MDO3000 Programmer Manual.pdf - 第785页

Commands Listed in Alphabetical Order Group V ertical Syntax SELect:{BUS<x >|B<x>} {OFF| ON|0|1} SELect:{BUS<x >|B<x>}? Arguments ON or 1 turns the specified waveform display on. OFF or 0 turns the…

100%1 / 1158
Commands Listed in Alphabetical Order
End—endvaluei
n percent of record
Focus focus value in percent of record
X value horizontal value of the focus point in the mark’s waveform units
Y value vertical value of the focus point in the mark’s waveform units
Delta X value difference from the Reference Marker.
Delta Y value difference from the Reference Marker.
If no user t
ime marks have been set, the query returns
NONe,0.0E0,0.0E0,0.0E0,0.0E0,0.0E0,0.0E0,0.0E0.
Examples
SEARCH:SPECTral:LIST? might return
RF_NORMAL,50.2000,50.2000,50.2000,100.0000E+6,3.6624,0.0E+0,0.0E+0;
RF_NORMAL,70.2000,70.20
00,70.2000,200.0000E+6,-45.7444,0.0E+0,0.0E+0;
RF_NORMAL,90.2000,90.2000,90.2000,300.0000E+ 6,-43.6428,0.0E+0,0.0E+0
This indicates that the markers automatically mark three peaks on the RF Normal
trace. Each marker is delineated by a semicolon. The values that follow each
RF_NORMAL record are Start, End, Focus, X value, Y value, Delta X value,
Delta Y value, respectively.
SELect? (Query Only)
Queries which waveforms are displayed.
Group
Vertical
Syntax
SELect?
Examples
SELECT?? might return the following
:SELECT:BUS1 0;BUS2 0;CH1 1;CH2 0;CH3 0;CH4 0;MAT H 0;REF1
0;REF2 0;REF3 0;REF4 0;CONTROL CH1
SELect:{BUS<x>|B<x>}
This command turns on and off the display of the waveform for <x>, where x is
the bus number 1-4, 1-3 for the MDO4000C and 1-2 for the MDO3000. The
query returns whether the channel is on or off but does not indicate whether it is
the specified waveform.
2-754 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual
Commands Listed in Alphabetical Order
Group
Vertical
Syntax
SELect:{BUS<x>|B<x>} {OFF|ON|0|1}
SELect:{BUS<x>|B<x>}?
Arguments
ON or 1 turns the specified waveform display on.
OFF or 0 turns the specified waveform display off.
SELect:CH<x>
Turns the display of the channel <x> waveform on or off, where <x > is the
channel number, which can be 1–4. This command also resets the acquisition.
The query returns whether the channel is on or off but does not indicate whether
it is the specified waveform.
Group
Vertical
Syntax
SELect:CH<x> {ON|OFF|1| 0}
SELect:CH<x>?
Arguments
ON or 1 turns on the display of the specified waveform. This waveform also
becomes the selected waveform.
OFF or 0 turns off the display of the specified waveform.
Examples
SELECT:CH2 ON turns the channel 2 waveform display on, and selects channel 2.
SELECT:CH1? might return :SELECT:CH1 1 indicating that channel 1 is being
displayed.
SELect:CONTROl
This command selects a waveform to be used with channel-related commands
such as the cursor commands. In addition to selecting the waveform, it turnson
the display of the waveform, if it wasn’t on already.
Group
Vertical
Syntax
SELect:CONTROl {CH1|CH2|CH3|CH4
|MATH|REF1|REF2|REF3|REF4|BUS1|BUS2|BUS3|BUS4|D0|D1|D2|D3|D4
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-755
Commands Listed in Alphabetical Order
|D5|D6|D7|D8|
D9|D10|D11|D12|D13|D14|D15|RF_AMPlitude|RF_FREQuency
|RF_PHASe|RF_NORMal|RF_AVErage| RF_MAXHold|RF_MINHold}
SELect:CONTR
Ol?
Arguments
CH1–4 specifies a channel waveform as the waveform affected by the front-panel
controls.
MATH specifies the math waveform as the waveform that is affected by the
front-panel controls.
BUS1–4 specifies a bus waveform as the waveform affected by the front-panel
controls.
D0–D15 specifies a digital waveform as the waveform affected by the front-panel
controls. (MSO/MDO4000/B models only as well as MDO3000 and MDO400C
models with option MDO3MSO or MDO4MSO installed.)
|
RF_AMPlitude|RF_FREQuency|RF_PHASe (MDO4000/B/C models only)
and
RF_NORMal|RF_AVErage|RF_MAXHold| RF_MINHold specify an RF
trace as the waveform affected by the front-panel controls.
Returns
NONE if all the channels are turned off. NONE is ignored on input.
Examples
SELECT:CONTR
OL CH2
resets acquisition displays on channel 2, and causes the
selected waveform to be the implied object of waveform commands.
SELECT:CONT
ROL?
might return :SELECT:CONTROL MATH indicating that math
is the implied object of waveform commands.
SELect:DAll (No Query Form)
This command turns on or off all digital channels (D0 D15).
NOTE. This command applies only to MSO/MDO4000/B models only as well
as MDO3000 and MDO400C models with option MDO3MSO or MDO4MSO
installe
d.
Group
Vertical
Syntax
SELect:DAll {ON|OFF|0| 1}
2-756 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual