MDO3000 Programmer Manual.pdf - 第856页
Commands Listed in A lphabetical Order Group T rigger Syntax TRIGger:A:BUS :B<x>:I2C:DATa:V ALue <QStrin g> TRIGger:A:BUS :B<x>:I2C:DATa:V ALue? Arguments <QString> is the binary data string, wher…

Commands Listed in Alphabetical Order
Group
Trigger
Syntax
TRIGger:A:BUS:B<x>:I2C:DAT
a:DIRection {READ|WRITE|N OCARE}
TRIGger:A:BUS:B<x>:I2C:DATa:DIRection?
Arguments
READ specifies read as the data direction.
WRITE specifies write as the data direction.
NOCARE specifies either as the data direction.
Examples
TRIGGER:A:BUS:B1:I2C:DATA:DIRECTION WRITE specifies write as the
I
2
C data direction.
TRIGger:A:BUS:B<x>:I2C:DATa:SIZe
This command specifies the length of the data string in bytes to be used for an
I2C trigger if the trigger condition is DATA or ADDRANDDATA. Applies to
bus <x>, where x
Conditions
This command requires a DPO4EMBD application module for DPO/MSO4000B
and MDO4000/B/C models or MDO3EMBD application module for
MDO3000 models.
Group
Trigger
Syntax
TRIGger:A:BUS:B<x>:I2C:DATa:SIZe <NR1>
TRIGger:A:BUS:B<x>:I2C:DATa:SIZe?
Arguments
<NR1> is the le
ngth of the data string in bytes.
TRIGger:A:BUS:B<x>:I2C:DATa:VALue
This command specifies the binary data string used for I2C triggering if the trigger
condition is
DATA or ADDRANDDATA.
B<x>
Conditions
This command requires a DPO4EMBD application module for DPO/MSO4000B
and MDO4000/B/C models or MDO3EMBD application module for
MDO3000 models.
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-825

Commands Listed in Alphabetical Order
Group
Trigger
Syntax
TRIGger:A:BUS:B<x>:I2C:DATa:VALue <QString>
TRIGger:A:BUS:B<x>:I2C:DATa:VALue?
Arguments
<QString> is the binary data string, where the number of bits is 8 times the
number of bytes specified. The only allowed characters in the string are 0, 1,
and X.
TRIGger:A:BUS:B<x>:LIN:CONDition
This command specifies the trigger condition for LIN. B<x>
Conditions
This command requires a DPO4AUTO application module for DPO/MSO4000B
and MDO4000/B/C models or MDO3AUTO application module for
MDO3000 models.
Group
Trigger
Syntax
TRIGger:A:BUS:B<x>:LIN:CONDition
{SYNCField|IDentifier|DATA|IDANDDATA
|WAKEup|SLEEP|ERROR}
TRIGger:A:BUS:B<x>:LIN:CONDition?
Arguments
SYNCField sets the LIN trigger condition to sync field.
IDentifier sets the LIN trigger condition to identifier.
DATA sets the LIN trigger condition to data.
IDANDDATA sets the LIN trigger condition to id and data.
WAKEup sets the LIN trigger condition to wake up.
SLEEP sets the LIN trigger condition to sleep.
ERROR sets the LIN trigger condition to error.
Examples
TRIGGER:A:BUS:B1:LIN:CONDITION ERROR sets the LIN trigger condition
to error.
TRIGGER:A:BUS:B1:LIN:CONDITION? might return
TRIGGER:A:BUS:B1:LIN:CONDITION SYNCFIELD indicating the LIN trigger
condition is sync field.
2-826 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual

Commands Listed in Alphabetical Order
TRIGger:A:BU
S:B<x>:LIN:DATa:HIVALue
This command specifies the binary data string to be used for LIN trigger if trigger
condition is ID or IDANDDATA.
B<x>
Conditions
This command requires a DPO4AUTO application module for DPO/MSO4000B
and MDO4000/B/C models or MDO3AUTO application module for
MDO3000 models.
Group
Trigger
Syntax
TRIGger:A:BUS:B<x>:LIN:DATa:HIVALue <QString>
TRIGger:A:BUS:B<x>:LIN:DATa:HIVALue?
Arguments
<QString> is a quoted string that is the binary data string used for LIN trigger if
the trigger condition is ID or IDANDDATA.
Examples
TRIGGER:A:BUS:B1:LIN:DATA:HIVALUE "11001010" sets the high value
to
11001010.
TRIGGER:A:BUS:B1:LIN:DATA:HIVALUE? might return
TRIGGER:A:BUS:B1:LIN:DATA:HIVALUE "XXXXXXXX"indi
cating
the high value is don't care.
TRIGger:A:BUS:B<x>:LIN:DATa:QUALifier
This command specifies the LIN data qualifier. This only applies if the trigger
condition is IDANDDATA or DATA.
B<x>
Conditions
This command requires a DPO4AUTO application module for DPO/MSO4000B
and MDO4000/B/C models or
MDO3AUTO application module for
MDO3000 models.
Group
Trigger
Syntax
TRIGger:A:BUS:B<x>:LIN:DATa:QUALifier
{LESSthan|MOREthan|EQual|UNEQual|LESSEQual|MOREEQual|
INrange|OUTrange}
TRIGger:A:BUS:B<x>:LIN:DATa:QUALifier?
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-827