MDO3000 Programmer Manual.pdf - 第800页
Commands Listed in A lphabetical Order Syntax TRIGger FORCe TRIGger? Arguments FORCe creates a trigger event. If TRIGger:ST A TE is set to READy , the acquisition will complete. Otherwise, th is command will be ignored. …

Commands Listed in Alphabetical Order
Arguments
<QString> is a t
ime in the form "hh:mm:ss" where hh refers to a two-digit hour
number, mm refers to a two-digit minute number from 00 to 59, and ss refers to a
two-digit second number from 00 to 59.
Examples
TIME "14:00:00" sets the time to exactly 2:00 p.m.
TIME? might return :TIME "14:05:17” indicating the current time is set to 2:05
p.m. and 17 seconds.
TOTaluptime? (Query Only)
This query returns the total number of hours that the oscilloscope has been
powered on since the nonvolatile memory was last programmed (usually sincethe
initial manufacturing process).
Group
Miscellaneous
Syntax
TOTaluptime?
*TRG (No Query Form)
Performs a group execute trigger on commands defined by *DDT.
Group
Miscellaneous
Syntax
*TRG
Related Commands
*DDT
Examples
*TRG immediately executes all commands that have been defined by *DDT.
TRIGger (No Query Form)
Forces a trigger event to occur.
Group
Trigger
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-769

Commands Listed in Alphabetical Order
Syntax
TRIGger FORCe
TRIGger?
Arguments
FORCe creates a trigger event. If TRIGger:STATE is set to READy, the acquisition
will complete. Otherwise, this command will be ignored.
Examples
TRIGGER FORCE forces a trigger event to occur.
TRIGger:A
Sets the A trigger level automatically to 50% of the range of the minimum and
maximum values of the trigger input signal. The query returns current A trigger
parameters. The trigger level is the voltage threshold through which the trigger
source signal must pass to generate a trigger event. This command w
orks for the
following cases: Edge Trigger (when source is not AC line or power), Logic
Trigger (when Clock Source is not Off or Logic Pattern is Don't Care), and Pulse
Width Trigger.
Group
Trigger
Syntax
TRIGger:A SETLevel
TRIGger:A?
Related Commands
TRIGger:A:EDGE?,
TRIGger:A:LOGIc?,
Arguments
SETLevel sets the A trigger level to 50% of the range of the minimum and
maximum values of the trigger input signal.
Examples
TRIGGER:A SETLEVEL sets the A trigger level to 50% of the range of the
minimum and maximum values of the trigger input signal.
TRIGGER:A? might return a long response with A trigger parameters,
some of which could be as follows:
:TRIGGER:A:MODE AUTO;TYP E
EDGE;LEVEL 20.0000E-3; LEVEL:CH1 20.0000E-3;CH2 0.0000
;
CH3 0.0 000;CH4 0.0000;:TRIGGER:A:UPPERTHRESHOLD:CH1
1.4000;CH2 800.0000E-3 ;CH3 8 00 .0000E-3;CH4
800.0000E-3;:TRIGGER:A:LOWERTHRESHOLD:CH1 20.0000E-3;CH2
0.0000;CH3 0.0000;CH4 0.0000;:TRIG GER:A:HOLDOFF:TIME
20.0000E-9;:TRIGGER:A:EDGE:SOURCE CH1;COUPLING
DC;SLOPE RISE;:TRIGGER :A:LOGIC:CLASS SETHOLD;FUNC
TION
2-770 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual

Commands Listed in Alphabetical Order
AND;THRESHOLD
: CH1 20.0000E-3;CH2 0.0000;CH3
0.0000;CH4 0.0000;:TRIG GER:A:LOGIC:INPUT:CH1
X;CH2 X ;CH3 X;CH4 X;CLOCK:SOURC E NONE;EDGE
RISE;:TRIGGER:A:LOGIC:PATTERN:INPUT:CH1 X; CH2
X;CH3 X ;CH4 X;:TRIGGER:A:LOGIC:PATTERN:WHEN
TRUE;WHEN:LESSLIMIT 4.0000E-9;MO RE LIMIT
4.0000E-9;
:TRIGGER:A:SETHOLD:CLOCK:SOURCE CH1;EDGE
RISE;THRESHOLD 20.000 0E-3;:TRIGGER :A:SETHOLD:DATA:SOURCE
CH2;THRESHOLD 0.0000;:T RIGGER:A:SETHOLD:HOLD TIME
4.0000E-9;SETTIME 4.0000E-9;:TRIGGER:A:PULSE:CLASS
TRANSITION;:TRIGGER:A:PULSEWIDTH:SOURCE
CH1;POLARITY POSITIVE;W HEN LESSTHAN;WIDTH
4.0000E-
9;:TRIGGER:A:RUNT:SOURCE CH1;POLARITY POSITIVE;WHEN
OCCURS;WIDTH 4.0000E-9; :TRIGGER:A:TRANSITION:SOURCE
CH1;POLARITY POSITIVE;W HEN SLOWER;DELTATIME
4.0000E-9;:TRIGGER:A:VIDEO :POLARITY POSITIVE;SOUR CE
CH1;STANDARD NTSC;SYNC ALLLINES;HOL DOFF:FIELD 0 .0000;:
TRIGGER:A:VIDEO:CUSTOM:FORMAT PROGRESSIVE;SCAN
RATE1
5K;:TRIGGER:A:VIDEO:LINE 1;H DTV:FOR MAT
HD1080I60;:TRIGGER:A:BUS:SOURCE B1;B1:I2C:CONDITION
START;DATA:VALUE "XXXXXXXX";SIZE 1; START 0.0000;DIRECTION
NOCARE;:TRIGGER:A:BUS:B1:I2C:ADDRESS:MODE
ADDR7;TYPE USER;VALUE
"XXXXXXX";:TRIGGER:A:BUS:B1:SPI:CONDITION MOSI;DATA:OUT
:VA
LUE "XX XXXXXX";:TRIGGER:A:BUS:B1:SPI:DATA:IN:VALUE
"XXXXXXXX";:TRIGGER:A:BUS: B1:SPI:DATA:SIZE 1;START 0.0000;:
TRIGger:A:BANDWidth:RF:HIGH? (Query Only)
Returns the high end of the power level trigger bandwidth range as an NR3 value
in hertz. This is the value that is displayed in the user interface when the trigger
source is RF.
This command requires the MDO4TRIG advanced trigger feature to be enabled.
Conditions
MDO4000/B and MDO4000C with option SA3 or SA6 only.
Group
Trigger
Syntax
TRIGger:A:BANDWidth:RF:HIGH?
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-771