Command Groups NOTE. Analog, d igital, and RF waveforms and tra ces and those wa veforms and traces derived from them ( such as math and r efer ence) can be saved to an ISF fi le. When saving all channels in ISF format, a…

100%1 / 1158
Command Groups
Table 2-37: RF Commands (cont.)
Item Description
SEARCH:SPECTral:LIST? This query will return a list o f all automatically placed peak markers displayed in the frequency
domain graticule. (To return information about manual markers, use the MARKER:M<x>
commands.)
SELect:RF_AMPlitude This command switches the RF Amplitude vs. Timetracedisplayonoroffinthetimedomain
graticule.
SELect:RF_AVErage This command switches the R F Average trace display on or off in the frequency domain
graticule.
SELect:RF_FREQuency This command switches the RF Frequency vs. Time trace display on or off in the time domain
graticule.
SELect:RF_MAXHold This command switches the frequency domain Max Hold trace display on or off in the
frequency domain graticule.
SELect:RF_MINHold This command switches the frequency domain Min Hold trace display on or off in the frequency
domain graticule.
SELect:RF_NORMal This command switches the frequency domain Normal trace display on or off in the frequency
domain graticule.
SELect:RF_PHASe This command switches the RF P hase vs. Time trace display on or off in the time domain
graticule.
Save and Recall Command Group
UsethecommandsintheSaveandRecallCommand Group to store and retrieve
wav
eforms and settings. When you save a setup, you save all the settings of the
oscilloscope. When you recall a setup, the oscilloscope restores itself tothestate
it was in when you originally saved the setting.
NOTE. External lestructureisasfollows:
E: is the USB memory device plugged into the rst USB port o n the front of the
oscilloscope.
F: is the USB memory device plugged into the second USB port on the front of the
oscilloscope. (Not available for MDO3000 models).
G: and H: are the USB memory device plugged into the USB ports on the rear o f
the oscilloscope. (Not available for MDO3000 models.)
I: Z are for network storage.
2-58 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual
Command Groups
NOTE. Analog, d
igital, and RF waveforms and traces and those waveforms and
traces derived from them ( such as math and reference) can be saved to an ISF
le. When saving all channels in ISF format, a group of les will be saved.
Each will have the same value for XXXX, but the YYY values will be set to the
different channels that were turned on when the Save All Waveforms operation
was performed.
The oscilloscope can save, but not recall, RF acquisitions as .TIQ les. You can
import .TIQ les into Tektronix SignalVu-PC software (PC based), SignalVu
software (
oscilloscope based), RSAVu software (PC based) or into a Tektronix
real-time spectrum analyzer for pulse analysis and demodulation analysis.
Table 2-3
8: Save and Recall Commands
Command
Descript
ion
FACtory Resets the oscilloscope to factory default s ettings
*RCL
Recalls
saved oscilloscope settings
RECAll:SETUp
Recall
s saved oscilloscope settings
RECAll:SETUp:DEMO<x> This command recalls one of the 6 specied built-in demonstration setups of RF functionality.
<x> can be 1 through 6. The demonstrations include 1. Multiple Peaks, 2. spectrogram, 3.
VCO/P
LL Turn On, 4. ASK Modulation, 5. Frequency Hop and 6. Capture BW.
RECA
ll:WAVEform
This
command (no query form) recalls a stored waveform to a reference memory location,
and, for instruments with the arbitrary waveform feature, to arbitrary waveform edit memory
(EMEM). Only the rst waveform in the .CSV le is recalled for multiple waveform .CSV les.
Reca
ll of digital waveforms (D0 through D15) is not supported.
*SA
V
Sto
res the state of the oscilloscope to a s pecied memory location
SA
Ve:ASSIgn:TYPe
Th
is command species the assignment of the save button
SA
Ve:EVENTtable:{BUS<x>|B<x>}
Sa
ves event table data from bus<x> to a specied le
S
AVe:IMAGe
S
aves a capture of the screen image to the specied le
S
AVe:IMAGe:FILEFormat
T
his command speci es the le format to use for saving screen images. The le format is
not automatically determined by the le name extension. You need to choose a le name
with an extension w hich is consistent with the selected le format
SAVe:IMAGe:INKSaver This command species the current inksaver setting for the SAVe:IMAGe command
SAVe:IMAGe:LAYout This command species the layout to use for saved screen images
SAVe:SETUp Saves the state of the oscilloscope to a specied memory location or le
SAVe:WAVEform This command saves the specied w aveform to the specied destination reference memory
slot, or saves the specied waveform(s) to the specied destination le. The same function
can be accomplished from the front panel Menu->save waveform m enu. The type of le
saved is dependent upon the SAVe:WAVEform:FILEFormat command.
SAVe:WAVEform:FILEFormat This command species the format for saving waveforms.
This command species the le format to be used when saving waveforms either an
internal format, .ISF, or an external comma-delimited spreadsheet format, .CSV, that
includes waveform header and timing information.
SAVe:WAVEform:FILEFormat:
RF_BB_IQ
This command species the le format for saving the RF baseband I & Q data. The default
format is TIQ.
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-59
Command Groups
Table 2-38: Save and Recall Commands (cont.)
Command
Description
SAVe:WAVEform:GATIng Species whether save waveform operations should save the entire w aveform or a specied
portion of the waveform
SETUP<x>:DATE? Returns the date when the specied oscilloscope setup was saved
SETUP<x>:LABEL This command species the specied oscilloscope setup label
SETUP<x>:TIME ? Returns the time when the specied oscilloscope setup was saved
2-60 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual