Commands Listed in A lphabetical Order Group T rigger Syntax TRIGger:A:VID eo:STANdard {NTSc|PAL|SECAM| BILevelcustom| TRILevelcusto m| HD480P60| HD576P50|HD720P3 0|HD720P50 |HD720P60|HD8 75I60|HD1080P24| HD1080SF24|HD10…

Commands Listed in Alphabetical Order
Arguments
POSitive argum
ent sets the oscilloscope to trigger on a positive video sync pulse.
NEGative argument sets the oscilloscope to trigger on a negative video sync
pulse.
Examples
TRIGGER:A:VIDEO:POLARITY NEGATIVE sets the oscilloscope to trigger on a
negative video pulse.
TRIGGER:A:VIDEO:POLARITY? might return :TRIGger:A:VIDeo:POLARITY
POSITIVE
indicating th
at the oscilloscope is set to trigger on a positive video
sync pulse.
TRIGger:A:VIDeo:SOUrce
This command sets the source channel to use for triggering on video signals
(CH1-4).
Conditions
This command requires a DPO4VID application module to use any standard
besides NTSc, PAL, or SECAM, except for MDO3000 models, which do not
require an application module.
Group
Trigger
Syntax
TRIGger:A:VIDeo:SOUrce
{CH1|CH2|CH3|CH4}
TRIGger:A:VIDeo:SOUrce?
Arguments
CH1–CH4 specifies the input channel to use as the A video trigger.
Examples
TRIGGER:A:VIDEO:SOURCE CH1 sets Channel 1 as the source for the video
trigger operation.
TRIGGER:A:VIDEO:SOURCE? might return :TRIGger:A:VIDeo:SOURCE CH2
indicating that the source for the video trigger operation is set to Channel2.
TRIGger:A:VIDeo:STANdard
This command sets the standard to use for triggering on video signals.
Conditions
This command requires a DPO4VID application module to use any standard
besides NTSc, PAL, or SECAM, except for MDO3000 models, which do not
require an application module.
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-917

Commands Listed in Alphabetical Order
Group
Trigger
Syntax
TRIGger:A:VIDeo:STANdard {NTSc|PAL|SECAM|BILevelcustom|
TRILevelcustom| HD480P60|HD576P50|HD720P30|HD720P50
|HD720P60|HD875I60|HD1080P24|HD1080SF24|HD1080I50
|HD1080I60|HD1080P25|HD1080P30|HD1080P50|HD1080P60}
TRIGger:A:VIDeo:STANdard?
Related Commands
TRIGger:A:VIDeo:CUSTom{:FORMat|:TYPe}, TRIGger:A:VIDeo:CUSTom:
LINEPeriod, TRIGger:A:VIDeo:CUSTom:SYNCInterval
Arguments
NTSc sets the oscilloscope to trigger on video signals that meet the NTSC
525/60/2:1 standard (a line rate of 525 lines per frame and a field rate of 60 Hz).
PAL sets the oscilloscope to trigger on video signals that meet the NTSC
625/50/2:1 standard (a line rate of 625 lines per frame and a field rate of 50 Hz).
SECAM sets the oscilloscope to trigger on video signals that meet the SECAM
standard.
BILevelcustom sets the oscilloscope to trigger on video horizontal scan rate
parameters defined by the TRIGger:A:VIDeo:CUSTom{:FORMat|:TYPe},
TRIGger:A:VIDeo:CUSTom:LINEPeriod and TRIGger:A:VIDeo:CUSTom:
SYNCInterval commands. (Requires installation of a DPO4VID application
module for 4000 series models.)
TRILevelcustom sets the oscilloscope to trigger on video horizontal scan rate
parameters defined by the TRIGger:A:VIDeo:CUSTom{:FORMat|:TYPe},
TRIGger:A:VIDeo:CUSTom:LINEPeriod command. (Requires installation of a
DPO4VID application module for 4000 series models.)
HD480P60|HD576P50|HD720P30|HD720P50|HD720P60|HD875I60
|HD1080P24|HD1080SF24|HD1080I50|HD1080I60|HD1080P25|HD1080P3 0
|HD1080P50|HD1080P60
set the oscilloscope to trigger on an HDTV video
signal that meets standards defined in the following table. (Requires installation
of a DPO4VID application module for 4000 series models.)
Table 2-57: Available HDTV formats
HDTV format Description
480p60
525 lines (480 active), 640 or 704 x 480
pixel, progressive, 60 fps
576p50
EDTV with 625 lines (576 active), 1024 x 576
pixel, progressive, 50 fps
720p30
750 lines (720 active), 1280 x 720 pixel,
progressive, 30 fps
2-918 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual

Commands Listed in Alphabetical Order
Table 2-57: Available HDTV formats (cont.)
HDTV format Description
720p50
750 lines (720 active), 1280 x 720 pixel,
progressive, 50 fps
720p60
750 lines (720 active), 1280 x 720 pixel,
progressive, 60 fps
875i60
875 active lines, interlaced, 60 fps
1080i50
1125 Lines (1080 active), 1920 x 1080 pixel,
interlaced, 50 fps
1080i60
1125 lines (1080 active), 1920 x 1080 pixel,
interlaced, 60 fps
1080p24
1125 lines (1080 active), 1920 x 1080 pixel,
progressive, 24 fps
1080sF24
1125 lines (1080 active), 1920 x 1080 pixel,
progressive (sF), 2 4 fps
1080p25
1125 lines (1080 active), 1920 x 1080 pixel,
progressive, 25 fps
1080P30
1125 lines (1080 active), 1920 x 1080 pixel,
progressive, 30 fps
1080P50
1125 lines (1080 active), 1920 x 1080 pixel,
progressive, 50 fps
1080P60
1125 lines (1080 active), 1920 x 1080 pixel,
progressive, 60 fps
Examples
TRIGGER:A:VIDEO:STANDARD NTSC sets the oscilloscope to trigger on
NTSC-standard video signals.
TRIGGER:A:VIDEO:STANDARD? might return :TRIGger:A:VIDeo:STANDARD
NTSC
indicating that the oscilloscope is set to trigger on an NTSC format.
TRIGger:A:VIDeo{:SYNC|:FIELD}
This command sets the video field to use for triggering on video signals (odd
fields, even fields, all fields, all lines, numeric).
Conditions
This command requires a DPO4VID application module to use any standard
besides NTSc, PAL, or SECAM, except for MDO3000 models, which do not
require an application module.
Group
Trigger