MDO3000 Programmer Manual.pdf - 第735页

Commands Listed in Alphabetical Order SEARCH:SEARC H<x>:TRIGger:A:BUS:B<x>:USB:DA T a:TYPe This command specifies the data typ e to be used in a USB s earch, when the search condition is set to DATAPacket. SEA…

100%1 / 1158
Commands Listed in Alphabetical Order
Conditions
This command re
quires the DPO4USB application module to be installed, except
for MDO3000 models, which require the MDO3USB. Note that 1 GHz bandwidth
models are recommended for high-speed (HS) USB.
Group
Search
Syntax
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:USB:DATa:OFFSet <NR1>
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:USB:DATa:OFFSet?
Arguments
<NR1> is number in the range 0 to 1024.
Examples
SEARCH:SEARCH1:TRIGGER:A:BUS:B1:USB:DATA:OFFSET 36 sets the data
offset to 36.
SEARCH:SEARCH1:TRIGGER:A:BUS:B1:USB:DATA:OFFSET? might return
:SEARCH:SEARCH1:TRIGGER:A:BUS:B1:USB:DATA:OFFSET 0, indicating
thatthedataoffsetisthedefaultvalue,0.
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:USB:DATa:SIZe
This command specifies the number of contiguous data bytes to search for in a
USB search. The minimum and default values are 1 and the maximum value is
16.
SEARCH<x> is the search number, which is always 1, and B<x>
Conditions
This command requires the DPO4USB application module to be installed, except
for MDO3000 models, which require the MDO3USB. Note that 1 GHz bandwidth
models are recommended for high-speed (HS) USB.
Group
Search
Syntax
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:USB:DATa:SIZe <NR1>
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:USB:DATa:SIZe?
Arguments
<NR1> is number in the range 1 to 16.
Examples
SEARCH:SEARCH1:TRIGGER:A:BUS:B1:USB:DATA:SIZE 4 sets the
oscilloscope to search on four contiguous data bytes.
SEARCH:SEARCH1:TRIGGER:A:BUS:B1:USB:DATA:SIZE? might return ,
:SEARCH:SEARCH1:TRIGGER:A:BUS:B1:USB:DATA:SIZE 6 indicating that
the oscilloscope will search on 6 contiguous data bytes.
2-704 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual
Commands Listed in Alphabetical Order
SEARCH:SEARC
H<x>:TRIGger:A:BUS:B<x>:USB:DATa:TYPe
This command specifies the data type to be used in a USB search, when the
search condition is set to
DATAPacket. SEARCH<x> is the search number, which
is always 1, a
nd
B<x>
Conditions
This command requires the DPO4USB application module to be installed, except
for MDO3000 models, which require the MDO3USB. Note that 1 GHz bandwidth
models are recommended for high-speed (HS) USB.
Group
Search
Syntax
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:USB:DATa:TYPe
{ANY|DATA0|DATA1}
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:USB:DATa:TYPe?
Related Commands
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:USB:CONDition
Arguments
ANY indicates either a DATA0 or DATA1 data packet type.
DATA0 indicates a DATA0 data packet type.
DATA1 indicates a DATA1 data packet type.
Examples
SEARCH:SEARCH1:TRIGGER:A:BUS:B1:USB:DATA:TYPE DATA0 sets the
oscilloscope to search on a DATA0 data packet type.
SEARCH:SEARCH1:TRIGGER:A:BUS:B1:USB:DATA:TYPE? might return
:SEARCH:SEARCH1:TRIGGER:A:BUS:B1:USB:DATA:TYPE DATA1, indicating
that the data packet type is set to DATA1.
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:USB:DATa:VALue
This command specifies the lower limit value for the inside-of-range and
outside-of-range qualifiers to be used in a USB search, when the search condition
is set to
DATAPacket. SEARCH<x> is the search number, which is always 1,
and
B<x>
Conditions
This command requires the DPO4USB application module to be installed, except
for MDO3000 models, which require the MDO3USB. Note that 1 GHz bandwidth
models are recommended for high-speed (HS) USB.
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-705
Commands Listed in Alphabetical Order
Group
Search
Syntax
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:USB:DATa:VALue <QString>
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:USB:DATa:VALue?
Related Commands
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:USB:CONDition
Arguments
<QString> within the range 00000000 to 11111111 (00 hex to FF hex).
Examples
SEARCH:SEARCH1:TRIGGER:A:BUS:B1:USB:DATA:VALUE “00001000 sets
the binary address to 00001000 (08 hex).
SEARCH:SEARCH1:TRIGGER:A:BUS:B1:USB:DATA:VALUE? might return
:SEARCH:SEARCH1:TRIGGER:A:BUS:B1:USB:DATA:VALUE “01000000”,
which indicates that the binary address is 0100000 (40 hex).
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:USB:ENDPoint:VALue
This command specifies the endpoint value to be used for a USB search.
SEARCH<x> is the search number, which is always 1, and B<x>
Conditions
This command requires the DPO4USB application module to be installed, except
for MDO3000 models, which require the MDO3USB. Note that 1 GHz bandwidth
models are recommended for high-speed (HS) USB.
Group
Search
Syntax
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:USB:ENDPoint:VALue
<QString>
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:USB:ENDPoint:VALue?
Related Commands
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:USB:CONDition
Arguments
<QString> within the range 0000 to 1111 (00 hex to 0F hex).
Examples
SEARCH:SEARCH1:TRIGGER:A:BUS:B1:USB:ENDPOINT:VALUE “1000 sets
the binary address to 1000 (08 hex).
2-706 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual