Conditions
This command requires a DPO4AERO application module for DPO/MSO4000B
and MDO4000/B/C models or MDO3AERO application module for
MDO3000 models.
Group
Search
Syntax
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:MIL1553B:STATus:ADDRess:
QUALifier {LESSthan|MOR Ethan|EQual|UNEQual|LESSEQual|
MOREEQual|INrange| OUTrange}
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:MIL1553B:STATus:ADDRess:
QUALifier?
Related Commands
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:MIL1553B:CONDition
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:MIL1553B:STATus:ADDRess:
QUALifier
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:MIL1553B:STATus:ADDRess:
HIVALue
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:MIL1553B:STATus:ADDRess:
VA L u e
Arguments
LESSthan sets the Status Address qualifier to less than.
MOREthan sets the Status Address qualifier to greater than.
EQual sets the Status Address qualifier to equal.
UNEQual sets the Status Address qualifier to not equal.
LESSEQual sets the Status Address qualifier to less than or equal.
MOREEQual sets the Status Address qualifier to greater than or equal.
INrange sets the Status Address qualifier to in range.
OUTrange sets the Status Address qualifier to out of range.
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-683

Commands Listed in Alphabetical Order
Examples
SEARCH:SEARCH
1:TRIGger:A:BUS:B1:MIL1553B:STATus:ADDRess:
QUALifier MOREthan
sets the qualifier to be used with the address field to
greater than.
SEARCH:SEARCH1:TRIGger:A:BUS:B1:MIL1553B:STATus:ADDRess:
QUALifier?
might return MORETHAN.
SEARCH:SEA
RCH<x>:TRIGger:A:BUS:B<x>:MIL1553B:STATus:ADDRess:VALue
When the MIL-STD-1553 bus search condition is set to STATus,andthe
qualifier is set to
LESSthan, MOREthan, EQual, UNEQual, LESSEQual
or MOREEQ
ual
, this command specifies the value of the 5–bit remote terminal
address to be used in the search. When the MIL-STD-1553 bus search condition
is set to
STATus, and the qualifier is set to INrange or OUTrange, this command
specifies the lower limit of the range. (Use the command SEARCH:SEARCH<x>:
TRIGger:A:BUS:B<x>:MIL1553B:STATus:ADDRess:HIVALue to specify the
upper limit of the range.) The default is all X’s (don’t care).
SEARCH<x> is the
search
number, which is always 1, and
B<x>
Conditions
This command requires a DPO4AERO application module for DPO/MSO4000B
and MDO4000/B/C models or MDO3AERO application module for
MDO3000 models.
Group
Search
Syntax
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:MIL1553B:STATus:ADDRess:
VALue < QString>
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:MIL1553B:STATus:ADDRess:
VALue?
Related Commands
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:MIL1553B:CONDition
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:MIL1553B:STATus:ADDRess:
QUALifier
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:MIL1553B:STATus:ADDRess:
HIVALue
Arguments
QString is a quoted string of up to 5 characters, where the allowable characters are
0, 1, and X.
Examples
SEARCH:SEARCH1:TRIGger:A:BUS:B1:MIL1553B:STATus:ADDRess:
VALue " 01000"
sets the value of the 5–bit remote terminal address to be used in
2-684 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual

Commands Listed in Alphabetical Order
the search, whe
n the qualifier has been set to
LESSthan, MOREthan, EQual,
UNEQual, LESSEQual
or MOREEQual.
SEARCH:SEARC
H1:TRIGger:A:BUS:B1:MIL1553B:STATus:ADDRess:
VALue " 01000"
sets the lower limit of the range to 01000, if the qualifier has
been set to
INrange or OUTrange.
SEARCH:SEARCH1:TRIGger:A:BUS:B1:MIL1553B:STATus:ADDRess:
VALue?
might return “01000”.
SEARCH:SE
ARCH<x>:TRIGger:A:BUS:B<x>:MIL1553B:STATus:BIT:BCR
When the MIL-STD-1553 bus search condition is set to STATus, this command
specifies the status word broadcast command received (BCR) bit value (bit 15)
to be use
dinthesearch.
SEARCH<x> is the search number, which is always 1,
and
B<x>
Conditions
This command requires a DPO4AERO application module for DPO/MSO4000B
and MDO4000/B/C models or MDO3AERO application module for
MDO3000 models.
Group
Search
Syntax
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:MIL1553B:STATus:BIT:BCR
{0|1|X|ZERo|ONE|NOCARE|OFF|ON}
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:MIL1553B:STATus:BIT:BCR?
Related Commands
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:MIL1553B:CONDition
Arguments
0
1
X
sets the value to X (“don’t care”) which is the default.
ZERO s
ets the value to 0.
ONE sets the value to 1.
NOCARE sets the value to X (“don’t care”) which is the default.
OFF sets the value to 0.
ON sets the value to 1.